Sling Options: Polyester Standing or Transfer Sling Attaching the Power Base Actuator 13 - 14 Weight Capacity: 350 I bs Weight without Sling: P460 - 109 lbs, P465 -121 lbs Attaching the Control Unit on Mounting Bracket Battery: 24V DC (RCHBL) 17 - 18...
PRODUCT DIAGRAM PRODUCT DIAGRAM SEMI ELECTRIC PATIENT LIFTER P460 SEMI ELECTRIC PATIENT LIFTER P465 A. Lift Arm B. Attachment (Sling) Cradle A. Lift Arm B. Attachment (Sling) Cradle C. Up-Down Actuator D. Knee Pad C. Up-Down Actuator D. Knee Pad E.
TUFFCARE patient lifters. can install a properly grounded outlet. Slings and accessories designed by other manufacturers are not to be utilized as a component of Tuffcare’s patient lifter system. 5. The Rhino stand up lifter is intended to be used for patient within weight limit indicated for the lifter.
Page 5
350lbs. transfer) from around the patient. ASSEMBLING THE LIFTER TUFFCARE DOES NOT suggest locking the rear casters DO NOT over-tighten mounting hardware. It will cause of the patient lifter when lifting a patient. When lift arm damage to the mounting brackets.
The lifter comes with pre-assembled main frame for easy assembly WARNING Use only Tuffcare parts to complete this stand up patient lifter. Use team Lift! This is a heavy product, asssistance is needed to preform the assembly. ASSEMBLING THE MAST ASSEMBLY TO THE BASE 3.
Page 7
ATTACH THE KNEE PAD TO THE MAIN FRAME WARNING Use only Tuffcare parts to complete this stand up patient lifter. Use team Lift! This is a heavy product, asssistance 1. Insert the knee pad is needed to preform the assembly.
Page 8
ATTACH THE POWER BASE ACTUATOR TO THE MAST ASSEMBLY (FOR P465 ONLY) 4. Place the small flat washer 1. Locate the base actuator at onto the bolt. the space between the two rear casters. 2. Turn counter clockwise to 5. Follow by the threading the remove the knob and the 1st knob onto the bolt.
ATTACHING DRIVE UNIT ATTACHING THE CONTROL UNIT ON THE MOUNTING ATTACHING DRIVE UNIT BRACKET 1. Temporarily remove the controller setting screw from the mounting plate. 2. Align the controller to the mounting plate at about a 5 degree angle. 3. Make sure the 2 grooves at the lower back of the controller WARNING slide are positioned over the 2 extrusions at the bottom of the Make sure there is an audible click when plugging battery...
2. Set the battery pack on top of the controller at about a 5 degree angle. 4. Place the controller mounting screw back into the same place and tighten. 3. While holding onto the lever latch inside the handle, move the battery pack parallel to the controller.
PREPARING THE PATIENT LIFT FOR USE CHARGING THE BATTERY CHARGING THE BATTERY 1. Plug the hand control, leg actuator (For P465 Power Base Lifters Only) and up-down actuator into the receptacles on NOTE: It is suggested to recharge the battery daily to prolong the bottom of the control unit.
“Absorption Charge”, leave charging for 4 to 6 hours. Patient Lift When the battery becomes fully charged, the battery meter indicator bars turn off. Unplug the charging cord and the lifter P460 P465 will be ready for use. - 21 -...
Page 13
CLOSING/OPENING THE LEGS ADJUSTING CALF PAD POSITION To Open and Close the Legs on P460 Step down on left side of “Leg Adjusting Stepper” to close the legs Step down on right side of “Leg Adjusting Stepper” to open the leg.
ACTIVATING A MECHANICAL EMERGENCY RELEASE LIFTING THE PATIENT LIFTING THE PATIENT Emergency Release To activate the emergency release, pull-down the Red Knob located on the top-end of Up-Down actuator drive shift. WARNING 1. DO NOT exceed the maximum weight limit of 450/650 lbs 2.DO NOT attempt to transfer until the user is assessed by a qualified professional.
Page 15
1. Open the leg section of the floor lift to obtain stability. FOR P460: Step down on right side of “Leg Adjusting Stepper” to open the legs to the maximum open position. Step down on right to OPEN 3.
5’10” Otherwise, injury or damage may occur. The legs of the Tuffcare’s Rhino Stand-Up Lift’s lift arm has three attachment knobs per patient lifter MUST be in the maximum open position for side. Use the above chart as a guideline to determine best attachment point.
Attach Standing Sling Before lifting the patient, make sure that the bottom edge of the standing sling is positioned on the lower back of the patient and the patient's arms are outside the standing sling. Check that the sling is properly connected Determine the best attachment point for the standing sling on each lift to the attachment points prior to lifting a patient.
Attach Buttock Strap Attach Transport Standing Sling Make sure the color coded strap sits flush against the attachment knob. Determine the best attachment point for the standing sling on each lift Loop the hanging strap on both end of the buttock strap on to the transport arm.
LIFT SLING TYPE REFERENCE TRANSPORT STANDING SLING is designed to be used only with a Tuffcare Rhino stand up lift slings are made with a three layer stand up sling. It will accommodate patients who have weight bearing soft but sturdy polyester fabric which provides comfort and capacity along with head and neck control and are able to bend at the hip, protects patients who may have sensitive skin.
(1) year. The warranty period commences on the original manufacture ® date. If within such warranty period any such product proven to Tuffcare DETECTING WEAR AND DAMAGE satisfaction to be defective, such product will be repaired or replaced at ®...