Instructions for use
VA131-17
VA131 ENG - Rev. 4
VA131-22

Advertisement

Table of Contents
loading

Summary of Contents for W&H Lisq VA131-17

  • Page 1 Posizione logo Instructions for use VA131-17 VA131 ENG - Rev. 4 VA131-22...
  • Page 2 Symbols Symbols displayed on the product and/or used in this manual: WARNING! ATTENTION! General explanations, without risk to Risk of injury To prevent damage occurring persons or objects Call service HOT SURFACES! HOT STEAM! Risk of burns Risk of burns Thermo washer Consult instruction for use Do not dispose of with...
  • Page 3: Table Of Contents

    Contents 1. Introduction ....................... . 2.
  • Page 4: Introduction

    1. Introduction About this manual This manual contains the Instructions for Use of the W&H sterilizers VA131-17 and VA131-22, hereinafter referred to as LISA 17 and LISA 22. All drawings, images and texts contained in this manual are the property of the manufacturer. All rights reserved.
  • Page 5 Introduction Qualifications of the users There are two types of users who may operate the sterilizer: The ADMINISTRATOR (ADM) is the head of the clinic/practice, who is legally responsible for the efficiency of the hygiene protocol in place as well as the sterilization process. He/she is also responsible for the USERS’ training and the correct operation and maintenance of the equipment.
  • Page 6: Unpacking

    2. Unpacking If the sterilizer comes from a cold location, wait until all external and internal surfaces are free from moisture before switching it ON. The sterilizer must be removed from the box and transported by two people. Weight: LISA 17: 50,2 kg LISA 22: 51,7 kg...
  • Page 7 Contents of the package Tray (5 pieces total) Mains cable Reversible rack USB pen drive Tray holder Fast guide Declaration of conformity Documentation CD Warranty card Drain tube Works tests report Maintenance sheet...
  • Page 8: Safety Advice

    3. Safety advice The user is responsible for the proper installation, the correct use and maintenance of the sterilizer in accordance with the • instructions listed in this manual. The safety devices of the sterilizer are impaired when the product itself is not installed, used and serviced in accordance •...
  • Page 9: Installation, Programming And Setup

    4. Installation, programming and setup Placement Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials. Do not place the sterilizer so that it is difficult to open the service door and operate on the controls behind it. Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
  • Page 10 Water tanks Filling the clean water tank Switch the sterilizer ON; • Remove the cap from the tank inlet; • Fill the water tank with app. 4 litres of distilled or demineralized water; • Once the clean water tank is almost full, an audible tone sounds; stop filling; •...
  • Page 11 Chamber furniture The chamber furniture consists of the trays and the tray rack. Tray rack Insert the rack into the sterilizer chamber, align it at the center/bottom of the chamber and push it gently into position until it clicks. The chamber rack is reversible and can accommodate 5 trays horizontally or 3 cassettes / containers vertically.
  • Page 12 Controls, commands, connections (front side) Service door Bacteriological filter Tank filling cover - cap Mains switch Water tank cover Chamber door USB port Touch screen Identification label Door pin Door seal Service door Used water drain (grey) Clean water drain (blue) Dust filter Sterilization chamber...
  • Page 13 Connections (rear side) Serial communication port (male) Ethernet port Mains plug socket USB ports Mains cable guide Used water drain Condenser grid Pressure safety valve cover Water supply inlet The water supply system must deliver demineralized water meeting the requirements as listed in ANNEX 7. Do not add any chemical / additive to the water.
  • Page 14 Controls and commands Switching ON the sterilizer Press the mains switch behind the service door to switch ON the sterilizer. The visual indicator on the mains switch turns green and the CYCLE SELECTION screen (HOME screen) (see next page) appears. Energy management: the SLEEP mode In the SLEEP mode the display remains dark and the sterilizer chamber is no longer heated to save energy.
  • Page 15 Touch screen and Commands TYPICAL SELECTION SCREEN (example: CYCLE SELECTION SCREEN/HOME screen) HEADER: Shows the title/purpose of the screen or the cycle number and the current date and time. ICONS: show the available options/actions. By touching an icon, you: select an item in the menu, or undertake an action.
  • Page 16 Touch screen and Commands TYPICAL KEYBOARD SCREEN (to enter letters and numbers) HEADER: Shows the title/purpose of the screen and the current date and time. DATA ENTERING FIELD: shows the letters/numbers you are typing. TYPING FIELD: touch a key to type into the data entering field. WHAT IT DOES SPECIAL KEYS: see table Activates capital letters.
  • Page 17 Touch screen and Commands TYPICAL SCREEN TO SELECT FROM A LIST (example: language selection) The active selection (example: English) appears in a frame in green colour. To activate another selection (example: French), simply touch the item in the list, then press BACK. If the list is longer than displayed, you may scroll it by touching the icons SCROLL UP –...
  • Page 18 Control icons The following icons appear in the bottom or in other specific areas of the screen. They permit to activate specific functions that are appropriate for the situation (e.g. choosing options, entering values, replacing consumables, etc.). ICON WHAT IT DOES / MEANS SLEEP MODE Enters or exits the SLEEP mode.
  • Page 19 Control icons ICON WHAT IT DOES / MEANS ECO-DRY If shown during a cycle, reminds you that the ECO – DRY option with extended drying is operating automatically. ECO-DRY plus If shown during a cycle, reminds you that the ECO – DRY PLUS option is operating automatically. UP / DOWN Increases (UP) and decreases (DOWN) a number or a parameter.
  • Page 20 List of the MENU options MENU SUB-MENU WHAT IT DOES ets the language from a "selection from list" screen. Sets time and date values and format. TIME AND DATE: Touch the figure you want to change and use the UP / DOWN keys to set. FORMAT: Touch the format you want to set.
  • Page 21 List of the MENU options Continued from the previous page MENU SUB-MENU WHAT IT DOES See section CONNECTION TO THE LAN. Available only if a WI-FI or a ETHERNET connection is set. Provides information about the status of the connection. Available only if a connection to a network is set.
  • Page 22 List of the MENU options MENU SUB-MENU WHAT IT DOES Shows the list of the recorded cycles. Scroll the list with the UP / DOWN keys or with the scroll bar. Displays the report of a cycle by touching it in the list. Prints the HTML file or the traceability labels by touching the icons in the footer.
  • Page 23 List of the MENU options MENU SUB-MENU WHAT IT DOES Formats your USB pen drive in case you want to erase all the data on it. Select the printer model. Calibrate (Adjusts the label printer to the edge of the label). Prints a Test label (**) Set the horizontal (X) and vertical (Y) offset values in order to have the printout duly centred in...
  • Page 24 Initial setup Initial setup Before using the sterilizer please program important parameters such as date, time, language, display backlight and contrast, sound, energy management and sterilizer name. From the CYCLE SELECTION screen (HOME screen): select MENU – SETTINGS – DEVICE, then set the parameters following the icons in the page. Go to the second page to set up further parameters...
  • Page 25 Connection to the LAN Connection via Ethernet Use a standard Ethernet cable to connect the port on the rear of the sterilizer to the Ethernet socket on the wall. Switch ON the sterilizer. The sterilizer connects automatically to the LAN after installation. Two IP configuration modes can be set - dynamic configuration: this is the default option and is applicable to most LANs.
  • Page 26 Connection to the LAN (static IP) Skip this step if your network supports dynamic IP To switch to static IP do as follows: MENU – SETTINGS – CONNECTIVITY – ETHERNET. Touch the STATIC buttons. This enables the data fields (they turn to dark grey). Then, touch each data field data fields containing invalid data appear in a dark grey frame.
  • Page 27 User authentication From the CYCLE SELECTION screen (HOME screen), select: MENU – TRACEABILITY – USER MANAGEMENT. Choose your user name and key the PIN code to log in. If you log in as the ADMINISTRATOR, you may create and delete users, reset the PIN code of a user (the initial value of all PINs, after being created or reset, is 0000, and must be changed by the user), change your own PIN code.
  • Page 28 User Management (ADMINISTRATOR only) From the CYCLE SELECTION screen (HOME screen), select: MENU – TRACEABILITY – USER MANAGEMENT and log in as the ADMINISTRATOR. To add a new user: Touch this field to type the new user ID (the keyboard screen appears). Select this option if you want to give the administrator authority to the new user.
  • Page 29 User Management (ADMINISTRATOR only) - Continued To delete a user: In the selection screen, select the user you want to delete and confirm the further step.
  • Page 30 Pin Code Management To change your PIN code: From the CYCLE SELECTION screen (HOME screen), select: MENU – TRACEABILITY – USER MANAGEMENT and log in with your own PIN code. Select the option CHANGE YOUR PIN CODE. This activates the numeric keyboard: enter your new PIN code and confirm. NOTE If you have ADMINISTRATOR authority, you will see other icons in addition to this one.
  • Page 31 Traceability options (ADMINISTRATOR only) As the ADMINISTRATOR, in the TRACEABILITY screen you may access other traceability settings. From the CYCLE SELECTION screen (HOME screen), select: MENU – TRACEABILITY and log in with your own PIN code, then select USER TRACEABILITY OPTIONS. Enable and disable the identification options by touching the relevant buttons.
  • Page 32 Cycle Exclusion (ADMINISTRATOR only) The ADMINISTRATOR can hide / unhide a cycle program to users. From the CYCLE SELECTION screen (HOME screen), select: MENU – SETTINGS– CYCLE – CYCLE EXCLUSION. NOTE you will have to enter the ADMINISTRATOR's password). Enable or disable the cycles that appear on the CYCLE SELECTION screen (HOME screen) by touching the relevant switch icon.
  • Page 33 Printer setup This menu permits to select the model of cycle report printer connected. From the CYCLE SELECTION screen (HOME screen), select: MENU – ACCESSORIES – PRINTER. A selection screen opens, select the model of the printer connected from the list. For a typical cycle report printout, see ANNEX 8.
  • Page 34 Label printer setup Select MENU – ACCESSORIES – LABEL PRINTER. It is possible that an activation key code is requested. Enter the code, CONFIRM and proceed. Choose between LOCAL PRINTER (if the printer is connected to the sterilizer) and SHARED PRINTER (if the printer is connected to another sterilizer in the network).
  • Page 35 Shared label printer To select a shared printer, go to the sterilizer to which the printer is physically connected (the sterilizer must be ON and no cycle shall be running). From the CYCLE SELECTION screen (HOME screen), select MENU – INFO. Scroll the pages and take note of the IP address (Ethernet or Wi-Fi, depending on the LAN connection).
  • Page 36 Label printing options To set the label printer options, select MENU – TRACEABILITY – LABEL PRINTER (This menu is available only if a label printer is properly configured). For a typical label layout, see ANNEX 9. AUTOMATIC PRINTING: if activated, a fixed number of labels will be printed after a successful (*) sterilization cycle.
  • Page 37 USB pen drive Lisa sterilizers are equipped with a digital cycle data recording system. All data are recorded on the removable/rewritable USB pen drive. For details about the data that are saved and the file formats, see Chapter 7 – CYCLE HISTORY. The USB connection for the pen drive is located in the front, behind the service door.
  • Page 38 Remote data storage If connected to a LAN, the sterilizer can record the cycle reports on a specific location in the network. Cycle reports are saved in HTML format files (See Chapter 7 – CYCLE HISTORY). From the CYCLE SELECTION screen (HOME screen), select MENU –...
  • Page 39 Remote data storage: settings From the CYCLE SELECTION screen (HOME screen), select MENU – SETTINGS – REMOTE DATA STORAGE – SETTINGS. Another screen (see below) is displayed. The switch-shaped icon enables the remote storage. Touch these fields to enter the path and the other settings to reach the location where the files have to be saved. NOTES: The path consists of the name of the shared folder, followed by the sub-folders, if any (and NOT the full Windows path).
  • Page 40 Remote data storage: test From the CYCLE SELECTION screen (HOME screen), select MENU – SETTINGS – REMOTE DATA STORAGE – TEST. A sequence of tests is automatically performed (see below). The sterilizer attempts to access the specified location to check if the parameters entered are valid. Positive tests are marked with a green checkmark.
  • Page 41: The Available Sterilization Cycles

    5. The available sterilization cycles In total there are four sterilization cycles available, all of them complying with the European Standard EN13060: three B-type cycles; one S-type cycle. CYCLE CYCLE NAME PURPOSE TYPE Designed for all your general items like hand instruments, handpieces, forceps, etc. Features an extended sterilization time, if this is required for your load or mandated in your country.
  • Page 42 The available sterilization cycles STERILIZATION CYCLES 134°C 134°C 121°C 134°C Sterilization temperature Sterilization pressure 3.03 barA 3.03 barA 2.04 barA 3.03 barA Duration of the holding time phase 18‘ 30” 15‘ 30” 3’ 30” Duration of the drying phase (ECO-DRY PLUS mode) 5’-16' 5’-16' 6’-24'...
  • Page 43 The available sterilization cycles For your safety and for the safety of your patients Never process objects different from those specified in the cycle program table and never exceed the maximum load weight limits specified in it. Such actions: could result in non-sterile conditions at the end of the cycle; could expose people to the hazard of cross-infections;...
  • Page 44 Profile of a Type B sterilization cycle All type B sterilization cycles feature the same basic pressure profile as shown in the graph below. The duration and the temperature sterilization phase differ between the various cycles. LEGEND Pressure Pre-heating (this is not considered a part of the cycle) Vacuum pulse (removal of air from PV1 - PV3...
  • Page 45 Profile of a Type S sterilization cycle cycle is specifically designed to sterilize unwrapped instruments, for immediate use on patients, not requiring a complete drying. Thus, the drying phase of this cycle is short, making this cycle the fastest available. LEGEND Pressure Pre-heating (this is not considered a...
  • Page 46: Running A Sterilization Cycle

    6. Running a sterilization cycle Place the sterilization load in the sterilization chamber and close the door. See ANNEXES 2 and 3 on how to properly prepare and place the load. Switch the sterilizer ON and wait the sterilizer is initialized (the CYCLE SELECTION/HOME screen is displayed). Select a cycle by touching the relevant icon.
  • Page 47 Cycle in progress Information displayed on the screen while a cycle is running Current date and time Name of the current cycle Eco-Dry plus option active Padlock icon Indicates that the door is securely locked Current cycle phase Countdown (approximate residual time Progress disc until cycle completion) Current chamber pressure...
  • Page 48 The INFO screen INFO screen While a cycle is in progress, you can view the main cycle parameters in real time. On the CYCLE IN PROGRESS screen, touch the INFO icon. The list is made of several pages. Scroll them with BACK and NEXT. The INFO screen is displayed.
  • Page 49 End of a sterilization cycle When a cycle is successfully finished, the CYCLE COMPLETED message appears on the screen. To open the door,touch the OPEN icon. Wait the door to unlock, then open the chamber door. If an error message appears at the end of the cycle, consult Chapter 8 (TROUBLESHOOTING) and, if the problem persists, call for technical service.
  • Page 50 Manual stop While a cycle is in progress, you CONFIRM icon: you confirm that you want to abort the can abort it manually at any time cycle. by touching the STOP icon. CANCEL icon: you cancel the command. The cycle goes on as originally programmed.
  • Page 51 Customization of cycle settings ECO-DRY The ECO-DRY feature (represented by the icon ) is active whenever you initiate a type B sterilization cycle. The feature is designed to provide the shortest possible cycle times, by adjusting the drying time to the load. This means the smaller the load, the faster the cycle, thus eliminating the need for selecting a quick or a flash cycle.
  • Page 52 Customization of cycle settings ECO-DRY ECO-DRY plus This is the default option. The drying time is automatically adjusted to the shortest possible for the total amount of load, but enough to ensure perfect drying of standard loads. ECO-DRY With this option the drying time is automatically adjusted to the total amount of load, but is extended compared to the previous one.
  • Page 53 Customization of cycle settings - delayed cycle start You may delay the start of a cycle (e.g., if you want to load the sterilizer in the evening and run standard sterilization cycle early the next morning before office hours). It is possible to set both the date and the time in which the selected cycle(s) will start. Load the chamber, close the chamber door, and choose the cycle you want to launch.
  • Page 54 Programming a sequence of cycles for use on daily basis You may even program a sequence of two cycles (e.g., if you want to run a vacuum test plus a HELIX test in the early morning before office hours). Load the chamber and close the chamber door. From the CYCLE SELECTION screen (HOME screen), select: MENU –...
  • Page 55 Programming a sequence of cycles for use on daily basis In the ADD CYCLES screen: In this area the programmed cycles are listed. Touch this icon to ADD a cycle to the list. Touch this icon to DELETE a cycle from the list. This opens a new screen: Touch the date/time area to open the DATE AND TIME SETTINGS screen.
  • Page 56: Cycle History

    7. Cycle history To access the cycle history: from the CYCLE SELECTION screen (HOME screen), select MENU – TRACEABILITY – CYCLE HISTORY. This displays the whole list of cycles, identified by number, name, date and time. NOTE: the sterilizer stores in memory the summarized reports of the last 400 cycles in HTML format, and the analytical reports of the last 50 cycles in SCL format (all parameters recorded every second).
  • Page 57 Cycle history Print the cycle report on paper. In the CYCLE REPORT screen, you may Save the cycle report on simply view the data, or you may touch the USB pen drive. the OTHER OPTIONS icon to go ahead. From the next screen, you Print traceability labels for have further choices: the selected cycle.
  • Page 58: Maintenance

    8. Maintenance Before carrying out any maintenance on the sterilizer, switch the unit OFF and remove the mains cable. During any maintenance work, all health, safety, cross-infection and cross-contamination protocols must be followed. Before accessing the chamber and the connected parts, be sure that the sterilizer is cold. Follow the instructions in this chapter when carrying out any maintenance on the sterilizer.
  • Page 59 Resetting the maintenance counters The sterilizer monitors the wear of consumables by counting the number of cycles executed since the last replacement. When the number of cycles is close to the maximum, a pre-alert about the concerned consumable is ..displayed.
  • Page 60 Resetting the maintenance counters from the main menu If you replace the consumables before the request of replacement appears, the counters should be manually reset through the following procedure. From the CYCLE SELECTION screen (HOME screen), select: MENU – MAINTENANCE Then select the consumable to be replaced (e.g.
  • Page 61 Monthly or 50-cycle maintenance Cleaning the door seal and the chamber face side Carefully clean the seal seat and the chamber face side with a damp, lint-free cloth. If you use a detergent solution, be careful not to get in contact with the plastic body of the front cover.
  • Page 62 Monthly or 50-cycle maintenance Cleaning the chamber filter Empty the sterilizer chamber by removing the trays and the rack. 1–2: Remove the filter cap at the back of the chamber (bottom/center) by turning it counter-clockwise. 3: Remove the cartridge filter and rinse it with tap water. 4–5–6: Insert the filter in the cap, attach the filter cap and lock it by turning clockwise.
  • Page 63 400-cycle maintenance Replacing the bacteriological filter Open the service door. Unscrew the bacteriological filter by hand (counter-clockwise). Screw on the new bacteriological filter (clockwise) and tighten it snug. Remember to reset the counter after replacement (see RESETTING THE MAINTENANCE COUNTERS). See video animation directly on Lisa display.
  • Page 64 6 month or 800-cycle maintenance Cleaning the water tanks Switch OFF the sterilizer and disconnect the mains cable. Completely drain both tanks. Leave the drain tube attached to one of the drain quick connectors. Lift the cover of the tank to gain access to the tanks. Tap with your fingers on the rubber membrane to remove any condensate.
  • Page 65 800-cycle maintenance Replacing the door seal Fully open the chamber door. Pull out the used door seal by hand (easy if seal and fingers are dry). Carefully clean the seal seat and the chamber face side with a damp, lint-free cloth. Moisten the new seal with water.
  • Page 66 4000 cycle/5 year general check and service Regular service is imperative to ensure continuous and effective operation of the sterilizer. It is recommended to carry out a general service every 4000 cycles or five years by an authorized service technician. The service includes replacement of consumables and other important internal components, a check of the entire unit with special care for the safety systems, and cleaning of areas and components that cannot be accessed by the user.
  • Page 67: Troubleshooting, Errors, Messages And Alerts

    9. Troubleshooting, Errors, Messages and Alerts During the sterilization cycle, the sterilizer is continuously monitored by a sophisticated control system. If an anomaly is detected, the cycle is aborted automatically, and the sterilizer starts a reset phase. The following information is displayed: Error symbol Error code Warning message:...
  • Page 68 Error table (continues in the next page) ERROR CODE DESCRIPTION AND ACTIONS Load cannot be considered sterile (See END OF STERILIZATION CYCLE). Check if the main switch or network circuit breaker is OFF. Check if the mains cable is properly connected Switch the sterilizer OFF and ON.
  • Page 69 Error table (continued) ERROR CODE DESCRIPTION AND ACTIONS Chamber filter clogged. Remove and clean the chamber filter. See error 13X-16X Clean the door seal and the chamber face side. Switch the sterilizer OFF and ON. Wait for the chamber to cool down. Reset the thermal overload (see next pages). Check the door seal.
  • Page 70 Messages and alerts Message / alert DESCRIPTION AND ACTIONS Fill the water tank as requested. There is not enough water in the tank to perform a cycle. The used water tank is full. Drain the water tank as requested. The door must be locked, but you didn't close it. Close the door so it can be locked.
  • Page 71 Messages and alerts Message / alert DESCRIPTION AND ACTIONS Run a vacuum test. Air was detected in the chamber: a vacuum leak is suspected. The Call for service if an anomaly is detected. cycle was completed but a vacuum test is required. Check the settings, the network resource and the Error when saving the files in the network.
  • Page 72 Resetting the safety thermostat The sterilizer is fitted with a safety thermostat to prevent overheating. If the safety thermostat operates because of too high temperatures, the Error 240 or a timeout error is displayed. The thermostat must be reset manually. Proceed as follows: - Switch the sterilizer OFF and remove the mains cable.
  • Page 73 Battery-operated door opening tool The door locking system is electrically activated. In case the door remains locked due to a black-out or an electric fault, an auxiliary unlocking procedure is available. Warning! The following procedure should be used only if necessary and only when NO RESIDUAL PRESSURE IS IN THE CHAMBER and the sterilizer has completely cooled down.
  • Page 74 Troubleshooting PROBLEM POSSIBLE CAUSE SOLUTIONS The main switch or network circuit breaker is OFF Activate the main switch or network circuit breaker (ON). The sterilizer remains switched OFF. No voltage at the socket Check the electric circuit. The power cord is not connected properly Check and connect the power cord properly.
  • Page 75 Troubleshooting PROBLEM POSSIBLE CAUSE SOLUTIONS Door seal not properly placed; seal sticking out Ensure that the door seal is evenly inserted on the entire circumference. When starting a cycle, the chamber door locks but re-opens immediately. The “Open the Remove any objects interfering with the chamber door. Check the door does not Door jammed by external objects or by the load itself door”...
  • Page 76: Recycling And Disposal

    10. Recycling and disposal LISA sterilizers are mainly built from fiber-reinforced polymers, metals and electric / electronic components. In case of disposal: - separate the various components according to the materials they are made of; - drop the sterilizer with a company that specializes on the recycling of related products; - do not abandon the sterilizer in unsecured places;...
  • Page 77: Technical Data

    ANNEX 1. Technical data TECHNICAL DATA CONFORMITY Electrical supply: STERILIZER featuring sterilization cycles conform with the following standards: Nominal voltage: 200 - 240 VAC – 50/60 Hz, single - phase 93/42/CEE Medical Device Directive (MDD) Max. current: 10 or 12 A (refer to product label) PED 2014/68/EU Pressure Equipment Directive (PED) 2012/19/EU...
  • Page 78: Maintenance Of Dental Handpieces

    ANNEX 2. Maintenance of dental handpieces External disinfection This procedure reduces the risk of infection during cleaning and maintenance of the instrument. Wear protective gloves during disinfection. Refer to the instructions of the instrument manufacturer. Avoid using abrasive disinfectants (pH-value 2.5 – 9; no chlorine based disinfectants). We recommend the use of disinfectant wipes rather than spray disinfection.
  • Page 79: Sterilization Load Preparation

    ANNEX 3. Sterilization load preparation Cleaning the instruments Clean all instruments thoroughly prior to sterilization. If possible, clean instruments immediately after use; always follow the instrument manufacturer ‘s instructions. Remove all traces of disinfectants and detergents. Rinse and dry carefully all instruments. Lubricate dental handpieces after cleaning and prior to sterilization in accordance with the manufacturer’s instructions.
  • Page 80 Loading the chamber Tubes Rinse, drain and dry tubes after washing. Place tubes on a tray allowing the ends to remain open. Do not bend tubes. Partial loads If the chamber is just partially loaded, place the load in such a way that the space in-between the trays is maximized (see example with three trays on the left).
  • Page 81: Bowie & Dick Test

    ANNEX 4. Bowie and Dick test Description The Bowie & Dick (B&D) test device is used to validate the sterilizer performance for textile load sterilization. It is made of several sheets of paper wrapped in a small packet in the middle of which there is a chemical heat-sensitive indicator sheet. How to carry out the test The test must be performed in an empty chamber (EN13060) without load but with the standard chamber accessories (chamber rack and trays) mounted.
  • Page 82: Helix Test

    ANNEX 5. Helix test Description The Helix test device is used to validate the sterilizer performance for hollow items. It consists of a 1,500 mm long tube that is open on one side and closed with a capsule on the other side. A chemical indicator strip is placed inside of the capsule.
  • Page 83: Vacuum Test

    ANNEX 6. Vacuum test Description The vacuum test is designed to validate the sterilizer performance in terms of: • Efficiency of the vacuum pump; • Tightness of the pneumatic circuit. It consists of a vacuum phase, followed by a stabilization period of 5 minutes and a testing period of 10 minutes. During the 10-minute testing period the internal pressure is monitored.
  • Page 84: Water Quality

    ANNEX 7. Water quality LISA sterilizers use distilled or demineralized water to generate steam for the sterilization process. The table below lists the maximum content of minerals and the specifications for the water used for steam sterilization (see EN13060 ANNEX C). FEED WATER SPECIFICATIONS Value/Specification Contaminants/minerals/qualities...
  • Page 85: Example Of A Cycle Data Report

    ANNEX 8. Example of a cycle data report Sterilizer brand Sterilizer model and serial number Software Revision number Surgery – practice – doctor name Cycle name Cycle counter Programmed sterilization temperature Programmed holding time Cycle start date and time Headers for the table below Cycle start Pressure and vacuum pulses Phase of pressure rise to sterilization conditions...
  • Page 86: Example Of A Traceability Label

    ANNEX 9. Example of a traceability label Sterilizer model Serial number Software release Depending on the traceability settings (see TRACEABILITY), this field may contain: - the user who released the cycle, or - the user who started the cycle, or - the sterilizer ID LISA 17 131685...
  • Page 87: Accessories And Spare Parts

    ANNEX 10. Accessories and spare parts Aluminium tray Printer model S'Print LISA17: part n. F523204x part n. 19721108 LISA 22: part n. F523205x Label printer LisaSafe Tray holder part n. 19721100 Label Printer (Only) part n. F523001x part n. 19721123 LisaSafe Connection Kit USB pen drive Includes: USB connection cable + 1 roll of part n.
  • Page 88 Connection schemes Data communication Water treatment, supply and drain MULTIDEM ETHERNET Wi-Fi Key Osmo Mdist External LisaSafe USB drive S’Print water filtration system (*) (*) The water filtration system must be fitted with a backflow preventing device complying to IEC 61770 and to national and local regulations. Maximum pressure 8.6 bar, minimum flow 2.5 litres/min., maximum temperature 35 °C...
  • Page 89 Accessories and spare parts Standard chamber rack for 3 cassettes / containers (*) 190x43x300/375 * LISA 17: part n. F523008x 190x43x300/375 * LISA 22: part n. F523009x 190x43x300/375 * Standard chamber rack for 3 USA size cassettes (*) 205x35x300/375 * LISA 17: part n.
  • Page 90 Consumables Bacteriological filter - part n. W322400x Replace every 400 cycles Door seal - part n. F460504x Replace every 800 cycles Dust filter - part n. F364502x Replace every 400 cycles 400-800 cycle consumable kit - part n. X050326x This kit consists in a stock of consumables suitable to run 800 cycles. It includes: - 1 door seal;...
  • Page 91: Helix Test Documentation Form

    ANNEX 11. Helix test documentation form Use this page to create a logbook tracing the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. Date Cycle N. Operator Released Signature Chemical indicator YES NO YES NO YES NO YES NO YES NO YES NO...
  • Page 92 YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO YES NO...
  • Page 93 Authorized W&H service partners Find your nearest W&H service partner at http://wh.com Simply go to the menu option »Service« for full details. Alternatively please contact: W&H UK LIMITED, 6 Stroud Wood Business Centre, Park Street, St.Albans, Herts AL2 2NJ t +44 1727 874990 f +44 1727 874628 technical.uk@wh.com A-DEC AUSTRALIA CO.INC., Unit 8, 5-9 Ricketty Street, Mascot NSW 2020,...
  • Page 94 Manufacturer VA131 ENG - Rev. 4 W&H Sterilization S.r.l. Italy, I-24060 Brusaporto (Bg), via Bolgara, 2 t +39/035/66 63 000 f +39/035/50 96 988 Subject to alterations wh.com 29.11.2017...