Vishay G4-PM Technical Manual page 61

G4 series multi channel force instrument. pm/dt/he types
Hide thumbs Also See for G4-PM:
Table of Contents

Advertisement

Range/Alternatives
<default value>
1:C Value Cal. P1
Range: +/–999999
Unit:
Measurement Unit
<0>
1:C Transd. Signal P1
Range: +/–9.99999
Unit: mV/V
<0.00000>
1:C Value Cal. P2
Range: +/–999999
Unit:
Measurement Unit
<1000>
1:C Transd. Signal P2
Range: +/–9.99999
Unit: mV/V
<2.00000>
1:C Set Zero
Range +/-999999
Unit:
Measurement Unit
< 0 >
1:C Zero Offset
Range +/-999999
Unit:
Measurement Unit
< 0 >
Explanation and
result of alternatives.
This parameter defines, for Input C, the load in the lowest
calibration point, normally 0.
Note: This parameter is only shown if parameter
'Func. Block 1 Type' is set to '4 Channel' and 'F.Block 1 Inp.C' is
not set to 'Not In Use'.
This parameter defines the transducer signal, for Input C, in the
lowest calibration point.
Note: This parameter value is automatically acquired, and
cannot be edited, if '1:Calibration Type' is set to 'Known Force'.
It is only shown if parameter 'Func. Block 1 Type' is set to
'4 Channel' and 'F.Block 1 Inp.C' is not set to 'Not In Use'.
This parameter defines, for Input C, the load in the second
calibration point.
Note: This parameter is only shown if parameter
'Func. Block 1 Type' is set to '4 Channel' and 'F.Block 1 Inp.C' is
not set to 'Not In Use'.
This parameter defines the transducer signal, for Input C, in the
second calibration point.
Note: This parameter value is automatically acquired, and
cannot be edited, if '1:Calibration Type' is set to 'Known Force'.
It is only shown if parameter 'Func. Block 1 Type' is set to
'4 Channel' and 'F.Block 1 Inp.C' is not set to 'Not In Use'.
This parameter is used to zero input C for Function Block 1.
Enter wanted value for the actual force on Input C, usually '0',
i.e. unloaded transducer.
Note: This parameter is only shown if parameter
'Func. Block 1 Type' is set to '4 Channel' and 'F.Block 1 Inp.C' is
not set to 'Not In Use'.
This parameter shows the offset value acquired, for Input C, by
zeroing with '1:C Set Zero'. If this parameter is edited, the
zeroing of Input C will be influenced.
Note: This parameter is only shown if parameter
'Func. Block 1 Type' is set to '4 Channel' and 'F.Block 1 Inp.C' is
not set to 'Not In Use'.
G4 Multi Channel Force Instrument
4-21

Advertisement

Table of Contents
loading

Related Products for Vishay G4-PM

This manual is also suitable for:

G4-dtG4-he

Table of Contents