DC VOLTS
Ranges: 200mV,2V,20V,200V,600V
Resolution: 100µV
Accuracy: ±(0.35%rdg + 1dgt)
Input impedance: 10MW
Overload protection: 500VDC or AC rms on 200mV range
AC VOLTS (50Hz - 500Hz)
Ranges: 200mV,2V,20V,200V,600V
Resolution: 100µV
Accuracy: ±(1.0%rdg + 4dgts) on 200mV to 200V ranges
±(2.0%rdg + 4dgts) on 600V range
Input impedance: 10MW
Overload protection: 500VDC or AC rms on 200mV range
DC CURRENT
Ranges: 200µA,20mA,200mA,10A
Accuracy: ±(0.5%rdg + 1dgt) on µA,mA ranges
±(3.0%rdg + 1dgt) on 10A range
Input protection: 0.5A / 250V fast blow ceramic fuse
AC CURRENT (50Hz - 500Hz)
Ranges: 200µA,20mA,200mA,10A
Accuracy: ±(1.2%rdg + 4dgts) on µA,mA ranges
±(3.5%rdg + 4dgts) on 10A range
Input protection: 0.5A / 250V fast blow ceramic fuse
600VDC or AC rms on all other ranges
600VDC or AC rms on all other ranges
10A / 600V fast blow ceramic fuse
10A / 600V fast blow ceramic fuse