EMC Information
EMC Information
Important information regarding Electromagnetic Compatibility
(EMC)
This device manufactured by MD Biomedical, Inc. conforms to
IEC60601-1-2 Electromagnetic Compatibility (EMC) standard.
Nevertheless, special precautions need to be observed:
*The use of accessories and cables other than those specified or
provided by MD Biomedical, Inc. could result in increased
electromagnetic emission or decreased electromagnetic immunity of
the device and result in improper operation.
*During measurement, the use of the device adjacent to or stacked
with other device should be avoided because it could result in
improper operation. In case such use is necessary, the device and
other device should be observed to verify that they are operating
normally.
*During measurement, Portable RF communications device (including
peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the device,
including cables specified by MD Biomedical, Inc Otherwise,
degradation of the performance of the device could result.
*Refer to further guidance below regarding the EMC environment in
which the device should be used.