Performing Measurements
Creating Limit Lines
Indications of pass/fail
conditions
PASS
N O T E
Limits are checked only at the actual measured data points. If the point density is insucient, a
device can be out of specication without displaying
measurement points.
Limit lines can be
shifted
N O T E
The three oset keys (
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
POSITION OFFSET
the limit segments look out of place, check to see if they have been shifted in position or amplitude.
All manuals and user guides at all-guides.com
FAIL
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
AMPLITUDE OFFSET
) shift the limit segments but do not change the limit values in the table. If
FAIL
. Be sure to specify a suitable number of
!
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
AMP.OFF.
,
, and