LORD TECHNICAL NOTE
®
V-Link
-LXRS
Using Pressure Transducers
Overview
The LORD MicroStrain
transducers on channels 1, 2, 3 and 4 and the
support pressure transducers on channel 1. This technical note explains use of a typical pressure transducer with SG-
®
®
Link
-LXRS
and Node Commander
Pressure Transducer
For our example, we are using a
designed to give accurate level measurement in up to 1,200 feet of water, and is used for both general water
applications and corrosive applications.
Wiring
Refer to the Model 817
Start Guide
(frame 2C) to connect the sensor to the node.
Model 817 Pressure Transducer
Lead color
Red
+ Power
Black
+ Signal
Green
- Power
White
- Signal
Shield
GND
Step 1
Locate the specific calibration sheet provided by the pressure transducer
manufacturer and any other documents as necessary to determine the
following parameters:
o
Calibration Voltage
o
Full Scale Reading
o
Measurement Range
We find the Calibration Voltage embossed on the barrel of the
transducer (10 V).
We find the Full Scale Reading (86.073 mV) on the calibration
document, as shown in Figure 2.
We find the Measurement Range embossed on the barrel of the
transducer (20 feet).
1
®
and SG-Link
®
®
®
V-Link
-LXRS
Wireless 7 Channel Analog Input Sensor Node will support pressure
®
software; the V-Link
Viatran Model 817 pressure
Installation Data Manual
Lead use
Pin number
1
2
4
3
4
Table 1: Wiring
®
®
SG-Link
-LXRS
Wireless 2 Channel Analog Input Sensor Node will
®
®
-LXRS
mirrors the instruction.
transducer. The Model 817 is a submersible instrument
(page 4) and the SG-Link
SG-Link-LXRS
Pin nomenclature
VXC
S+
GND
S-
GND
®
-LXRS
®
®
-LXRS
Quick
Figure 1: Model 817
Pressure Transducers
Figure 2: Calibration Document
®

Advertisement

loading

Summary of Contents for Lord MicroStrain V-Link-LXRS

  • Page 1 -LXRS and SG-Link -LXRS Using Pressure Transducers Overview ® ® ® The LORD MicroStrain V-Link -LXRS Wireless 7 Channel Analog Input Sensor Node will support pressure ® ® transducers on channels 1, 2, 3 and 4 and the SG-Link -LXRS Wireless 2 Channel Analog Input Sensor Node will support pressure transducers on channel 1.
  • Page 2 Note: by varying the User Settable Gain and the Balance (high, mid, low), we would be able to make setups ® with higher and lower resolution, and differing ranges. Please speak with your LORD MicroStrain support engineer if further detail is needed.
  • Page 3 ® ® ® ® V-Link -LXRS and SG-Link -LXRS Using Pressure Transducers  Select Other in the Units drop-down; Other will now represent ‘inches’.  Click the Modify button.  Enter 0.06546 in the Slope textbox; this value is pulled from the orange Slope cell, calculated in our spreadsheet.
  • Page 4  The system is now configured and ready for use. Support ® LORD MicroStrain support engineers are always available to expand on this subject and support you in any way we can. Copyright © 2013 LORD Corporation LORD Corporation ®...

This manual is also suitable for:

Sg-link-lxrs