Page 1
Dear Client Thank you for Purchasing our HTJL-H High Precision SF6 Gas Leak Detector. Please read the manual in detail prior to first use, which will help you use the equipment skillfully. Our aim is to improve and perfect the...
Page 2
SERIOUS COMMITMENT All products of our company carry one year limited warranty from the date of shipment. If any such product proves defective during this warranty period we will maintain it for free. Meanwhile we implement lifetime service. Except otherwise agreed by contract.
Page 3
shock, the grounding conductor must be connected to the ground. Make sure the product has been grounded correctly before connecting with the input/output port. Pay Attention to the Ratings of All Terminals To prevent the fire hazard or electric shock, please be care of all ratings and labels/marks of this product.
Page 4
-Security Terms Warning: indicates that death or severe personal injury may result if proper precautions are not taken Caution: indicates that property damage may result if proper precautions are not taken.
Contents Warning:.................... 6 I. Product introduction ..............6 II. Technical data................7 III. Product structure..............7 IV. Operation procedure and method ........8 V. Adjustment method..............9 VI. Announcements..............11...
I. Product introduction HTJL-H SF6 Gas leak Detector’s working principle is that when the concentration of sulfur hexafluoride in the air changes its degree of ionization under an electromagnetic field of high frequency varies at the same time. The concentration degree of sulfur hexafluoride can be displayed by a detection of the degree of ionization.
II. Technical data Range of measurement(SF ): 0.01ppm~500ppm(Volume ratio) Response time: not more than 3 seconds Indication pattern: crystal digital display harmony, light signal alarm Setting range of alarming: 0.01ppm~500ppm(Volume ratio) Pumping speed of vacuum pump 0.5 litre/second Length of the vacuum tube: 4 m Continual working time:...
1 detection switch 2 power switch 3 power outlet 4 ac fuse (0.5A) 5 dc fuse (5A) 6 the host Seven of the needle valve 8 handheld probe 9 LCD 10 buttons 11 ionization chamber observation window 12 gas electric tube IV.
inspection. If there is a gas leak, the reading on the LCD screen will increase to show the specific concentration of volume. 6, press “√” key to enter setup interface can change the emergency alarm value (using “◄”、 “►”The key changes).When the volume ratio is greater than the alarm point, the device will emit a beeping alarm.
Page 10
( 3 ) Operation: Extract 1 millilitre of pure sulfur hexafluoride with a one-litre cylinder from the steel bottle. Inject it into 100-millilitre cylinder and dilute it with air outdoors. The concentration degree is 10000ppm. Again, extract 5 millilitre 10000ppm sulfur hexafluoride with a twenty-milliliter from a 100-millilitre cylinder and inject it into another 100-millilitre cylinder.
PPM sample gas. Press OK and then to go to the next point. In this way the demarcation of all the seven points is finished ( six sample gas value and one zero point value ). After all the sample gas’s demarcation is finished, the user can restart the leakage detector for the next detection.
Need help?
Do you have a question about the HTJL-H and is the answer not in the manual?
Questions and answers