OEM PDD-300 User Manual page 15

Pulsed diode driver
Table of Contents

Advertisement

Rated power frequency
magnetic fields (IEC 61000-4-8)
Voltage dips / Voltage
interruptions (IEC 61000-4-11)
30 A/m, 50 Hz
0% U
for 0.5 cycle at 0°, 45°,
T
90°, 135°, 180°, 225°, 270°,
315°
0% U
for 1 cycle at 0°
T
70% U
for 25/30 cycles at 0°
T
0% U
for 250/300 cycles 0°
T
Recommended separation distance:
d = 1.2√P for 150 kHz to 80MHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Mains power quality should be that of a typical
environment. If the user of the device requires
continued operation during power mains interruptions,
it is recommended that the device is powered from an
uninterruptible power supply or battery.