6.3.5 Temperature field check and adjustment using a
calibrated reference probe
Complete the steps listed in
(page 62)
1. Check that the probes are located in equal conditions and wait until the readings have
stabilized (can take 30 minutes or more). If you are close to the probes, do not breathe in
their direction.
2. Press ADJUST to continue adjusting.
3. Choose To same as THI/II from the MI70 adjustment menu and press SELECT. MI70
automatically recognizes which port the HMP70 series probe is connected to.
4. Press YES to confirm the adjustment.
If the temperature difference between the reference probe and the
transmitter is too large, the adjustment cannot be done (MI70 notifies you of
this). The available reserve for T adjustment of the transmitter depends on
the initial temperature calibration of the unit.
5. The adjustment is done. Press BACK and EXIT to return to the basic display.
6. Turn off the MI70 and detach the connection cable.
66
Adjusting RH and T with MI70 handheld indicator
before continuing with this adjustment option.