Device for the 24hour measurement of arterial function (stiffness) and peripheral blood pressure User’s Manual Please read the user’s manual carefully before the first use! With wireless communication Bluetooth TensioMed® Arteriograph24-03v6-00 Revised: 01-07-2021...
1. Introduction and the intended use of the device Thank you for choosing TensioMed® Arteriograph24™. The intended use of the device The TensioMed® Arteriograph24™ device measures the brachial blood pressure, the heart rate, and the arterial function parameters intermittently throughout one or more days.
1.3. Warranty TensioMed Ltd. offers a 2-year warranty for the device. Any repair and device calibration within or beyond the warranty period are performed by TensioMed Ltd. in the service station of the company (97. Kossuth Lajos str. Budapest H- 1181).
For optimum performance it is recommended to use Nickel-Metal Hybrid (Ni-MH) rechargeable batteries, or Nickel Cadmium (NiCd) rechargeable batteries, with1500 mAh capacity, size AA. The TensioMed® Arteriograph24™ can also be used with 1.5 V long life batteries, size AA. ________________________________________________________________________________________________________________________________...
Page 8
Devices, which are withdrawn from the use, should be collected selectively in order to provide an effective recycling, reducing the unfavorable effects on the environment and human health. Only use with cuffs supplied by TensioMed Ltd! Use of cuffs supplied by a third party can lead to erroneous measurement results.
Page 9
The Arteriograph24™ device can be used on pregnant women, as well. However, the validation of BP measurements of the device has not been done on preeclamptic patients, yet. Do not use the device on neonates or infants! Cannot be used on persons below 3 years of age! ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
2. The TensioMed® Arteriograph24™ device 2.1. Explanation of symbols The front of the device is shown in Figure 1. Function button (Offering four menu options) Command symbols Air connector Figure 1. ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
Page 11
18 Level of protection against any liquid or grainy material filtering into the device (IP =2: Protected against solid foreign objects of 12,5mm and greater =0: Not protected against water Figure 2. ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
TensioMed® Arteriograph24™ device (Figure 1), is supplied with the following accessories: 3 different cuffs TensioMed® TensioWin™ software on CD and on flash drive with Instruction for use Instruction for use The cuff is the patient's part of the device that necessarily comes into physical contact with the patient during the intended use of the device to perform its function (see section 2.4.5...
PC indicating an active Bluetooth service. Then if the device is within 10 m from the Bluetooth, the computer will get into connection with the device via the software. To transmit data, it is necessary to use the TensioMed® TensioWin™ program, of course.
If the battery voltage is adequate, the device will be D 09‐39 ready for measuring and the current computer time will be displayed. TensioMed® Arteriograph24™ is ready for operation: 2.4.1.1. One short push of the button starts a manual measurement.
If the memory of the device is full, this will appear on the FULL display. 2.4.1.3. Three short pushes on the button allows the D 09‐39 patient to indicate the time of going to bed and waking ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
- the device and the PC will connect automatically. This sign will appear on the LCD. The actual data transmission does not happen, yet this can be started by the TensioMed® TensioWin™ program. During the operation of the Bluetooth communication CONNECT...
(e.g. because of movement), the measurement will be interrupted. With the TensioMed® TensioWin™ program - in case of a faulty measurement - it is possible to set the device to repeat it according to the measuring plan after 1 minute.
Although the Arteriograph24™ is appropriate for longer (e.g. 48 hours) measurement period, but in this case the physician should evaluate the advantages and additive clinical value from the extended measuring period, compared to the inconvenience, caused by the longer duration of the measuring process. ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
Average difference from the auscultatic (Korotkov) measurements: (systolic / diastolic): 0.5/-0.4 mmHg The range of the difference (systolic/diastolic): 2.8/2.8 mmHg Pressure sensor: Piezo-resistive Inflation: Automatic motor-driven pump Safety: Maximum inflation 300 mmHg ; Maximum measurement time: 180 seconds Deflation: Automatic, stepwise ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...
3.1. Electromagnetic compatibility Electromagnetic emissions The TensioMed® Arteriograph™ TD3 device is intended for use in the electromagnetic environment specified below. The customer or the user of the TensioMed® Arteriograph™ TD3 device should assure that it is used in such an environment.
Page 21
Electromagnetic immunity The TensioMed® Arteriograph™ TD3 device is intended for use in the electromagnetic environment specified below. The customer or the user of the TensioMed® Arteriograph™ TD3 device should assure that it is used in such an environment. Immunity test...
See sections 1., 2.4.1.1, 2.4.1.2., 2.4.1.3. and 2.4.1.4. for details. Use environment Programmed and measurement results data read back in a medical environment, automatic measurements during the daily life (active and/or sleep periods) of the patients. Operating principle Oscillometric method ________________________________________________________________________________________________________________________________ TensioMed® Arteriograph24™...