Advertisement

Quick Links

NEO
Veinlite
LED Window [A]
Side Button [C]
Battery Cover [D]
READ INSTRUCTIONS BEFORE USE
TransLite LLC
345 Commerce Green Blvd
Sugar Land TX 77478 USA
+1 281 240 3111
www.veinlite.com
info@veinlite.com
Caution
Veinlite NEO is designed for external examination
only.
To avoid contamination of the device, Veinlite NEO
must be used with the disposable plastic cover.
To avoid the spread of infection, always change the
disposable plastic cover between patients and wipe
the new cover with a recommended germicidal wipe.
Always turn the LEDs on after the device is in
position. Do not look directly at the LEDs when they
are illuminated or shine the light from them directly
into anyone's eyes.
Do not immerse the device in liquid.
Do not autoclave the device.

Introduction

The Veinlite NEO is a hand-held battery powered
transillumination device for visualizing veins and
arteries in neonates.

Applications

To assist in venous and arterial access in neonatal
pediatric patients.
End Button [B]
Cap [E]
Cover [F]

Operating Directions

Device:
1 Place Veinlite NEO against the skin with the LED
Window [A] in contact with the skin.
2 Depress the End Button [B] to turn on
the LED.
3 Apply gentle pressure to the Veinlite
NEO to maintain optimal skin contact.
4 Device will turn off after cycling through the
colors or you can depress and hold the End
Button [B] for one second to turn device off.
5 When you turn on a device with a low battery the
device will shut itself off.
6 Device will shut off after approximately 4 minutes.

Disposable Covers:

The Veinlite NEO should always be used with a
single-use disposable plastic cover to protect it from
contamination and to prevent the spread of infection
between patients.
1 Place the Veinlite NEO inside
the disposable plastic cover.
Check the cover to make sure it is
intact and the flap is on the side
away from the patient.
2 These covers are not sterile. They
should be wiped with a germicidal
wipe prior to use.
3 Discard the single-use
disposable plastic cover
after each patient.
The device is shipped with 50 covers.
Additional covers can be ordered from
TransLite or any of its authorized dealers.

Brightness Control:

Depress the Side Button [C] to change
the brightness. Each depression of the
button will cycle through low, medium, and
high brightness levels. The lowest brightness level
is the default setting.
 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NEO and is the answer not in the manual?

Questions and answers

Summary of Contents for Veinlite NEO

  • Page 1 Operating Directions Veinlite Device: 1 Place Veinlite NEO against the skin with the LED Window [A] in contact with the skin. LED Window [A] 2 Depress the End Button [B] to turn on End Button [B] the LED. Side Button [C] 3 Apply gentle pressure to the Veinlite NEO to maintain optimal skin contact.
  • Page 2 Selecting Colors: Warranty Depress the End Button [B] to cycle through these The Veinlite NEO has a one year parts & labor modes: warranty. Orange Optimal wavelength for highlighting the contrast between the venous blood Troubleshooting vessels and surrounding tissue.

This manual is also suitable for:

Vneo

Table of Contents