Changing the span does
not a ect calibrations
Keep the reference cable stationary
Do not move the reference-extension cable connected between
B
the calibration. If you have an Option 001 accessory kit, use the kit's cable tray to help keep the
reference extension cable stationary.
Unlike the reference-extension cable, the test-port cable is una ected by changes in polarization due to
cable bending. Therefore, this cable may be moved without degrading the calibration.
Keep the balance knobs stationary
Do not move either of the front panel reference-polarization balance knobs after calibration. Moving
the knobs compromises the calibration.
All manuals and user guides at all-guides.com
on the front during or after a calibration. Moving this cable changes the polarization and voids
Performing Guided Setups and Calibrations