Summary of Contents for bg edelstahl euroMAT basic
Page 1
Bedpan cleaning and disinfection unit With controller, operating manual and technical information Mit Steuerung Bedienungsanleitung und Technische Hinweise Last updated: 18/05/2020...
The contents of this publication may not be reproduced or copied in whole or in part without the prior consent of bg edelstahl und kunststofftechnik für krankenhaus, industrie und wasserwirt- schaft gmbh.
3.1 Overview of functions / intended purpose 1. Introduction The euroMAT basic bedpan cleaning unit is used for cleaning and disin- This operating and maintenance manual is intended to ensure that the 1. Einleitung fecting urine bottles, bedpans and commode buckets. Three programs following automatic cleaning and disinfection units are used properly 3.1 Funktionsübersicht...
Text display for program 4. Control elements and status messages Kurzprogramm mablauf Normalprogramm euroMAT basic Select a program Randspülung oder WC-Spülung Intensivprogramm Short program Spout program Standard program Intensive program enfeld 4.3 Loading 4.1 Operation 4.2 Beladen ahren sie detailliert die durchzuführenden Schritte, um Now that you have briefly checked that the unit is ready for operation, igungs- und Desinfektionsergebnis zu erhalten.
24-hour interval – When not in use, euroMAT units are rinsed automati- cally 24 hours after the last rinse. No additional rinsing is activated when euroMAT basic Select a program the units are used regularly. This function improves hygiene (the lines are rinsed and the units are always ready for immediate use).
Cause: The door contact’s switch is generating invalid signal Error – door contact Basic-Programm combinations; switch or cable Disinf. A0 = 2 defective Remedy: Check the switch and cable and replace them if necessary (Service) Basic-Programm Cause: The set temperature is not Disinfect.
4.6 Program end and unloading Cause: The tank is filled higher Close the door than the setpoint (inlet valve If none of the above error messages are visible in the text display once IMMEDIATELY defective) or automatic 24 h the entire cycle has ended, the cleaning program has been completed disinfection is on.
If the temperature displayed on the programming unit’s control panel is 85°C – 87°C, the value on the external measuring device • The surface of the euroMAT basic bedpan cleaning unit should be must be noted simultaneously (within one second), otherwise treated with the manufacturer’s stainless steel care oil.
6.2 Maintenance schedule The following parts must be inspected, cleaned and (if necessary) replaced during annual maintenance. Abbreviations: I = Inspect C = Clean R = Replace V = Verification CA = Calibrate Year of operation Part CW supply hoses HW supply hoses Existing dirt strainers Solenoid valves...
6.3 Proof of maintenance The following tasks were carried out during maintenance work. Abbreviations: I = Inspect C = Clean R = Replace V = Verification CA = Calibrate Year of operation Part CW supply hoses HW supply hoses Existing dirt strainers Solenoid valves Rinsing chamber Rinsing impeller / rotary nozzle...
6.4 Uninstalling / disposal Technical data Please proceed as follows to uninstall the bedpan cleaning unit: euroMAT basic is monitored – compliant – inspected in accordance with: • Unplug the mains plug. • Medical Devices Directive • Switch off the water supply (indoor installation).
Page 12
8. Medical devices book 8.1 Unit master data Nature and type Serial number Heating number Year of acquisition Initial commissioning (MM/YYYY) Manufacturer Supplier Company ID number Location and operational assignment Deadlines for safety checks (see sticker on the product) Last updated: 18/05/2020...
8.2 Functional test and briefing Inventory no. Functional test Briefings Person responsible: Name of the people respon- Time Manufacturer / supplier Signature sible for the briefing Last updated: 18/05/2020...
8.4 Malfunctions or repeated similar operating errors Inventory no. Time Description of nature and consequences Last updated: 18/05/2020...
Page 16
8.5 Malfunctions or repeated similar operating errors Inventory no. Time Description of the reported incident Last updated: 18/05/2020...
Page 17
Installation and commissioning log IMPORTANT! PLEASE FILL OUT THIS LOG CAREFULLY AND RETURN IT TO THE MANUFACTURER OR TO THE RELEVANT SALES PARTNER! Postal address: Name of the company: Contact: Street / no.: Postcode / Town/city: Serial number: Unit location: Unit type / designation: Installation: Unit installed Housing...
Need help?
Do you have a question about the euroMAT basic and is the answer not in the manual?
Questions and answers