SCI 6000 Series Quick Start Manual

6-in-1 multi-function tester

Advertisement

Quick Links

6000 Series
6-in-1 MULTI-FUNCTION TESTER
M odels: 6330
SAFETY CHECKLIST
Survey the test station. Make sure it is safe & orderly.
Always keep unqualified/unauthorized personnel away from the test area.
Familiarize yourself with safety protocols in the event of a problem.
Exercise caution and never touch products or connections during a test.
Train operators. Connect the return lead first and never touch clips directly.
You should always know when a test is being performed.
WARNING:
ELECTRICAL SAFETY TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND
CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL
INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING
AND USING A TEST INSTRUMENT.
QUICK START GUIDE
THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH
HIPOT.COM

Advertisement

Table of Contents
loading

Summary of Contents for SCI 6000 Series

  • Page 1 QUICK START GUIDE 6000 Series 6-in-1 MULTI-FUNCTION TESTER M odels: 6330 SAFETY CHECKLIST Survey the test station. Make sure it is safe & orderly. Always keep unqualified/unauthorized personnel away from the test area. Familiarize yourself with safety protocols in the event of a problem.
  • Page 2: Front Panel Controls

    Front Panel Controls 12 13 Reset Button: Red momentary contact switch used to reset the instrument in case of a failure. Also serves as an abort signal to stop any test in progress. Test Button: Starts a test. LCD Display: 20 x 2 character display. Lock Indicator: This LED indicates that the lock feature has been enabled.
  • Page 3: Back Panel Controls

    Back Panel Controls 3 4 5 External Measuring Device: Provides an access port for a custom measuring device. The port may be activated during Line Leakage testing from the instrument’s menu system. The operator must supply the appropriate measuring device as none is provided. Power Control Fuse Receptacle: To change the fuse, unplug the power (mains) cord and turn the fuse receptacle counterclockwise.
  • Page 4 This is required in order to run a test. Turn the power switch ON. Upon power up, you will see the SCI company name, model number, and the current software version briefly appear on the display. After this, you will see the default test parameters.
  • Page 5: Selecting A Memory Location

    SET SYS Edit System Parameters Configure the instrument system parameters to your preferences. The instrument system parameters are global and will affect all tests that are performed regardless of memory location and memory step. SET SYSTEM PARAMETERS SET SYSTEM PARAMETERS Use the up or down arrow Press the EXIT key to access Press the SET key to toggle...
  • Page 6 Edit Test Parameters Default test type parameters are shown below. Press the SET key a third Press the SET key from the Press the SET key again to EDITING TEST PARAMETERS EDITING TEST PARAMETERS time until the test types Perform Test screen to confirm the memory step EDITING TEST PARAMETERS appear.
  • Page 7 Edit Test Parameters Default test test parameters for each test type are as follows: AC Hipot Mode DC Hipot Mode Insulation Resistance • Voltage: 1.24 kV • Voltage: 1.50 kV • Voltage: 500 V • HI-Limit: 10.00 mA • HI-Limit: 5.00 mA •...
  • Page 8: Conducting A Test

    Conducting a Test Adapter Box Connections The adapter box may be used for products that are terminated in either a two prong or three prong line cord. It is easy to use the adapter box for all six test functions. Remove the cover plate on the rear panel of the 6330 to access the DUT Output Connections.
  • Page 9: Test Result

    Test Result CAUTION HIGH VOLTAGE 5KVAC MAX. 6KVDC MAX. Once all the test connections are properly made power ON the 6330 tester and load the memory and test steps you need to perform. Once the desired tests are loaded press the TEST button to execute the test.
  • Page 10 Testers For Electrical Safety Compliance 28105 N. Keith Drive, Lake Forest, IL 60045 USA +1-847-932-3662 • +1-800-504-0055 • Fax +1-847-932-3665 • hipot.com • info@hipot.com © 2019 SCI, A Division of Ikonix USA SAFETY MADE SIMPLE® V2.0 6/19...