Metrohm 872 Manual

Metrohm 872 Manual

Extension module, suppression - mcs
Hide thumbs Also See for 872:
Table of Contents

Advertisement

Quick Links

872 Extension Module
2.872.0120 – Suppression – MCS
Manual
8.872.8007EN

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 872 and is the answer not in the manual?

Questions and answers

Summary of Contents for Metrohm 872

  • Page 1 872 Extension Module 2.872.0120 – Suppression – MCS Manual 8.872.8007EN...
  • Page 3 Metrohm AG CH-9101 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com 872 Extension Module 2.872.0120 – Suppression – MCS Manual 8.872.8007EN 10.2010 zst...
  • Page 4 Teachware Metrohm AG CH-9101 Herisau teachware@metrohm.com This documentation is protected by copyright. All rights reserved. Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
  • Page 5: Table Of Contents

    Inserting the rotors ..............21 3.6.2 Connecting the suppressor ............ 24 Metrohm CO Suppressor (MCS) ........26 3.7.1 General information on the MCS ........... 26 3.7.2 Connecting MCS ..............26 3.7.3 Installing the adsorption cartridges ........28 ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 6 Replacing the CO adsorption cartridge ......... 41 5.5.2 Regenerating the H O adsorption cartridge ......41 Quality Management and validation with Metrohm ..42 6 Troubleshooting Problems and their solutions ..........43 7 Technical specifications Reference conditions ............45 Instrument ................45 Ambient conditions ............
  • Page 7 ■■■■■■■■■■■■■■■■■■■■■■ Table of contents Warranty (guarantee) ............50 9 Accessories Scope of delivery ..............51 Optional accessories ............55 Index ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 8 ■■■■■■■■■■■■■■■■■■■■■■ Table of figures Table of figures Figure 1 Front 872 Extension Module Suppression – MCS ....... 6 Figure 2 Rear 872 Extension Module Suppression – MCS ........ 7 Figure 3 Proposed setup ................. 8 Figure 4 Peristaltic pump ................16 Figure 5 Installing the pump tubing ..............
  • Page 9: Introduction

    Instrument description Existing 850 Professional IC instruments can be expanded to include addi- tional functions by means of 872 Extension Modules. Every 850 Professio- nal IC instrument can be supplemented with up to 3 extension modules. The 872 Extension ModuleSuppression – MCS permits to add an additional suppression module to a Professional IC system.
  • Page 10: Intended Use

    Intended use This instrument is suitable for processing chemicals and flammable sam- ples. The usage of the 872 Extension Module therefore requires that the user has basic knowledge and experience in the handling of toxic and caustic substances. Knowledge with respect to the application of the fire prevention measures prescribed for laboratories is also mandatory.
  • Page 11: Safety Instructions

    The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010. Warning Only personnel qualified by Metrohm are authorized to carry out service work on electronic components. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 12: Tubing And Capillary Connections

    Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections. Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are manda- tory. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 13: Flammable Solvents And Chemicals

    The correct disposal of your old equipment will help to prevent negative effects on the environment and public health. More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 14: Overview Of The Instrument

    ■■■■■■■■■■■■■■■■■■■■■■ 2.1 Front 2 Overview of the instrument Front Figure 1 Front 872 Extension Module Suppression – MCS Peristaltic pump (see Chapter 3.5, page 15) (see Chapter 3.7, page 26) Standby indicator Suppressor drive (see Chapter 3.6, page 21) ■■■■■■■■...
  • Page 15: Rear

    ■■■■■■■■■■■■■■■■■■■■■■ 2 Overview of the instrument Rear Figure 2 Rear 872 Extension Module Suppression – MCS Waste air opening Transport locking screws For extracting the air from the vacuum For securing the vacuum pump when trans- chamber. Labeled with Exhaust.
  • Page 16: Figure 3 Proposed Setup

    If this is not possible, then the extension modules that are located at greater distances from one another must be connected with one another by means of the longer 6.2156.070 connection cable. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 17: Installation

    ■■■■■■■■■■■■■■■■■■■■■■ 3 Installation 3 Installation Installation overview The 872 Extension Module Suppression – MCS can be installed carrying out the following easy steps: 1 Preparing the instrument Set up the instrument (see Chapter 3.2, page 10) ■ Mount base tray / bottle holder (see Chapter 3.3, page 10) ■...
  • Page 18: Setting Up The Instrument

    Bottle holder and base tray must be removed or mounted every time one of the following instruments is mounted on or under an 850 Professional IC instrument: One or more 872 Extension Module. ■ or another instrument of the same support surface.
  • Page 19 There are no loose parts in the instrument. ■ The instrument is lying on it side, and the bottom surface is visible. ■ To mount the base tray, you need the 3 mm hexagon key (6.2621.100). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 20: Removing / Mounting The Bottle Holder

    The bottle holder is cleared. ■ Drainage tubing is disconnected from the drainage tubing connection ■ of the bottle holder. To remove the bottle holder, you need a 3 mm hexagon key (6.2621.100). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 21 Mounting the bottle holder Before you can mount the bottle holder, the following preconditions must be met: The instrument is switched off. ■ To mount the bottle holder, you need a 6.2621.100 hexagon key 3 mm. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 22 (see also the manual for the IC instrument). 5 If one of the instruments in the stack is equipped with a leak sensor connection socket, connect the leak sensor (see manual of the IC instrument). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 23: Transport Locking Screws

    The pump tubing is clamped between the rollers (4-3) and the tub- ing cartridge (4-5). During operation, the peristaltic pump drive rotates the roller hub (4-2), so that the rollers (4-3) push the liquid forward in the pump tubing. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 24: Figure 4 Peristaltic Pump

    ■■■■■■■■■■■■■■■■■■■■■■ 3.5 Peristaltic pump Figure 4 Peristaltic pump Knurled screw in the mounting pin Roller hub Rollers Cartridge holder Tubing cartridges 6.2755.000 Contact pressure lever Snap-action lever ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 25: Installing The Peristaltic Pump

    Release the tubing cartridge from the cartridge holder by pressing the snap-action lever and unhooking from the mounting pins (4-1). 2 Connecting the aspiration side Place a 6.2744.034 tubing olive (5-2) on the aspiration side of the pump tubing. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 26: Figure 6 Install Pump Tubing Connection With Filter

    Place the tubing olive with filter holder (6-3) onto the pump tub- ■ ing. Screw the union nut (6-1) onto the tubing olive (6-3). ■ Case B: 6.2744.160 pump tubing connection without filter: ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 27: Figure 7 Install Pump Tubing Connection Without Filter

    Screw the respective capillaries tightly to the two tubing olives ■ with PEEK pressure screws (5-1). Table 1 Pump tubings and suitable adapters Pump tubing Adapter 6.1826.020 (blue/blue) 6.1826.310 (orange/green) 6.1826.320 (orange/yellow) 6.1826.330 (orange/white) 6.1826.340 (black/black) ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 28 Note Pump tubings are consumable material. The service life of the pump tubings depends on the contact pressure amongst other factors. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 29: Suppressor Drive

    ■■■■■■■■■■■■■■■■■■■■■■ 3 Installation Suppressor drive The suppressor housing of the 872 Extension Module Suppression – MCS can hold various rotors. Large rotors like the SPM rotors A (6.2835.000) and the MSM-HC rotor A (6.2842.000) can be inserted directly. Small rotors like the MSM rotor A (6.2832.000) and the MSM-LC rotor A (6.2844.000) must first be fitted into the adapter (6.2842.010) which can...
  • Page 30 If the rotor is correctly inserted, its sealing surface will be approx. 4 mm within the housing. If this is not the case, the rotor must be moved into the right posi- tion from below using a sharp object (e.g. screwdriver). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 31 2 Inserting the adapter Insert the adapter as you would insert a large rotor (see "Inserting large rotors", page 22). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 32: Connecting The Suppressor

    The rinsing and regeneration solutions are pumped with a peristaltic pump (see Chapter 3.5, page 15). Connect the PTFE capillaries firmly mounted on the connecting piece to the other components of the IC system as follows: ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 33 5 Connecting the regeneration solution inlet capillary With a short PEEK pressure screw (6.2744.070), fasten the end of ■ the inlet capillary labeled with regenerant to the pump tubing connection of the pump tubing which carries the regeneration solution. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 34: Metrohm Co 2 Suppressor (Mcs)

    CO to diffuse out of the eluent flow. 3.7.2 Connecting MCS The MCS is connected between the MSM and detector. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 35: Figure 10 Mcs - Connection

    Connect the detector inlet capillary to the MCS output (10-2) using a long PEEK pressure screw (6.2744.090) (10-4). Caution If the MCS is not used, the input and output must be sealed with 6.2744.220 stoppers. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 36: Installing The Adsorption Cartridges

    3-layer filled, blue-brown-gray. Adapter (6.1808.190) PVC tubing For connecting the H O adsorption cartridge For connecting the H O adsorption cartridge and CO adsorption cartridge. and CO adsorption cartridge. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 37 O adsorption cartridge into the two upper clips (11-3) ■ of the adsorption cartridge holder (11-8). 6 Placing the adsorption cartridge holder in the instrument Place the adsorption cartridge holder with cartridges into the ■ detector chamber of the instrument ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 38: Start-Up

    A rotor is correctly inserted. ■ The MCS is connected. ■ The 872 Extension ModuleSuppression – MCS is connected to the 850 ■ Professional IC instrument. 1 Start MagIC Net. 2 Connect the 850 Professional IC instrument to the PC and switch on.
  • Page 39: Operation And Maintenance

    Metrohm. If working frequently with caustic and corrosive chemicals, a shorter mainte- nance interval is recommended. The Metrohm service department offers every form of technical advice for maintenance and service of all Metrohm instruments. 5.1.3...
  • Page 40: Shutting Down

    Therefore fully lift the tubing cartridges by loosening the snap- action lever (5-10) on the right-hand side if the peristaltic pump is to be turned off for a longer period. Once set, the contact pressure remains unaffected. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 41: Maintenance

    Pump tubing for bromate (orange/green), 3- determination using the stopper triiodide method. 6.1826.320 Pump tubing LFL PVC (Tygon) 0.48 mm For suppressor solutions, (orange/yellow), 3- acceptor solutions for stopper inline dialysis and for inline ultrafiltration. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 42: Figure 12 Pump Tubing Connection - Changing The Filter

    6.2621.000 adjustable wrenches. 2 Replacing the filter Remove the old filter (12-2) with tweezers. ■ Place the new filter (12-2) flat in the tubing olive (12-1) with ■ tweezers. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 43: Suppressor/Spm

    Caution In the case of reduced capacity or high backpressure, the suppressor must be regenerated (see Chapter 5.4.3.2, page 36), cleaned (see Chapter 5.4.3.3, page 37) or replaced (see Chapter 5.4.3.4, page 39). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 44: Maintenance

    / acetone≥ 20% 0.2 mol/L H Caution The pump tubings made of PVC must not be used for rinsing with solu- tions containing organic solvents. We recommend to use the high pressure pump for regeneration. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 45 Blockage of the suppressor which cannot be eliminated (solutions can ■ no longer be pumped through the suppressor). Jamming of the suppressor which cannot be eliminated (suppressor ■ can no longer be switched over). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 46 (13-5). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 47 1 Disconnecting the suppressor from the IC system Switch off the instrument. ■ Disconnect all capillaries of the suppressor from the IC system. ■ 2 Dismantling the suppressor Unscrew union nut (13-1) from the housing (13-4). ■ ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 48 7 Connecting and conditioning the suppressor Reconnect all capillaries of the suppressor to the IC system. ■ Before switching the suppressor over for the first time, rinse the ■ three suppressor units with solution for 5 minutes. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 49: Metrohm Co Suppressor (Mcs)

    1 Allow material to dry loose (not in cartridge) at 140 °C overnight and refill. Or dispose of the old material, and fill with new material. 2 Cover the packed material with cotton. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 50: Quality Management And Validation With Metrohm

    Metrohm» available from your local Metrohm agent. Validation Please contact your local Metrohm agent for support in validating instru- ments and software. Here you can also obtain validation documentation to provide help for carrying out the Installation Qualification (IQ) and the Operational Qualification (OQ).
  • Page 51: Troubleshooting

    Correctly set contact pressure (see "Set the ing of regeneration pressure too weak. contact pressure", page 20). or rinsing solution Peristaltic pump – filter Replace the filter (see "Replacing the filter", blocked . page 34). ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 52 Regenerate the suppressor (see Chapter baseline capacity. 5.4.3.2, page 36). Precision problems - Check connectors. If they are ok: MCS – vacuum too low. ■ significant scatter- Contact the Metrohm Service. ■ ing of the measured values ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 53: Technical Specifications

    Painted polyurethane hard foam without CFCs, fire class V0 Material Intelligent com- MagIC Net ponents Ambient conditions Operation Ambient tem- +5…+45 °C perature Humidity 20…80 % relative humidity Storage Ambient tem- –20…+70 °C perature Transport Ambient tem- –40…+70 °C perature ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 54: Housing

    Material fluoropolymer Resistance to sol- No restriction (apart from PFC) vents Vacuum Working area Microprocessor-controlled / stabilized Build-up time < 30 s after start Capillary volume 400 µL Recommended 0.1…1.0 mL flow range ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 55: Interfaces

    Emission EN/IEC 61326-1 ■ EN/IEC 61000-6-3 ■ EN 55011 / CISPR 11 ■ Immunity EN/IEC 61326-1 ■ EN/IEC 61000-6-1 ■ EN/IEC 61000-4-2 ■ EN/IEC 61000-4-3 ■ 7.11 Weight 1.872.0120 6.7 kg (without accessories) ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 56: Conformity And Warranty

    Name of commodity 872 Extension Module The 872 Extension Module is an expansion tool for upgrading all 850 Professional IC instruments. This instrument has been built and has undergone final type testing...
  • Page 57: Quality Management Principles

    Components All components used in the Metrohm instruments have to satisfy the qual- ity standards that are defined and implemented for our products. Suppli- ers of components are audited by Metrohm as the need arises.
  • Page 58 Warranty (guarantee) Metrohm guarantees that the deliveries and services it provides are free from material, design or manufacturing errors. The warranty period is 36 months from the day of delivery; for day and night operation it is 18 months.
  • Page 59 ■■■■■■■■■■■■■■■■■■■■■■ 9 Accessories 9 Accessories Note Subject to change without notice. Scope of delivery 2.872.0120 872 Extension Module Suppression – MCS Qty. Order no. Description 1.872.0120 872 Extension Module Suppression – MCS 6.1602.150 Bottle attachment / GL 45 - 3 x UNF 10/32 For connecting capillary tubing 1/16 in.
  • Page 60 Cable for the connection of an extension module to a Professional IC instrument Length (m): 6.2744.034 Coupling olive/UNF 10/32 2x Connection of pressure screw and pump tubing. 2 pieces. For IC instruments with peristaltic pumps ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 61 PEEK capillaries Material: PEEK Length (mm): 6.2744.090 Pressure screw long Long version. With UNF 10/32 connection. 2 pieces. For the connec- tion of PEEK capillaries. (MCS and sample degasser) Material: PEEK ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 62 Connection piece to MSM-HC and SPM module With inlets and outlets. For inline sample preparation 6.2837.000 adsorption cartridge Adsorption cartridge for purifying the air. 6.2837.010 O adsorption cartridge To CO Suppressor. Water adsorption cartridge for the aspirated air. ■■■■■■■■ 872 Extension Module Suppression – MCS...
  • Page 63 Suppressor rotor for all IC instruments with MSM Length (mm): 6.2842.000 MSM-HC Rotor A Suppressor rotor for all IC instruments with MSM-HC (Metrohm Suppressor Module with high capacity) 6.2842.010 Adapter to MSM The adapter allows the use of the standard MSM rotor in the MSM-HC unit.
  • Page 64 Validation ......... 42 Regeneration ...... 41 Installing ......17 Heavy metals Service life ......32 Contamination suppressor . . 36 Warranty ........50 Housing ........46 Humidity ........45 Quality Management ....42 ■■■■■■■■ 872 Extension Module Suppression – MCS...