Calibration Procedure - Endress+Hauser nivotester FTC 422 Installation And Operating Instructions Manual

Hide thumbs Also See for nivotester FTC 422:
Table of Contents

Advertisement

Available languages

Available languages

Calibration
Switch position
Range
I
II
III
Setting
Position
1
2
3

Calibration procedure

This calibration requires that
• the vessel be filled to the upper switch point,
• the vessel be emptied to the lower switch point.
Step Remarks
1
Fill the vessel to the upper switch point.
2
Select Position 1 at hook switch 2.
3
Using Adjuster 1 and hook switch 1, find the operating
range as described in »Range seeking«, see p 28.
4
Depending on the mode chosen, calibrate for minimum or
maximum fail-safe mode - see p 28.
5
Calibrate at the lower switch point
Step Remarks
1
Empty the vessel to the lower switch point.
2
Select Position 3 at hook switch 2.
3
Using Adjuster 2 and hook switch 3, find the operating
range as described in »Range seeking«, see p 28.
4
Depending on the mode chosen, calibrate for minimum or
maximum fail-safe mode - see p 28.
5
Select Position 2 at hook switch 2 - the Nivotester 422 is
ready for operation.
- Check that the system functions correctly.
Input voltage
Capacitance
3...6.6 V
10...120 pF
6...8.8 V
110...350 pF
8...12 V
300...1200 pF
Remarks
Adjust upper switch point
Operating position
Adjust lower switch point
Table 2
Capacitance value as a function of
hook switch 1 + 3 position
Table 3
Function of hook switch 2
Calibration at upper switch
point
The calibration can also be
started at the lower switch point
Calibration at lower switch
point
27