AGFA DRYSTAR 5302 5366 Service Manual page 192

Tabletop, multi-size imager
Table of Contents

Advertisement

DD+DIS150.05E
HC306012 : outputmodule.cdm.home
Type
Context
Action
Reason
Cure
HC306013 : outputmodule.home
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC306014 : outputmodule.cdminstall
Type
Context
Action
Reason
Cure
HC306015 : outputmodule.intensity.unstable
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Accidental
None
CDM homing accidently failed
Self resolving accident
Accidental
None
Output homing accidently failed
Self resolving accident
Output module defect
Replace
MOD06
Failure
W306014
None
CDM needs calibration
Calibrate CDM
Failure
W306015
None
CDM intensity is unstable
Replace CDM module MOD06
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 36 of 72