Operation and Maintenance Manual GASMAX II + GDS-IR GASMAX Gas Detector with GDS-IR Infrared Sensor for Combustibles and Carbon Dioxide GDS Corp. 1245 Butler Road League City, TX 77573 409-927-2980 409-927-4180 (Fax) ww.gdscorp.com...
Page 2
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 CAUTION: FOR SAFETY REASONS THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UNDERSTAND INSTRUCTION MANUAL COMPLETELY BEFORE OPERATING OR SERVICING. ATTENTION: POUR DES RAISONS DE SÉCURITÉ, CET ÉQUIPEMENT DOIT ÊTRE UTILISÉ, ENTRETENU ET RÉPARÉ...
Do not paint transmitter or sensor assembly. • Do not operate the GASMAX II + GDS-IR if its enclosure is damaged or cracked or has missing components. Make sure the cover, internal PCB’s and field wiring are securely in place before applying power.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 2 OVERVIEW The GASMAX II + GDS-IR gas monitor integrates the GASMAX II gas monitor with the GDS-IR standalone infrared gas sensor. The combination provides an ideal solution with local or remote display, built-in calibration, convenient IR % L E L Zero options and diagnostics with Fault display.
Page 8
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 GDS-IR INFRARED SENSOR FOR COMBUSTIBLES AND CO2 The GDS-IR is ideally suited for use in harsh environments and where the cost of required maintenance for conventional catalytic detectors is prohibitive. The GDS-IR Infrared gas detector will perform reliably in the...
To make calibration easier, GDS Corp recommends the remote sensor configuration for heavy gases. This allows the display to be located at eye level while the sensor is located near the floor.
Page 10
CAUTION: KEEP EXPLOSION PROOF COVER TIGHT WHILE CIRCUITS ARE ALIVE MISE EN GARDE : GARDER EXPLOSION COUVERTURE PREUVE TIGHT CIRCUITS PENDANT QUE SONT ALIVE INTRINSICALLY SAFE INSTALLATION The GASMAX II + GDS-IR is not certified for use as an Intrinsically Safe device. Page 10...
Plus 24VDC connects to TB2-1. DC Ground connects to TB2-4. Channel One 4-20mA output is available on TB2-2 and Channel Two 4-20mA output is available on TB2-3. GDS Corp always recommends using shielded wire for signal and power cable. TB2 – 1: +12-30VDC Input TB2 –...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 OPTIONAL RELAY + MODBUS INTERFACE The optional 10-0234 Relay / Modbus board is connected “piggyback” to the back of the GASMAX II Display Assembly and supplies two level alarm relays (“K ” and “K ”), a FAULT relay (“K3”) and an RS-485 Modbus RTU slave port.
MODBUS / RS-485 transmissions use low-voltage differential signaling to achieve reasonable data rates over very long distances, up to 4000 feet without a repeater. For MODBUS data signals, GDS Corp recommends 24GA shielded cable for best results. Power distribution, however may require larger gauge wire since it is critical that the supply voltage for the GASMAX at the far end of the string not fall below 18VDC during power-up and normal operation.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 disconnect the GASMAX II power or MODBUS connections without affecting any other MODBUS devices. “ ” “B” - 33- IO/Power upply ibbon to isplay “ ” “B” Figure 3-4: MODBUS Wiring (MODBUS Junction Box) The MODBUS wiring junction box is attached to the GASMAX, either on the bottom fitting or one of the side fittings, depending on installation requirements.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 OPTIONAL ISOLATED 4-20MA OUTPUT BOARD The optional 10-0250 Isolated 4-20mA output board provides two 4-20mA outputs that are electrically isolated from sensor inputs and the 24 VDC power source. Each 4-20mA output shares the same common ground and are therefore not isolated from each other.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 4 INITIAL SETUP When power is applied to the GASMAX II, the screen will show ‘G M II”. G -IR sensors may initially indicate off-scale high or low values, but should quickly return to zero within a few minutes if no target gas is present.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 Press the EDIT key to access the USER MENU display mode. When in user menu display mode, use UP and DOWN to select an item, EDIT to change an item, and NEXT to exit the menu or function and return the GASMAX II to display mode.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 Above the LCD display, three LEDs indicate the status of the level and fault alarms. If relays are installed, the LEDs indicate whether the relays are active. Note that if any relay is set for FAILSAFE operation, if LED is ON the relay will be de-energized.
Page 19
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 ALARM 1, ALARM 2 Alarm 1 and Alarm 2 are identical functions that will trigger front panel LED indicators and (if installed) optional alarm relays. Both can be set to activate whenever the input signal goes above or below a given...
5 minutes. Follow these GASMAX II calibration guidelines: • Calibration accuracy is only as good as the calibration gas accuracy. GDS Corp calibration gases are traceable to NIST (National Institute of Standards and Technology). • Never use calibration gas that has passed its expiration date.
% nitrogen. If the GDS-IR is configured to detect Carbon Dioxide, the zero air mixture should contain nitrogen 99.9% by volume. 3) A fixed flow regulator and a length of flexible tubing. GDS Corp recommends a flow rate of 0.5 liters / minute for standard combustible gases.
Page 22
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 IR ZERO PROCEDURE NOTE: This step should be performed initially and every 3 months for the first year, then every six months thereafter. If not performing the IR Zero procedure, proceed to GASMAX CALIBRATION PROCEDURE shown below.
6 MAINTENANCE NORMAL MAINTENANCE Normal maintenance for the GASMAX II + GDS-IR primarily involves periodic inspection of the unit followed by zeroing the GDS-IR and then calibrating the GASMAX + GDS-IR combination. GDS Corp recommends calibration at least every three to six months, or more often if temperature extremes, vibration, the presence of incompatible gases or other environmental factors may exist.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 7 USER MENUS MAIN MENU GASMAX II setup variables are stored in non-volatile memory and can be modified by the end user to better match a particular application. The GASMAX II Main Menu tree is shown below. To access the Main Menu, swipe the magnetic wand over the EDIT key.
0, 1 & 2 decimal points displays as “0”, “0.0” & “0.00”. CAL SPAN VALUE sets the engineering units value of the calibration gas. GDS Corp recommends that the calibration gas value be between 25% and 75% of full scale. Calibrating at 100% of scale is not recommended.
75%, the GASMAX II transmits 4.75mA during the 5-second interval NOTE: TX SENSOR LIFE SHOULD ALWAYS BE SET FOR NO UNLESS THE 4-20MA RECEIVER IS CAPABLE OF INTERPRETING THE SENSOR LIFE SIGNAL. MOST GDS CORP PROTECTOR SERIES CONTROLLERS INCLUDE THIS CAPABILITY.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 ALARM SETTINGS PAGE The Alarm Settings page covers the Alarm 1, 2, 3 Setup and Event Log menu items. Alarm 1, Alarm 2 and Alarm 3/Fail menus are identical and are therefore described only once. If the optional alarm relays are not installed, the alarm settings only affect the operation of the front panel LEDs.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 DISPLAY EVENT LOG displays the stored events in the Event Log. These include power-on, cold-boot, alarms and alarm acknowledge events. The event log stores the date and time of the most recent 300 events.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 INPUT TYPE indicates what kind of input or sensor the GASMAX II is configured to accept and is typically pre-configured at the factory. When installed, Smart Sensors upload sensor type and other data to the GASMAX II and this data may be viewed on the SMART SENSOR information screen.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 HELP SCREEN PAGE The Help screen contains several pages of information describing GASMAX II operation. DIAGNOSTICS PAGE The Diagnostics page provides tools for use during setup or testing. Since relays are not available on the GASMAX II, the Relay Test option is not functional.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 CAL Mode until the correct Pass Code is entered. LOW and HIGH security levels always allow static viewing of configuration menus. TECHNICIANS MENU GROUP The TECHNICIANS MENU group consists of the XMITTER CONFIG menu and TECHNICAL PAGE menu. These menus contain items that are typically factory configured depending upon the type sensor and input connected.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 TECHNICAL PAGE MENU The TECHNICAL PAGE menu is used to adjust certain parameters associated with local or remote sensors. To access the TECHNICAL PAGE menu, enter the Main Menu, move the cursor to the second page and select Technicians Only.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 WARNING: E CEE I G O ’ E E CI IO VO GE M Y M GE O OY HE SENSOR. Figure 7-13: Sensor Balance & Voltage Set Menu SET GAIN TO UNITY clears any previous calibration O E and G I values to “...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 TROUBLESHOOTING GDS-IR SHOWS FAULT INDICATION • GDS-IR output < 4mA due to internal error condition. See Maintenance for more information. • Incorrect channel activated. Verify channel connected to GDS-IR is enabled.
Page 35
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 • 03 – Sensor Calibration Error – Contact factory. • 04 – Zero calibration failure – Zero readings must be within 10% of zero. For example, if the range is 0-100 ppm, the zero reading cannot be above 10 ppm. For bridge-type sensors, it may be possible to readjust the balance to restore a zero reading.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 9 MODBUS INTERFACE The GASMAX II features a full complement of user-accessible MODBUS registers that can provide a complete snapshot of each gas detectors configuration. This includes all real-time data, preset zero, span and calibration values and user-programmable text.
Page 37
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 CH 2 Floating Point 33002 See Channel Floating Point definition Ch 1 Measurement Name 40401 - 40408 16 ASCII characters (2 per register) Ch 2 Measurement Name 40409 – 40416...
Page 39
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 ALARM STATUS WORD BIT DEFINITION Alarm Status Word Bit 0 Ch 1 Alarm 1 (1), no alarm (0) Bit 1 Ch 1 Alarm 2 (1), no alarm (0) Bit 2...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 10 SPECIFICATIONS Model GASMAX II + GDS-IR Single / Dual Channel Gas Monitor Power Input 18-30VDC at < 9 watts (375 mA) with relay board. Max Inrush current < 5x steady state current for < 33 mSec.
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 12 DRAWINGS AND DIMENSIONS 5.00 ” ¾” NPT (2x) 2.5” M E A S U R E M E N T N A M E xx.00” Cal Gas Inlet Figure 12-1: GASMAX II + GDS-IR Dimensions (Local Vertical Sensor)
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 5.00 ” 2.5” Cal Gas Inlet M E A S U R E M E N T N A M E xx.00” xx.00” Figure 12-2: GASMAX II + GDS-IR Dimensions (Local Horizontal Sensor)
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 m Input IO/Power upply ibbon to isplay Use magnetic wand or local pushbutton to Set Zero Figure 12-8: Wiring Diagram - Local GDS-IR Sensor Page 49...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 Use magnetic wand or local pushbutton to Set Zero at IR Sensor m Input IO/Power upply ibbon to isplay IR Zero Unavailable at Display Figure 12-9: Wiring Diagram - Remote GDS-IR Sensor (3-wire Interface)
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 Use magnetic wand or local pushbutton to Set Zero at IR Sensor m Input IO/Power upply ibbon to isplay Use magnetic wand or local pushbutton to Set Zero at Display...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 - m Output - 33-B IO/Power upply ibbon to isplay + V Input Power Common ocal G -I Infrared ensor ocal G -I Figure 12-11: Wiring Diagram - Local GDS-IR using p/n 20-0235 Interface PCB...
GASMAX II + GDS-IR Operation & Maintenance Manual, Revision 1.1 - m Output - 33-B IO/Power upply ibbon to isplay + V Input Power Common emote G -I Infrared ensor emote G Figure 12-12: Wiring Diagram - Remote GDS-IR using p/n 20-0235 Interface PCB...