is pushed to enter the next character. After all characters of user's name are entered by this
way, push enter key (F4) to enter the mail server of the e-mail address. After mail server is
entered in the same way, push enter key (F4) to pass the "password" submenu.
Enter the password and push enter key (F4).
The port provided by internet server is entered on the part of "Port". After port is entered,
enter key (F4) is pushed to return the main menu. Tab key (F3) is used for changes from one
character of address to another.
6.1.6.

SET TEMPERATURE

In case of active audible alarm, firstly, push the alarm mute key (F2) to access the
working settings page.
6.1.7.

GRAPH WORKING SCREEN

When the graph button (F3) is pushed in the operation mode, the graph page comes to the
screen. Daily operation can be followed on this graph. If you want to analyze old operations,
visit the memory page (See Section 6.1.3).
6.1.8.

TEMPERATURE ALARM RANGE

Temperature alarm range can be set between 5°C to 10°C. In the factory, temperature alarm
range is set to 10°C. Follow the steps below to change this value.
"Login" is selected on the e-mail page by the
value increase (F1) and decrease (F2) keys. The
page on the left appears when enter key (F4) is
pushed. Each character of e-mail addresses is
written one by one. The first character is entered
by pushing value increase (F1) and decrease keys
(F2) and tab key (F3) is pushed to enter the next
character. The second character is entered by
pushing value increase (F1) and decrease keys
(F2) and tab key (F3)
When the setting key (F2) is pushed in the
operation mode, password page appears. This
password is the user's password. The password is
0000 for the first usage. "Working Settings" page
on the left appears. The set temperature can be
adjusted between -20⁰C and -45⁰C by pushing the
value increase (F1) and decrease (F2) keys. After
set temperature value is entered, push the enter
key (F4) to save the value. If you want to exit this
page without any change, push the backspace key
(F3).