Troubleshooting; Specifications - HoMedics Walgreens BPA-430WGN Manual

Table of Contents

Advertisement

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
run down.
when the
START/STOP
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
incorrectly.
symbol shown on dis-
play or the blood pres-
sure value is displayed
excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with
the cuff on.
Er error symbol shown
Did you move during
on display - measure-
measurement?
ment result exceeds
measurement range.
Shaking of the arm with
the cuff on
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol shown
Inflation pressure
on display
exceeding 300 mmHg.
E3 error symbol shown
Data error.
on display
CA Appears on
Recalibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.
29
CORRECTION
Replace them with
four new AAA
alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Did you move during
measurement?
Shaking of the arm with
the cuff on
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Remove and reload the
batteries.
Refer to page 25 for
more information.

SPECIFICATIONS

Four 1.5V DC (AAA) batteries or
Power Source:
optional 12V DC AC adapter (not included)
Measurement
Oscillometric
Method:
Pressure: 40~250 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Pump driven
Deflation:
Automatic Pressure release valve
Memory Capacity:
48 memories for each User (96 total)
Auto-shut-off:
1 minute after last button operation
Operation
Temperature: 50°F~104°F (10°C ~40°C)
Environment:
Humidity: 40~85% RH max.
Temperature: 14°F~140°F (-10°C~60°C)
Storage Environment:
Humidity: 10~90% RH max.
Weight:
1.0 lb (462 g) (without batteries)
Arm Circumference:
Standard size cuff: 9.5"–14.2" (24-36 cm)
Large size cuff:13.4"–18.1" (34-46 cm)
5.6"(L) x 3.7"(W) x 2.8"(H)
Dimensions:
142mm(L) x 95mm(W) x 71mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
Note: These specifications are subject to change without notice.
30

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents