Data Connection; Specifications - Ozone Solutions ES-600 User Manual

Ozone switch

Advertisement

Data Connection

Data from the ES-600 and connected SM-module can be received by a computer via RS-232
serial cable. Basic communication parameters are listed in the table to the right. For more
details
please
consult
the
www.ecosensors.com/support.
Serial data is updated from the SM-module approximately once per second. An example serial
data output stream is given below for 1 ppm ozone, 23 °C, and 33% relative humidity:
Ozone (ppb), Temperature (°C), and Relative Humidity (%):
1000, 23, 33
Analog Output Full Scale Selection
The full-scale range for the 0-2 V and 4-20 mA analog outputs may be
selected by adding or removing up to three push-on jumpers on header
J4, located on the back of the display circuit board. Access this area by
removing four screws from the display bezel.
Changing the analog output full-scale range will not change the ES-600
display, which will always read in the within range 00.00 to 99.99 ppm.
The actual measurement range of the ES-600 display is determined by
the attached SM-module. Refer to the specification sheet of the attached
SM-module to determine the measurement range.
The ES-600 is shipped from the factory with the Analog Output Full Scale
Range set to match the installed SM-module. For further details, refer to
the specification sheet of the attached SM-module.

Specifications

Enclosure:
Wall mounted polycarbonate enclosure resists water spray and splash (NEMA 4X).
H = 6.3"/160mm, W = 6.5"/166mm, D = 4.6"/118mm.
Size:
Wiring:
Conduit or cable gland knock-outs along bottom of enclosure.
Power in:
12-24 VDC.
Power out:
+5 or +12-24 VDC to sensor.
Data in:
RS-232 serial data from remote sensor module (SM-module).
Data out:
RS-232, 9600 baud, 8N1 format.
Analog out:
0-2 V and 4-20 mA, internal jumpers select full scale ppm range.
Control in:
Generator enable (contact closure).
Control out:
Relays: 12 A @ 30 VDC (resistive load), 12 A @ 250 VAC (resistive load).
Relay functions:
Ozone generator control, >100 ppb safety limit, failure to control ozone.
User controls (internal):
MAX set point, MIN set point, FAIL time, increase, decrease.
4-digit, 0.5" (12 mm), auto-ranging, backlight.
Numeric display:
LED indicators:
Sensor, above MAX, below MIN, >100 ppb, O
Ozone Solutions
sales@ozonesolutions.com
(888) 892‐0303
RS-232
Communication
Ozone Controller, Model ES-600
User Manual, Rev. 1.0
Manual,
available
at
J4 Jumper Positions
A
B
OFF
OFF
ON
OFF
OFF
ON
ON
ON
OFF
OFF
ON
OFF
OFF
ON
ON
ON
generator control, failure.
3
Serial Communication Parameters
Value
Parameter
Speed (baud)
9600
Data bits
8
Stop bits
1
Parity
None
Flow Control
None
Analog Output Full-
Scale Range (ppm)
C
OFF
0-2
OFF
0-5
OFF
0-10
OFF
0-20
ON
0-50
ON
0-100
ON
0-200
ON
0-500