The contents of this publication may not be reproduced or copied in whole or in part without the prior consent of bg edelstahl und kunststofftechnik für krankenhaus, industrie und wasserwirtschaft gmbh.
1. Introduction 3.1 Overview of functions / intended purpose 1. Einleitung The euroMAT pro bedpan cleaning unit is used for cleaning and disin- This operating and maintenance manual is intended to ensure that the 3.1 Funktionsübersicht Diese Bedienungs- und Wartungsanleitung soll den korrekten Gebrauch fecting urine bottles, bedpans and commode buckets.
Intensive program Standard program Short program Edge rinsing or WC rinsing 4. The control panel 4.2 Loading Now that you have briefly checked that the unit is ready for operation, Below, you will be provided with a detailed guide to the steps to be you can begin the cleaning process.
5. The programs The authorised specialist reserves the right to change these parame- ters. However, they can be changed on request. The manufacturer of- The door is locked throughout the entire cleaning process and cannot fers training (for a fee) in this respect. The effectiveness of the modified be opened until cleaning, disinfection and cooling are all complete.
6. Notifications on the display Dishes’ drainage time. 6.1 Operating notifications In error-free operation, the individual stages of the selected program can be followed in the text display. Rinsing process finished. Here is a list of the possible messages and what they mean: Before program start in the case of general readiness for operation.
Cause: Different temperatures Cause: The SD card is not writable > +-3°C between processor and co-pro- or not present Temperatursensor defekt cessor; or co-processor is not responding Behebung: Check the SD card Fehler SD Karte Remedy: Check the sensor and cable; (Service) replace the controller (Service) Cause: The door cannot be locked;...
Verfügung gestellt werden müssen. Finally, perform a test rinse. Please run a temperature logger (A0 • The surface of the euroMAT pro bedpan cleaning unit should be value) for checking during this process. treated with the manufacturer’s stainless steel care oil. It can be applied as required.
8.4 Maintenance schedule The following parts must be inspected, cleaned and (if necessary) replaced during annual maintenance. Abbreviations: I = Inspect C = Clean R = Replace V = Verification CA = Calibrate Year of operation Part CW supply hoses HW supply hoses Existing dirt strainers Solenoid valves...
8.5 Proof of maintenance The following tasks were carried out during maintenance work. Abbreviations: I = Inspect C = Clean R = Replace V = Verification CA = Calibrate Year of operation Part CW supply hoses HW supply hoses Existing dirt strainers Solenoid valves Rinsing chamber Rinsing impeller / rotary nozzle...
10. Technical data re measuring device from ebro. • bg test plug, 87°C euroMAT pro is monitored – compliant – inspected in accordance with: • Medical Devices Directive 9. Spare parts • DIN EN ISO 15883 Parts 1 and 3 9.1 Electrical parts...
10.4 Noise emission • Start the short program (urine bottle) and check that all the connec- tions are tight. The workplace level is 45.3 dB (A) • Check that the hot and cold water are connected correctly. The hot The accuracy class 2 measurement was conducted in accordance with water hose must heat up.
Installation and commissioning log IMPORTANT! Please fill out this log carefully and RETURN it to the manufacturer or to the relevant sales partner! Postal address: Name of the company: Contact: Street / no.: Postcode / Town/city: Serial number: Unit location: Unit type / designation: Installation: Unit installed Housing...
13. Medical devices book Unit master data Nature and type Serial number Year of acquisition Initial commissioning (MM/YYYY) Manufacturer Supplier Company ID number Location and operational assignment Deadlines for safety checks (see sticker on the product) Last updated: 18/05/2020...
Functional test and briefing Inventory no. Functional test Briefings Person responsible: Name of the person respon- Time Manufacturer / supplier Signature sible for the briefing Last updated: 18/05/2020...