DD+DIS150
.05E
REQUIRED TOOLS:
•
Service PC with network interface and browser Internet Explorer
•
Calibrated densitometer MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite
310 or Gretag D200-2
•
Lint free cloth
•
Isopropyl - alcohol
•
Waterproof pen
•
Density reproduction check Excel sheets, available via MEDNET, GSO Library
•
Afssaps / EUREF Calibration Check Excel Sheet, available via MEDNET, GSO
Library
•
Connectivity Release Document for the corresponding modality to DRYSTAR
5302, available via MedNet HealthCare Library
Application
REQUIRED TIME:
60 to 120 min.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
General Info
Connectivity &
Chapter 3.6 / 18