Manamed PlasmaFlow User Manual
Hide thumbs Also See for PlasmaFlow:

Advertisement

USER GUIDE
Scan QR Code to watch
Instructions for Use
Vascular Therapy System
(Compressible Limb Sleeve Device)

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PlasmaFlow and is the answer not in the manual?

Questions and answers

Teresa Bowdy
March 19, 2025

I **** charging the device while using but it is not inflating. The light is red and a small light above it is flashing red and it has been plugged in for 30 minutes. How will I know it’s charging?

User image 67db31976d978
1 comments:
Mr. Anderson
March 19, 2025

If the Manamed PlasmaFlow device's light is red and flashing, it indicates that the battery voltage has dropped to a critical level, and the device has stopped operating. To confirm that it is charging, check for the charging light indicator, which should illuminate when the device is receiving a charge. Additionally, the battery icon on the LCD screen should show movement from right to left while charging.

This answer is automatically generated

Paul wWallace
April 21, 2025

I am required to use compression stockings, can I used this equipment without using the compression stockings?

Summary of Contents for Manamed PlasmaFlow

  • Page 1 USER GUIDE Scan QR Code to watch Instructions for Use Vascular Therapy System (Compressible Limb Sleeve Device)
  • Page 2: Table Of Contents

    Quick Start Instructions for Wearing and Operating PlasmaFlow ....... . .
  • Page 3: Plasmaflow Label Symbol Descriptions

    Federal Law (U.S.A.) restricts this device to sale, distribution or use by or on the order of a physician or properly licensed practitioner. PlasmaFlow is only intended for the use by the individual for whom it was prescribed. The device is ONLY for single patient use.
  • Page 4: Plasmaflow Labeling Information

    Indications for Use PlasmaFlow model PF0001 is intended to be an easy-to-use portable system, prescribed by a physician, for use in the home or in a clinical setting to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).
  • Page 5: Plasmaflow Overview

    Each house control unit contains a battery and internal electronics to ensure the pump is inflating and delivering air to the leg cuffs. PlasmaFlow control unit displays the mode, timer, and air pressure of the device while turned on. Refer to “Getting Started” section on page 4.
  • Page 6: Getting Started

    Important: Charge both device cuffs before first use. Warning: Use only the charger provided by ManaMed. The use of the wrong charger can cause excessive heat, damage to the circuit and shorten the life of the battery.
  • Page 7: Switching Modes

    The PlamsaFlow unit is pre-set to “Mode 1.” Only switch to “Mode 2” if instructed by your physician. In order to operate the PlasmaFLow unit in “Mode 2”, tap the Power Button once while the unit is powered on. The screen on the left side of the Power Button will display “O”...
  • Page 8: Cleaning And Disinfecting

    Do not use an extension cord with the battery charger as it may cause overheating. Do not use the battery charger with other devices as it may cause damage to the device or battery charger. If the battery area on the PlasmaFlow system becomes overheated, discontinue using and contact ManaMed Customer Service at 888-508-0712.
  • Page 9: Disposal

    PlasmaFlow Vascular Therapy System or refund the original purchase price. This will constitute the Purchaser’s sole remedy. This limited warranty does not extend to any resale or other transfer of the PlasmaFlow System by Purchaser to any other person or entity.
  • Page 10: Compliance Statements And Declarations Electromagnetic Compatibility (Emc) Tables

    Emissions Tests Compliance Electromagnetic Environment Guidance The PlasmaFlow uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely RF Emissions CISPR11 Group 1 to cause any interference in nearby electronic equipment.
  • Page 11 THE PLASMAFLOW The PlasmaFlow is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PlasmaFlow can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the PlasmaFlow as recommended below, according to the maximum output power of the communications equipment.
  • Page 12 Scan QR Code to watch Instructions for Use Manufactured By: 5240 W Charleston Blvd., Las Vegas, NV 89146 888-508-0712 (toll free) CustomerService@manamed.net www.manamed.com PlasmaFlow is a registered trademark of ManaMed, Inc. ManaMed and the ManaMed logo are trademarks of ManaMed, Inc.

This manual is also suitable for:

Plasmaflow e0676Pf0001

Table of Contents

Save PDF