Table 202 Guidance and Manufacturer's Declaration – Immunity
All Equipment and Systems
The Mighty Rest™ is intended for use in the electromagnetic environment specified below. The
customer, or user, of the Mighty Rest™ should ensure that it is used in such an environment.
Immunity Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips/Dropout
IEC 61000-4-11
Power Frequency
50-60Hz
Magnetic Field
IEC 61000-4-8
IEC 60601
Compliance
Test Level
Level
±6kV
Contact
±8kV Air
±2kV Mains
±1kV I/Os
±1kV
Differential
±2kV
Common
>95% Dip
for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip
for
5 seconds
3A/m
Electromagnetic Environment
Guidance
A
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
r/h should be at least 30%.
A
Mains power quality should be that of a
typical commercial or hospital
environment.
A
Mains power quality should be that of a
typical commercial or hospital
environment.
A
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the bed
requires continued operation during
power mains interruptions, it is
recommended that the bed be powered
from an uninterruptable power supply or
battery.
A
Power frequency magnetic fields should
be that of a typical commercial or
hospital environment.
33