Product Set-Up:

The incubator is designed to be installed on the bench-top. Before installing, always ensure that the
surface is clean, level and capable of withstanding the weight of the incubator. Once in place, plug
the electrical cord into a properly grounded outlet of the correct voltage. Once plugged in, press the
power button on the left side of the instrument.
III. Product Operation:
To select the desired temperature, press the "SET" key, "SP" is displayed the temperature can be set
using the Up, Down and Shift keys. Once the desired temperature value has been set, press the
"SET" key again to select the time value, "S
" is displayed. (When set to "0000" the time function
is disabled and the incubator will run continuously.) Once the desired time value has been
programmed, press the "SET" key again to store the values and begin the incubation. The incubator
will begin to heat to the desired temperature. Once
the temperature has been reached, the timer begins and the incubator will continue heating until the
time has expired. Upon completion, the incubator will beep to alert the user that the time is
completed and will display end. (To disable the beep press the down key. To clear the "End"
message, press and hold the down key for 3 seconds.)
When the time has been set to continuous, the display shows both the current chamber temperature
and the set chamber temperature. When a time other than continuous is chosen, the incubator
displays the measured chamber temperature as well as the remaining run time.
In addition to control over the temperature and run time, the fan speed can also be set to "High",
"Low" or "Off". With Dual Convection Technology, the SureTemp incubators provide accurate and
uniform results with the fan speed set to either of these three settings
WARNING: When loading samples, always avoid placing samples directly in front of fan vents. This
will minimize airflow and diminish temperature performance.
An internal outlet is included and provides the same electrical voltage as the outlet that the incubator
has been plugged into. In the event of condensation, always wipe the outlet down with a dry towel
prior to plugging in an electrical plug.
V. Care and Maintenance:
No routine maintenance is required other than to keep the instrument clean. When cleaning, use a
damp cloth to wipe down the internal or external parts of the incubator. Avoid the use of solvents as
they may attack the product housing. In addition, always ensure that the fan vents are clear of dirt or
any other obstructions.