Network; Diagnostics - Johnson & Johnson Sterrad 100NX Service Manual

Hide thumbs Also See for Sterrad 100NX:
Table of Contents

Advertisement

Network

Diagnostics

STERRAD
®
100NX
®
Service Guide
®
These STERRAD
100NX
be configured by an ASP Field Service Engineer working in conjunction
with the customer's network administrator. Call ASP for current
information on setting up the network.
The diagnostics function prompts you to select one of two types of tests
(either Temperature or Other Tests). If Other Tests is selected, the
sterilizer runs ten operator-assisted tests of the sterilizer subsystems. You
may skip one or more tests in the automatic sequence by touching
Cancel when a test begins. This causes the program to advance to the
next test in the sequence.
The ten tests and the sterilizer elements that are tested are listed in the
order in which they occur in the following table.
Touch Diagnostics to start automatic diagnostic testing of the sterilizer.
Note: The duration of the Temperature Test is 15-60
minutes depending on system temperature.
Figure 16. Touch Diagnostics to Run a Series of Tests to
Help You Resolve System Messages.