Optional Load/Unload Station Installation ............2-15 Load Station - iQ200 Connection ..............2-15 Load Station - AX-4280 Connection..............2-17 Unload Station....................2-19 Startup......................2-21 Run Focus ......................2-22 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 3
CCD Camera # 700-3565.................3-17 20 x Objective # 220-3009 ................3-17 Auto Focus System # 700-3505 ...............3-17 Power System Assembly # 230-3000 ..............3-18 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 4
Material required ....................4-6 Procedure......................4-6 Specimen Tube Detector Alignment..............4-7 Material required ....................4-7 Procedure......................4-7 Specimen Tube Detector Calibration ..............4-8 Material required ....................4-8 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 5
Tree View Pane....................5-5 Commands ......................5-6 Preset Moves .....................5-6 Sensors ......................5-6 Register 0x0A.....................5-7 Calibration Table ....................5-7 Calibration Auto-Sets ..................5-7 Sequences ......................5-7 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 6
Sequences & Sensor Identifications for Troubleshooting.........5-17 Master (Master Controller) ................5-17 Sequences .......................5-17 Level sensors ....................5-17 SPA ........................5-17 Pumps/Evacuation pump/Sensors ..............5-18 STM........................5-18 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 7
SP-KSP are high in all Runfiles................6-6 Bubbles are present in fluid system ..............6-6 Waste well overflow....................6-6 Event Log Error Messages.................6-7 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 8
Adjustment verification ..................9-5 Sampler Rack alignment ..................9-5 Specimen Tube Detector verification ..............9-5 Barcode Reader alignment verification ..............9-5 SPA alignment verification .................9-5 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 9
Optical Bench Verification ..................9-6 Replenish Consumables ..................9-6 Performance verification..................9-7 Software ......................9-7 Calibration run ....................9-7 Control run......................9-7 Documenting the Preventive Maintenance............9-7 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Warnings ........................1-4 Biological Warning......................1-4 Warranty.........................1-5 Limitation of Liability.......................1-6 Use of Third Party Computer Products ................1-6 Iris Diagnostics Contact Information ................1-7 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
® This manual is intended to be used by Iris Diagnostics trained personnel. How to use this Manual References are made to other sections within the Manual. In order to...
Page 12
Clicking on a bookmark will take you automatically to the starting page of the selected section. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Biological Warning WARNING: Wear protective gloves to prevent exposure to pathogens. Discard contaminated materials according to applicable regulations. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Purchaser. This limited warranty is made on condition that immediate written notice of any defect be given to Iris Diagnostics and that Iris inspection reveals that the Purchaser’s claim is valid under the terms of this warranty.
1. Introduction Limitation of Liability Iris Diagnostics shall not be liable for any loss of use, revenue or anticipated profits, or for any consequential or incidental damages resulting from the sale or use of the products. Use of Third Party Computer Products...
Chatsworth, CA 91311 Telephone: From U.S. locations (800) PRO-IRIS (776-4747) From outside the U.S. +1-818-709-1244 Fax: +1-(818) 700-9661 e-mail: sales@proiris.com Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Run Calibration Rack....................2-28 Run Control Rack .....................2-28 Verify Load/Unload and Bridge Connections............2-28 Preparation for Starting Operation ................2-29 Perform Setup.......................2-29 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
90 to 240V AC, 50-60Hz input source. Neither voltage ® nor frequency parameters are customer configurable. Uninterruptible power supplies are recommended and available from Iris Diagnostics for and the Workstation (and the AX-4280 if applicable) to maintain ® system operation during short power outages and brownouts. This allows for an orderly shutdown of instruments without the loss of data.
Lamina Cap & Filter 700-3701 Drain Tubing 700-3702 Lamina Tubing 250-3501 Serial Cable, PC to Microscopy Module 700-3011 Operators Manual CD Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Description 800-3810 Load and Unload Set 800-3514 Load Station (only) 800-3515 Unload Station (only) 800-3501 Ax-4280 Bridge Connection Kit Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
50 F (10 C) and 86 F (30 C), and humidity in a range of 20% to 80%. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
¾ inch. • After adjusting the height, tighten the locking nuts. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
If direct drainage is not an option, slide the Drain Tubing inside a waste container (approximately 2 in. (5cm)) Make sure the waste container is lower than the microscopy module. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
7. Connect one end of the RS232 cable to the serial port on the computer; connect the other end to the serial port 1 (top port) on the back of the ® Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Station. Currently, the only configuration for this port is all three jumpers in place. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 26
Align the brackets and join with the connection bracket. Secure with four (4, M3x10 pan head) screws. Iris Diagnostics, a Division of Iris International, Inc. ® 2-10 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 27
NOTE: Do not use the grey cable provided with the Arkray kit, the pin out for this cable will not work due to incorrect pin positioning. Iris Diagnostics, a Division of Iris International, Inc. ® 2-11...
Page 28
2. Installation Procedures REMOVE jumpers on COM2 “Load” port before connecting directly to AX-4280 Iris Diagnostics, a Division of Iris International, Inc. ® 2-12 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 29
Rack numbers 29-30 PASS Rack number 31 NONE 14. Save these settings when prompted. 15. Return to Standby mode. Iris Diagnostics, a Division of Iris International, Inc. ® 2-13 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 30
18. Click on Interfaces. 19. Verify that the “Enable direct connection” checkbox with the AX- 4280 is selected. Iris Diagnostics, a Division of Iris International, Inc. ® 2-14 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 31
The label ID’s should match in order to attach the chemistry and microscopy results together. Iris Diagnostics, a Division of Iris International, Inc. ® 2-15...
4. Assemble the bridge (see picture below) Securing Brace Note position of guide plate Load station bridge assembly Iris Diagnostics, a Division of Iris International, Inc. ® 2-16 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 33
(see picture below) 6. Attach the load station cable to port 2 labeled “Load” on the Rear Panel SIC (jumpers in.) Iris Diagnostics, a Division of Iris International, Inc. ® 2-17 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 34
Load Station with an adhesive strip. Secure the brace to the brace mount using the setscrew. (see picture below.) Brace Mount Securing Brace Set Screw Iris Diagnostics, a Division of Iris International, Inc. ® 2-18 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 35
NOTE: Smooth rack transfer between stations and samplers is critical for proper operation. 9. Turn “ON” the power switch located on the back of the Load Station. Iris Diagnostics, a Division of Iris International, Inc. ® 2-19 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
10. Secure the black optical sensor to the side of the base plate touching the front of the sampler, using the hardware included in the bridge connection kit. (see picture below) Iris Diagnostics, a Division of Iris International, Inc. ® 2-20...
Page 37
13. Assemble the bridge (see picture below) Note position of guide plate Load station bridge assembly Iris Diagnostics, a Division of Iris International, Inc. ® 2-21 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 38
(see picture below) 15. Adjust the leveling feet under the Load Station to assure a smooth transfer of racks. Iris Diagnostics, a Division of Iris International, Inc. ® 2-22 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 39
ENTER. 19. Save these settings when prompted. 20. Turn “ON” the power switch located on the back of the Load Station. Iris Diagnostics, a Division of Iris International, Inc. ® 2-23 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
7. Assemble the bridge (see picture below) Note position of Unload station bridge assembly guide plate Iris Diagnostics, a Division of Iris International, Inc. ® 2-24 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 41
Unload station. (see picture below) 9. Adjust the leveling feet under the Unload Station to assure a smooth transfer of racks. Iris Diagnostics, a Division of Iris International, Inc. ® 2-25 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 42
3mm from the left side of the unload station.(See Sampler Calibration Chapter 4) Iris Diagnostics, a Division of Iris International, Inc. ® 2-26 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Note: Verify NO red LEDs are present on the circuit boards inside the cardcage. If red LEDs are observed, clear the fault before continuing the installation. 7. Logon to the workstation. Identifier [IRIS] password iris2k1 (lowercases). Note: Verify the Microscopy Module is in Standby mode as indicated by a status light (green) on the instrument and displayed on the top left side of the screen.
8. Repeat Control rack 10 times and verify CV’s of < 5% for the positive control runs. Verify Load/Unload and Bridge Connections 1. Transfer 6 mL of Iris Diluent into five sample tubes. 2. Place one sample tube per rack at positions 1 or 10. 3. Place the five racks into the Load Station.
(relating to rack position) status. The SM was specifically designed to use the Arkray Sampler. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Microscopy Module to skip positions where there is no sample tube, avoiding using consumables for non-existent specimens, and achieving higher specimen throughput. Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics...
Barcode Reader # 700-3315 The Barcode Reader uses CCD Array optical scanning process to read barcodes. Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics 9172 Eton Ave Chatsworth, Ca 91311 Ph# 818-709-1244 Iris Diagnostics, a Division of Iris International, Inc.
SPA SIC will still allow upward vertical motion of the SPA arm, but not downward motion. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
(samples and Lamina) used during sample processing, then drains the fluids through the Waste Pump into the waste tubing. Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics 9172 Eton Ave...
This pump fills the rinse well for cleaning the pipette. Fill Pump # 700-3874 This pump maintains the level inside the Lamina tank. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 54
3. Components Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics 9172 Eton Ave Chatsworth, Ca 91311 Ph# 818-709-1244 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
The Cassette-In-Place detection senses the presence of a small tab on the cassette indicating that the cassette has been inserted into the pump. If the tab is not detected, the pump will not run. Iris Diagnostics, a Division of Iris International, Inc. ® 3-10...
Page 56
3. Components Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics 9172 Eton Ave Chatsworth, Ca 91311 Ph# 818-709-1244 Iris Diagnostics, a Division of Iris International, Inc. ® 3-11 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
An error message with high priority is generated; sample processing is stopped until a new container is installed. Iris Diagnostics, a Division of Iris International, Inc. ® 3-12 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Air Tank # 700-3802 The Air Tank maintains air pressure used for the air mixing of the specimen inside the sample tube before pipetting. Component drawings may change without notice. Please contact IRIS Diagnostics to determine the current revision level. IRIS Diagnostics...
Located on the Specimen Presentation Assembly. When this valve is opened, sample and/or Lamina go to the Flow Cell. When closed, sample is aspirated in the flow line. Iris Diagnostics, a Division of Iris International, Inc. ® 3-14 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Air Charge Valve # 700-3888 The Air Charge Valve is used to refill the pressure in the Air Line for the mixing circuit. Iris Diagnostics, a Division of Iris International, Inc. ® 3-15 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
3. CCD camera 4. 20 x Objective Auto Focus Motor Strobe Light CCD camera Flow Cell 5. Auto Focus Motor Iris Diagnostics, a Division of Iris International, Inc. ® 3-16 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Focus position is achieved with an ultra-fine-pitch stepper motor providing a movement of about 0.2 microns per full step. Iris Diagnostics, a Division of Iris International, Inc. ® 3-17 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
LED lights only when DC power is on (via the Power SIC) and the fuse is blown. Iris Diagnostics, a Division of Iris International, Inc. ® 3-18 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Do-All SIC power F13 D13 0.50A Rear Panel SIC power F14 D14 0.50A +24V Sample Transport Module SIC power Iris Diagnostics, a Division of Iris International, Inc. ® 3-19 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 65
F23 D23 2.5A +24V Strobe Power Supply power F24 D24 1.0A +24V Camera 24VDC to 12VDC Inverter power Iris Diagnostics, a Division of Iris International, Inc. ® 3-20 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Lamina ESN Bd 101-5048 container Waste Well FITD 101-5054 Waste well OBA Focus Home 101-5055 Above CCD camera Iris Diagnostics, a Division of Iris International, Inc. ® 3-21 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
There are two DCVB boards with green card ejectors. They are interchangeable. • Slot 4 Sample Presentation Assembly • Slot 5 Fluidic block Iris Diagnostics, a Division of Iris International, Inc. ® 3-22 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Lamina Pump SIC J1 Lamina Pump 2-pin receptacle. Cardcage Fan Cardcage fan 2-pin receptacle. Main SM Fan Main SM fan Iris Diagnostics, a Division of Iris International, Inc. ® 3-23 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
The signals are brought to SMCS backplane J25 through a 40-wire ribbon cable. The FSV SIC is mounted alongside the Do-All SIC on the Fluid System Valve Assembly. Iris Diagnostics, a Division of Iris International, Inc. ® 3-25...
Page 71
Operates EBV 12-pin female Not used 12-pin female Not used 12-pin female Not used 12-pin female Not used Iris Diagnostics, a Division of Iris International, Inc. ® 3-26 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Rot A, Rot B Green Crash Status Vert OK Green Crash Status Vert Crash Motors Disabled Motors Disabled Iris Diagnostics, a Division of Iris International, Inc. ® 3-27 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Sensor switch ESN switch SPA vertical & rotational 10-pin ribbon header SPA Opto Bd homes and other limit. Iris Diagnostics, a Division of Iris International, Inc. ® 3-28 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Iris config.: Dual opto-focus Focus Position female home. Newport config.: Optical Sensors P1 end limits for linear actuator. Iris Diagnostics, a Division of Iris International, Inc. ® 3-29 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
12-pin female Not used Not used 12-pin female Not used Not used 12-pin female Not used Not used Iris Diagnostics, a Division of Iris International, Inc. ® 3-30 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
STD PCBA # 101-5034 The Specimen Tube Detector board is a small electro-optics assembly that is part of the Specimen Tube Detector. It senses the presence of a Iris Diagnostics, a Division of Iris International, Inc. ® 3-31 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
(if any) of the 22 fuses are blown. 3. On the various remote circuit boards, they are appropriately placed for the application. Iris Diagnostics, a Division of Iris International, Inc. ® 3-32 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Collimation ........................4-22 What is a Collimation?....................4-22 Condenser Alignment ....................4-22 Run Auto Focus......................4-23 Flow Cell Replacement ....................4-24 Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 79
Flow Cell Removal....................4-24 Flow Cell Installation....................4-25 Flow Cell Lateral Adjustment ..................4-26 Flow Cell Tilt Adjustment ..................... 4-27 Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Results Processor, the system will prompt you to turn the Microscopy Module back on. After about 1 minute, the Service Screen will be displayed. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
1 to the difference of register 0x0A and calibration table entry 0. See Register 0x0A. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 82
4. Adjustments Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Note: Value “00000275” will position the rack short of the original value, “00000285” will move the rack further than the original value. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Note: Changing the value 10 Hex will change the rack move distance approximately 1 millimeter. 10. Repeat the “MN” “Move Next” command as needed to repeat this test. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
“Initialize Rack (‘M1’)” then “OK” to activate the Sampler and present the test tube rack to the #1 position. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 86
10. From the Service Application menu, click on STM’s “Calibration Table”, then click on “Tube Detect Threshold”, enter the calculated value and click “OK”. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Range should be greater than 0X30Hex Replace STD assembly if this test fails Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
NOTE: This should position the test tube located in position 1 into the correct sampling position. If not recalibrate the STM. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
#1. 12. Exit the service application. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Then click on “P2” pipette up and rotate and down to waste well and finally up from waste well. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 91
PH-Pipettor Home. 8. Perform the “SPA pipette depth calibration check” procedure. If the system hangs during this procedure, you will need to call Iris Technical Service for cal [31] to determine this new calibration value. 9. Exit Service Mode. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc.
Lamina Consumption Adjustment Cal 30 value is pre-set in factory. If any adjustment is needed, contact Iris Technical Services. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
If the sample overflows the top of the tube, decrease the air pressure. 9. Acceptance criteria: a. Air pressure: 5.0 ±0.5psi. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Prepare Barcode Reader Connections 1. Turn the iQ200 microscopy unit off. 2. Remove the Lower Front Panel assembly. 3. Use the Barcode Reader “Y” cable (IRIS part# 250-3084) for RS232 and 5 VDC power connection to the Barcode Reader. Connect to...
6. Turn on the iQ200 to power up the barcode scanner. 7. Connect the RS 232 cable to the communication port of the laptop. Load the Alignment Tool 1. Load the “Barcode Reader Alignment Tool” (IRIS part# 725-3090) like a normal rack in the iQ200 sampler. Alignment Graphic Barcode Reader Alignment Tool 2.
Page 96
Needs to move to the right RIGHT Needs to move to the left CLOSE Is to close Is to far Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 97
J13 connector it is keyed to go in one way (key is facing down for proper orientation). 13. Exit the Service Application. 14. Restart the entire iQ200 system. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
4. If the barcodes do not read properly, recheck the alignment (it will be necessary to reinstall the “y” cable and hook up the laptop again to do so). Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
3. In Video View, the Avg light should be about 210 and the Var. should be lower than .035. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
4. Adjustments 4. Turn the vertical knob clockwise to reach one edge of iris. Vertical adjustment Horizontal adjustment 5. Turn the vertical adjustment knob counter clockwise to the other edge of the iris. 6. Adjust the vertical knob until the condenser is centered.
The stainless steel cannulas are very fragile and any damage to the cannulas will destroy the flow cell. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
1. Connect the four tube assemblies to the flow cell. 2. Slide the flow cell into the mount, and then tighten the two locking screws. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
NOT display any cells. 9. If a lateral alignment is needed, use the knob highlighted on the right. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
A properly tilted flow cell will show the green dots form a circle that is centered in the camera view. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 105
When the FI is on the light side of focus, the field curvature will begin to show if your tilt is correct. Side adjustment knob Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 106
16. After the Focus passes, run a Calibration rack. 17. After the Calibration rack passes, run a Control Rack. Iris Diagnostics, a Division of International Remote Imaging Systems, Inc. ® Sprint™ Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Show Versions Report ..................5-11 Exit ........................5-11 Edit Menu .........................5-11 View Menu........................5-11 Toolbar........................5-11 Status Bar ......................5-12 Polling ........................5-12 Video........................5-12 Record Video Statistics..................5-12 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 108
FBA Pumps and Sequences ..................5-25 STM Motors 1 ......................5-26 STM Motors 2, Sensors, Calibration Table and Sequences........5-27 OBA ..........................5-28 Versions 1.........................5-29 Versions 2.........................5-30 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Microscopy Module. When connections are established with the Results Processor, the system will prompt to turn the Microscopy Module back on. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Tip: While waiting for the connection to be established, you may want to review the Event log. Using the mouse, select File, View Event Log. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Schematic Pane, but the tree contains many more. Types of items found in the tree: • Commands • Preset Moves • Sensors • Calibration Table • Sequences Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
If the question marks persist for a particular sensor it is probably an indication of a problem with the associated micro controller. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
M1 sequence, the program will let you queue up the MN sequence or the home command, but neither one will actually be run until the M1 sequence completes). Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
The peristaltic pump is shown surrounded by sensor values. If a sensor is inactive, no value is shown. If a sensor is active, it shows a corresponding letter: Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
S – Stall. Position error exceeded specified limits. • D – Done. Motor has reached target position. • M – Motor Energized. Drive output is enabled. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
The address line identifies the point in the state records where the system halted. Chapter 8 contains the state records that can be used to locate problems. Iris Diagnostics, a Division of Iris International, Inc. ® 5-10 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
View Menu Toolbar Check this item to show the toolbar. Uncheck it to hide the toolbar. Iris Diagnostics, a Division of Iris International, Inc. ® 5-11 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Tip: The Terminal Window should only be used by very advanced users or under the direct supervision of very advanced users. Iris Diagnostics, a Division of Iris International, Inc. ® 5-12...
(not all messages expect a reply, but many do). The message currently being processed is displayed in the current message status area. If no message is currently being Iris Diagnostics, a Division of Iris International, Inc. ® 5-13...
ASTM to a more or less raw serial protocol (gaining bandwidth at some sacrifice of robustness). Finally, it issues commands to the analyzer control system to put it into service mode. Iris Diagnostics, a Division of Iris International, Inc. ® 5-14...
Rear Panel SIC – Swap the Rear Panel SIC. “Connection established” This status is briefly displayed after successfully establishing the connection. Iris Diagnostics, a Division of Iris International, Inc. ® 5-15 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Scroll Lock Status Displays the status of the keyboard Scroll Lock key. Not particularly useful, but many programs have this. Iris Diagnostics, a Division of Iris International, Inc. ® 5-16 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Test tube down (TD) Test Tube Down (TD) Moves pipette down into the tube. Can only be used after To Tube (TT) Iris Diagnostics, a Division of Iris International, Inc. ® 5-17 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
0 = No cassette 1 = Cassette in place Cover in place 0 = Cover open 1 = Cover closed Iris Diagnostics, a Division of Iris International, Inc. ® 5-18 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
What Should I Do If the Service Application Appears to Hang? The most likely cause would be diagnosed with the Current Message Status indicator. See that section for details. Iris Diagnostics, a Division of Iris International, Inc. ® 5-20 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software Service Tree Master Iris Diagnostics, a Division of Iris International, Inc. ® 5-21 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software SPA Motors Iris Diagnostics, a Division of Iris International, Inc. ® 5-22 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software SPA Pumps and Valves Iris Diagnostics, a Division of Iris International, Inc. ® 5-23 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 130
5. Service Application Software SPA Calibration Table and Sequences Iris Diagnostics, a Division of Iris International, Inc. ® 5-24 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software FBA Pumps and Sequences Iris Diagnostics, a Division of Iris International, Inc. ® 5-25 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software STM Motors 1 Iris Diagnostics, a Division of Iris International, Inc. ® 5-26 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software STM Motors 2, Sensors, Calibration Table and Sequences Iris Diagnostics, a Division of Iris International, Inc. ® 5-27 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software Iris Diagnostics, a Division of Iris International, Inc. ® 5-28 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software Versions 1 Iris Diagnostics, a Division of Iris International, Inc. ® 5-29 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Service Application Software Versions 2 Iris Diagnostics, a Division of Iris International, Inc. ® 5-30 Sprint™ (2G)Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
SP-KSP are high in all Runfiles ..................6-6 Bubbles are present in fluid system................6-6 Waste well overflow....................6-7 Event Log Error Messages ..................6-7 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Bypassing Error Condition If the instrument displays a red flag error condition, reboot the instrument and login as IRIS as login name and iris2k1 as password. Check the error log for details of error condition. Decoding the Valves field in Valve Error Reports The Valves field in an iQ200 valve error report is a hexadecimal number indicating the valves that failed.
Focus is too far out for system to find starting point 1. Perform manual focus in Service mode Failed OBA nudger 1. Verify that nudger operates in Service mode Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
5. Barcode reader is misaligned 6. “White space” or “quiet zone” is too small 7. Labels too high or too low on tube Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Bubbles are present in fluid system 1. Cannula pump manifold leak 2. Loose fitting in fluid system 3. Faulty peristaltic pump cassettes Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Sequence 27 or 29, Error position fails to timeout reach down MC State 153 Sequence 6 Focus dial not moving Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 144
Detail: scanner error. Spare. SPA State 90 Sequence 25 Spa rotational Error: timeout motor failed to home Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 145
SPA State 119 Sequence 28 Rotation fails to Error: timeout, Detail: N/A reach position Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 146
Error: timeout, detail: N/A homing FBA State 42 Sequence 17 CP not moving Error: timeout, detail: N/A at proper velocity Iris Diagnostics, a Division of Iris International, Inc. ® 6-10 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 147
Module........................7-5 Addresses and Summary ...................7-6 Hardware Bit Assignments and Details ..............7-11 Hardware Assignment Cross Reference ..............7-21 ESN I2C Addresses....................7-27 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2004...
Page 148
The electronic serial number system, an electronic system for identifying the individual assemblies, subassemblies and other elements that represent the totality of the iQ200 Microscopy Module. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2004...
Page 175
700-3800 System Fluid Assembly..................8-3 Master Controller State records ..................8-4 SPA State records......................8-20 FBA State records......................8-24 STM State records .......................8-26 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 203
Fluidic system inspection..................9-12 Auto-Focus run ......................9-13 Optical Bench Verification ..................9-12 Performance verification...................9-13 Calibration run ......................9-13 Control run........................9-13 Reporting ........................9-13 Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
9. Preventive Maintenance ® 200 Preventive Maintenance Procedure. This document describes Iris Diagnostics’ Preventive Maintenance procedures ® for the iQ 200. There are two (2) classes of preventive maintenance: namely the Semi-Annual Preventive Maintenance and the Annual Certification. The Semi-Annual Preventive Maintenance should be performed six (6) months after the installation date or six (6) months after the last Annual Certification regardless of the number of samples the instrument has processed.
Fluidic system flush 1. Prepare a CONTROL rack with Iris System Cleanser solution in position 1 and two (2) tubes of Iris Diluent in positions 2 and 3. 2. Run this rack three (3) times. Iris Diagnostics, a Division of Iris International, Inc.
1. Clean the surfaces of the Rinse/Waste Well Assembly with 1:10 diluted Iris System Cleanser. 2. Turn-off the instrument and remove the front panel and the top panels. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
3. Clamp the tube between the Lamina Bottle and the Lamina Pump (green connector line) with a hemostat. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 208
Once the new cassette has been secured, remove the hemostat. 5. Verify that all the peristaltic pump doors are closed properly. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
2. Replace the Tubing fitting assembly (PN#700-3857) between the EBV Common Port and the line to the Evacuation Pump input. Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 210
DRV valve and the Flow Cell de-bubble port. 4. Replace the Specimen filter (PN#700-3891), the arrow points away from the pipettor. (Only for Annual Certification procedure). Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 211
Remove the other portion by unscrewing it off the top of the Flow Cell, use caution not to damage the Flow Cell! Iris Diagnostics, a Division of Iris International, Inc. ® Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Page 212
Connect the green fitting to the PBV valve COM Port. 7. Replace the Tubing Assembly (PN#700-3806) from the Waste Well Yellow connector to the Evacuation Pump. Iris Diagnostics, a Division of Iris International, Inc. ® 9-10 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Barcode scanner alignment verification 1. Use the Barcode Utility Software and verify the scanner alignment. Adjust if necessary. Iris Diagnostics, a Division of Iris International, Inc. ® 9-11 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...
Chapter 4 Adjustments for the correct procedure. Fluidic system inspection Run a CONTROL rack with Iris System Cleanser in position 1 and Iris Diluent in positions 2 and 3. Run this rack three times to assure the entire fluidic system is primed with no air bubbles in any lines.
Reporting Report all the results on the PM check list form and send the form to your regional service contact at Iris Diagnostics. Iris Diagnostics, a Division of Iris International, Inc. ® 9-13 Sprint™ (2G) Automated Urine Microscopy Analyzer Service Manual 300-4949 Rev A 01/2005...