Summary of Contents for ForaCare FORA ACTIVE P30 Plus Series
Page 1
Příručka vlastníka Tonometr Tonometer Návod na použitie FORA P30 Plus / FORA P30 plus BT MedNet EC-REP GmbH ForaCare Suisse AG Neugasse 55, 9000 St. Gallen, Switzerland Borkstraβe 10, 48163 Mϋnster, For self-testing / Zur Selbstmessung / www.foracare.ch Germany Pour l’automesure / Per l’automisurazione / Para autotest / Para auto-teste / Voor zelftest / Pentru autotestare / Pro vlastní...
Contents Maintenance Safety Information Changing Battery Introduction Using an AC Adapter Intended Use Caring for Your Device Test Principle Device Storage Product Overview Troubleshooting Screen Display Error Message Getting Started Blood Pressure Measurement 11 Testing Your Blood Pressure Symbol Information Fitting the Cuff Proper Measurement Position Specifications...
Safety Information (a) the death of a patient, user, or other people, Read the following Safety Information thoroughly before using (b) the temporary or permanent serious deterioration of a the device. patient’s, user’s or other person’s state of health, • Use this device ONLY for the intended use described in this (c) a serious public health threat.
Product Overview Screen Display P30 Plus BT P30 Plus 1. Date 8. Day- / Night-Time Symbol 1. Screen Display 7. Battery Compartment 2. Time 9. Day Average Symbol 2. M Button 8. AC Adapter Port 3. Systolic Pressure Symbol 10. Units for Blood Pressure 3.
Getting Started Before using the device for the first time, do the following initial setup: Step 1: Enter the Setting Mode Press and hold for 3 seconds until the device turns on. Step 2: Configuring the Settings (Date, Time, Time Format, Set [Year] Set [Month] Set [Day]...
Testing Your Blood Pressure 3. Stretch your left (right) arm in front of you with your palm facing up. Slide and place the cuff onto your arm with the air Before Measument tube and artery mark region (in red) toward the lower arm. •...
Important! If the cuff is relatively lower (higher) than the heart, the obtained blood pressure value could be higher (lower) than the actual value. A 15 cm difference in height may result in an error around 10 mmHg. 4. Press .
Page 8
is more than 15mmHg, the time interval for third measurement will be 40 seconds. 3. After taken three measurements, the results are averaged to produce the blood pressure measurement with AVERAGE symbol display on the screen. Press to turn the device off. 3.
Reviewing Test Results 1. Press and hold for 3 seconds until the “ ” icon appears on the screen. Your device stores the 200 (P30 Plus BT only) most recent blood pressure test results along with respective dates and times in the 2.
For P30 Plus and P30 Plus BT Category Systolic Diastolic Press to review the average and general memory recall mode. Optimal < 120 mmHg < 80 mmHg Note: Normal 120 –129 mmHg and/or 80–84 mmHg High normal 130 –139 mmHg and/or 85–89 mmHg •...
To use an AC adapter, do the following: • the “ “ icon appears with E-b: This indicates 1. Connect one end of the AC adapter to the AC adapter jack of the power is not enough for a measurement. the device.
Troubleshooting Blood Pressure Measurement If you follow the recommended action but the problem persists, Symptom Cause What To Do or error messages other than the ones below appear, please call Replace the Batteries exhausted. your local customer service. batteries. Nothing is displayed Batteries incorrectly Check that the Error Message...
Symbol Information Maximum 200 memory records Memory (P30 Plus BT); 60 (P30 Plus) Symbol Referent Automatic power off if system is Power Saving idle for 3 minutes Read instructions before use 10°C to 40°C (50°F to 104°F), below System Operating Conditions Manufacturer 85% RH Device Storage/Transport...
RF disturbances are controlled. The customer or the product manufactured by ForaCare Suisse shall be free of defects in user of the FORA P30 Plus Series can help prevent electromagnetic interference by maintaining a...
Page 15
Manufacturer’s declaration-electromagnetic immunity Manufacturer’s declaration-electromagnetic immunity The FORA P30 Plus Series is intended for use in the electromagnetic environment (for home healthcare The FORA P30 Plus Series is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. and professional healthcare) specified below.
Page 16
Manufacturer’s declaration-electromagnetic immunity Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment The FORA P30 Plus Series is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. The customer or the user of the FORA P30 Plus Series should assure that it is used in such an environment. Band a) Test frequency Maximum power...