AGFA DRYSTAR 5302 5366 Service Manual page 165

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
HC301020 : uppertraymodule.s11.flagproblem
Type
Context
Action
Reason
Cure
Part
HC301021 : uppertraymodule.s24.flagproblem
Type
Context
Action
Reason
Cure
Part
HC301022 : uppertraymodule.sensors
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Failure
W301020
None
S11 defect
Replace
MOD01
Failure
W301021
None
S24 defect
Replace
MOD01
Failure
W301022
None
one of the takeoverdetection sensors is not functioning
Remove jam
Upper tray module defect
Replace
MOD01
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 9 of 72

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents