Specifications; Electrical Specifications - AEMC 1005 User Manual

Megohmmeter
Table of Contents

Advertisement

CHAPTER 3

SPECIFICATIONS

3.1

Electrical Specifications

INSULATION TESTS
DC Test Voltage: 500V (Model 1005)
500V and 1000 V (Model 1015)
Megohm Range: 0.1 to 1000MΩ
Short Circuit Current: ≤ 6mA
Accuracy: ± 5% of reading
Discharging Time (Auto): 1s/µF
CONTINUITY TESTS
Range: 0 to 10Ω- ; 0 to 10Ω +
Short Circuit Current: ≥ 200mA
Open Circuit Voltage: 4.5 to 6.5V
Accuracy: ± 3% full scale
RESISTANCE TESTS
(Model 1015 Only)
Range: 0 to 1000Ω
Short Circuit Current: ≥ 2mA
Open Circuit Voltage: 4.5 to 6.5V
Accuracy: ± 3% full scale
VOLTAGE TESTS
(Safety Check)
Voltage Range: 0 to 600V
AC
Frequency: 45 to 400Hz
Accuracy: 3% of full scale