The operator or his agent is responsible for observing any additional standards, directives or laws if required due to operating conditions or place of installation. This applies in particular to the use of easily detachable process connections when measuring flammable media. 08/2017 - 4000422304 - AD DK32-34 II2G R04 en www.krohne.com...
5 Rating data: temperature & pressure rating 6 Ex data according to PTB 99 ATEX 2191 and IECEx PTB 09.0025 7 Built-in equipment 8 Data matrix 9 Safety instructions and disposal symbol www.krohne.com 08/2017 - 4000422304 - AD DK32-34 II2G R04 en...
Protection type intrinsic safety, protection level "ia" Suitable for gas group IIC, IIB and IIA T6...T1 T6...T1 Suitable for temperature class T6...T1 T6...T1 T6...T1 EPL, suitable for zone 1 08/2017 - 4000422304 - AD DK32-34 II2G R04 en www.krohne.com...
16 VDC 25 mA 64 mW Irrespective of the device design, the following values are to be observed for each intrinsically safe circuit in case of interconnection: 150 nF 150 µH 08/2017 - 4000422304 - AD DK32-34 II2G R04 en www.krohne.com...
The manufacturer is not liable for any damage resulting from improper use or use other than the intended purpose. This applies in particular to hazards due to insufficient corrosion resistance and suitability of the materials in contact with product. www.krohne.com 08/2017 - 4000422304 - AD DK32-34 II2G R04 en...
(nominal diameter range 6...9 mm). Connection diagrams Contact connection Cable colors of assembled cable Min minus white Min plus yellow Max minus green Max plus brown Lift slot Fastening screw of connection box 08/2017 - 4000422304 - AD DK32-34 II2G R04 en www.krohne.com...
1. The position of the ground terminal is illustrated below. This connection only ensures electrostatic grounding of the device and does not meet the requirements for equipotential bonding. DK32 - DK34 www.krohne.com 08/2017 - 4000422304 - AD DK32-34 II2G R04 en...
Setting of the limit switches may be carried out during operation. To do so, remove the housing cover. Close the housing cover immediately after the limit switches have been set. 08/2017 - 4000422304 - AD DK32-34 II2G R04 en www.krohne.com...
In the case of environmentally critical or hazardous products, appropriate safety precautions • must be taken with regard to residual liquids in the measuring unit. New gaskets have to be used when re-installing the device in the piping. • www.krohne.com 08/2017 - 4000422304 - AD DK32-34 II2G R04 en...