IRIS M510 Operator's Manual

Statspin express 4

Advertisement

Horizontal
StatSpin
Express 4
Centrifuge
®
Operator's Manual
A Division of IRIS International, Inc.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M510 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for IRIS M510

  • Page 1 Horizontal StatSpin Express 4 Centrifuge ® Operator’s Manual A Division of IRIS International, Inc.
  • Page 2 Express 4 ® Horizontal Centrifuge Model Number M510 For in-vitro diagnostic use Product Number ® SSH4 – StatSpin Express 4 for 100-240 VAC, 50/60 Hz StatSpin and Iris are registered trademarks of IRIS International, Inc. Copyright 2008 Printed in U.S.A.
  • Page 3: Table Of Contents

    Table of Contents How to use this manual…………………………………………..……….………… ii Section 1 ……………………………………………………………..……….……….. 1 Unpacking and Installation ..………………………………………………………... 1 Inspect Packaging ………………………………………………………………...………....Verify Contents ………………………………………………………………...………......Install ………………………………………………………………...………..........Connect Power ………………………………………………………………...………......Section 2 ……………………………………………………………..……….……….. 2 System Overview ……………………………………………………………………. . 2 Principle and Intended Use ………………………………………………………………...………...
  • Page 4: How To Use This Manual

    How to use this manual This manual along with information contained on product labels and in package inserts should ® provide you with all the information you need to operate and maintain the StatSpin Express 4. When the information requires special attention, a caution symbol appears next to the italicized text.
  • Page 5: Section 1

    Section Unpacking and Installation Inspect Packaging ® The StatSpin Express 4 and its accessories are delivered in a single carton. If the centrifuge or accessories have been damaged in transport, please inform your carrier immediately. NOTE: Save the shipping carton and packing materials to facilitate return should service be required. Verify Contents The package contains: ®...
  • Page 6: Section 2

    The 10 minute cycle generates 3100 xg at 4500 rpm. This setting provides lower g-forces for procedures which require a traditional spin. Symbols and Definitions Product/Reference Indicates the Iris Sample Processing product/catalog Number number Caution Statement of caution/warning: Read carefully.
  • Page 7: Operator Controls

    Operator Controls Start The Start button initiates the selected cycle. The Stop/Open button interrupts the cycle and Stop/Open releases the cover when the rotor has stopped. The 3 min button selects a cycle of three minutes at 3 Minute Cycle 3 min 5100 rpm.
  • Page 8: Error Codes Table

    Error Codes Table Failure to reach speed in 30 seconds. 3 min Confirm no obstruction in bowl area, contact customer service. Latch is not locked. 5 min Confirm cover is fully seated in closed position. Rotor exceeded set speed. 3 min 5 min Remove and inspect rotor, reinstall.
  • Page 10: Section 3

    Section Operating Instructions Opening and Closing the Cover An electronic operated cover interlock mechanism prevents operation if the cover is not completely closed and locked, and prevents the cover from opening when the rotor is turning. An operating cycle can be initiated only if the cover latch is engaged. Cover Interlock By-pass In case of power failure or malfunction, the Stop/Open button may not release the cover.
  • Page 11: Spinning The Sample

    Reinstall center screw and tighten with supplied hex key. Reinstall shield into housing. Align shield hooks with cut-outs in housing. Grasp shield at center …….and rotate counter-clockwise. If shield is not locked into place instrument will not operate. Spinning the Sample ®...
  • Page 12: Section 4

    Section Maintenance Overview Iris Sample Processing recommends that instrument operators perform periodic inspections and preventative maintenance on all instruments. Contact customer service or your distributor if the instrument is not functioning properly. ® CAUTION - Unplug the StatSpin Express 4 from the wall outlet before cleaning or performing maintenance.
  • Page 13: Troubleshooting

    Troubleshooting Symptom: Solution: Red Error / Service indicator flashing, unit will not spin. Refer to Error Code Table of manual to identify and correct problem. Be sure the correct tube inserts are in use. Check insert listing Tubes are loose and do not fit into buckets. available under accessories to match with appropriate tube sizes.
  • Page 14: Service

    Service Refer all service to Iris Sample Processing or an Authorized Repair Facility. Instrument fuse(s) are located on the rear panel. Remove the line cord to access the fuse holder. Replace with fuses of the same type and rating. 2.5A 250V F 5 x 20 mm Be sure to complete the Warranty Registration Card as directed.
  • Page 15: Section 5

    Appendix Specifications Product Number SSH4 Model Number M510 Spin Parameters 3 min: 180 seconds @ 5100 rpm / 4000 xg 5 min: 300 seconds @ 5100 rpm / 4000 xg 10 min: 600 seconds @ 4500 rpm / 3100 xg ≤...
  • Page 16: References

    CDC. Updated: US Public Health Service Guidelines for the Management of Occupational Exposures to HBV, HCV and HIV and Recommendations for Post Exposure Prophylaxis. Appendix A and B. MMWR 50 (RR-11): 1-42, June 29, 2001. A Division of IRIS International, Inc. Westwood, MA 02090 USA Phone: 800-782-8774...
  • Page 17: Warranty

    The purchaser shall be deemed liable for any and all claims, losses, or damages incurred by the use or misuse of the Iris Sample Processing instrument by the purchaser, its employees or others, following receipt of the instrument or other items.