Chattanooga Hydrocollator SS-2 User Manual

Chattanooga Hydrocollator SS-2 User Manual

Mobile heating units
Hide thumbs Also See for Hydrocollator SS-2:

Advertisement

Hydrocollator
®
MOBILE HEATING UNITS
User Manual
120 - 220/240 Volt Units
M-4
SS-2
M-2
ISO 13485 CERTIFIED

Advertisement

Table of Contents
loading

Summary of Contents for Chattanooga Hydrocollator SS-2

  • Page 1 Hydrocollator ® MOBILE HEATING UNITS User Manual 120 - 220/240 Volt Units SS-2 ISO 13485 CERTIFIED...
  • Page 2: Table Of Contents

    Any use of editorial, pictorial or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by Chattanooga Group of Encore Medical, L.P. ISO 13485 CERTIFIED...
  • Page 3: Foreword

    CAUTION • Read, understand and practice the precautionary and operating instructions. Know the limitations associated with the Hydrocollator SS-2, M-2, M-4 Heating Units. Observe the cautionary and operational decals installed on the unit. • The thermostat is extremely sensitive and the slightest adjustment will alter the temperature several degrees.
  • Page 4 WARNING • Grounding the Hydrocollator SS-2, M-2, and M-4 Mobile Heating Unit: Make certain the unit is electrically grounded by plugging into a hospital grade electrical outlet with a ground terminal receptacle (U-ground outlet), 120 VAC only. Follow the National Electrical Code.
  • Page 5: Nomenclature

    The operation of the unit may be restored by depressing the red button after the water/element temperatures have been allowed to cool. If the fault condition persists, please contact Chattanooga Group or an authorized agent.
  • Page 6 NOMENCLATURE Hydrocollator ® Mobile Heating Units Diagram 1. Heating Element 9. Junction Box (J-Box) Cover 2. Drain 10. IEC connector 3. Overtemp Thermostat 11. Power Cord 4. Thermostat, Hydraulic 12. Drain Valve (Not Shown) 5. Bracket (Brkt. & Bulb Themo A/B Assembly) 13.
  • Page 7 NOMENCLATURE Hydrocollator ® Mobile Heating Units Diagram (continued) 17. Hot Pac 23. Side Handle 18. Neck Pac Insert 24. Thermostat, Hydraulic 19. Drain, Copper Tube 25. Thermostat Control Knob 20. Lid Hinge nuts 26. On/Off Switch 21. Lid Knob/Handle 27. Internal Wiring Harness 22.
  • Page 8: Technical Specifications

    TECHNICAL SPECIFICATIONS Hydrocollator ® Mobile Heating Units SS-2 Model Power Supply 110-120V~ 220-240V~ Rated Frequency 50/60Hz Power Consumption 1000 W Tank Capacity 49.32 Liters Temperature Range 160°F - 165°F (71°C - 74°C) Thermal Cut-out Temp 190°F ± 14°F (88°C ± 8°C) Accuracy +/- 10% Heating Up Time...
  • Page 9 TECHNICAL SPECIFICATIONS Hydrocollator ® Mobile Heating Units M-4 Model Power Supply 110-120V~ 220-240V~ Rated Frequency 50/60Hz Power Consumption 1500 W Tank Capacity 135.76 Liters Temperature Range 160°F - 165°F (71°C - 74°C) Thermal Cut-out Temp 190°F ± 14°F (88°C ± 8°C) Accuracy +/- 10% Heating Up Time...
  • Page 10: Setup

    SETUP Hydrocollator ® Mobile Heating Units Remove all of the contents inside the Hydrocollator and then replace the wire rack. Make certain the drain valve is closed before filling the tank with water. The tank should be approximately 3/4 full with packs removed. The tank may be filled from any standard type faucet.
  • Page 11: Troubleshooting

    TROUBLESHOOTING Hydrocollator ® Mobile Heating Units The following are solutions to minor performance problems. Only qualified personnel should service equipment problems. PROBLEM POSSIBLE CAUSE POSSIBLE SOLUTION 1. Breaker for electric circuit off. 1. Turn breaker ON. Unit Plugged In, 2. Thermostat not functioning. 2.
  • Page 12: Replacement Parts

    REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model M-2 – Standard Cat. No. – 2402 w/packs Key* Part Description Part No. Element, Heating 120 Volt, 1000 Watt 10630 Hydraulic Thermostat 10631 Brkt. & Bulb Thermostat A/B Assembly (2 each)** 21023 22297 Caster 3 in 21259...
  • Page 13 REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model SS-2 – Standard Cat. No. – 2302 w/packs Key* Part Description Part No. Element, Heating 120 Volt, 1000 Watt 10630 Hydraulic Thermostat 10631 Brkt. & Bulb Thermostat A/B Assembly (2 each)** 21023 Lid SS-2 22219 Caster 3 inch (4)
  • Page 14 REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model M-4 – Standard Cat. No. – 2502 w/packs Key* Part Description Part No. Heating Element M-4, 120 Volt, 1000 Watt 20570 Hydraulic Thermostat 10631 Brkt. & Bulb Thermostat A/B Assembly (2 each)** 21023 22348 Caster 3 in...
  • Page 15 REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model SS-2 220-240 VOLT – (Low Water) Cat. No. – 2307 w/packs Cat. No. – 2309 w/o packs Key* Part Description Part No. Element, Heating, 240V~ CE 22245 Thermostat, Hydraulic 10631 Lamp, Pilot, Green 22146 Cord Set (Country Specific) Switch...
  • Page 16 REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model M-2 220-240 VOLT – (Low Water) Cat. No. – 2407 w/packs Cat. No. – 2409 w/o packs Key* Part Description Part No. Element, Heating, 220 Volt~, 1000 Watt 22245 Thermostat, Hydraulic 10631 Lamp, Pilot, Green 22146 Cord Set (Country Specific)
  • Page 17 REPLACEMENT PARTS Hydrocollator ® Mobile Heating Units Model M-4 220-240 VOLT – (Low Water) Cat. No. – 2506 w/packs Cat. No. – 2508 w/o packs Key* Part Description Part No. Element, Heating, 240 Volt~, 1500 Watt 22246 Thermostat, Hydraulic 10631 Lamp, Green, 1607-5107, 220-240 Volt~ 22146 Cord Set (Country Specific)
  • Page 18: Maintenance

    MAINTENANCE Hydrocollator ® Mobile Heating Units Care and Cleaning The Hydrocollator is equipped with an immersion type heating element and a hydraulic capillary type thermostat which evenly maintains the HotPac temperature in the water and provides a ready supply of heated packs.
  • Page 19 The pack should be replaced when it reaches this state. Original replacement Hydrocollator HotPacs may be ordered from an authorized Chattanooga Group dealer. NOTE: If HotPacs are to be stored for extended periods, they may be placed - while wet - in plastic bags and...
  • Page 20: Warranty

    WARRANTY Hydrocollator ® Mobile Heating Units Chattanooga Group ("Company") warrants that the Hydrocollator Models SS-2, M-2, and M-4 Heating Units ("Products") are free of defects in material and workmanship. This warranty shall remain in effect for one year from the date of original consumer purchase. If these Products fail to function during the...
  • Page 21 ISO 13485 CERTIFIED 4717 Adams Road P.O. Box 489 Hixson, TN 37343 1.423.870.2281 U.S.A. 1.800.361.6661 Canada 1.423.870.7200 Outside U.S.A. 1.423.870.2046 chattgroup.com © 2006 Encore Medical, L.P. 22151B...

This manual is also suitable for:

Hydrocollator m-2Hydrocollator m-4

Table of Contents