Maintenance; Troubleshooting; Lens Cleaning - Calex PyroMini Series Operator's Manual

Hide thumbs Also See for PyroMini Series:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

maintenance

Our customer service representatives are available for application assistance, calibration, repair, and
solutions to specific problems. Contact our Service Department before returning any equipment. In
many cases, problems can be solved over the telephone. If the sensor is not performing as it should,
try to match the symptom below to the problem. If the table does not help, call Calex for further advice.

Troubleshooting

symptom
No output
Erroneous temperature
Erroneous temperature
Erroneous temperature

lens cleaninG

Keep the lens clean at all times. Any foreign matter on the lens would affect measurement accuracy.
Blow off loose particles (if not using the air purge accessory) with an air 'puffer'.
gUarantee
Calex guarantees each instrument it manufactures to be free from defect in material and workmanship
under normal use and service for the period of two years from the date of purchase. This guarantee
extends only to the original buyer according to Calex terms and conditions of Sale.
probable cause
No power to sensor
Incorrect wire connection
Faulty sensor cable
Field of view obstruction
SEN TRONIC
056 222 38 18
AG
solution
Check power supply
Check wire colour codes
Verify cable continuity
Remove obstruction
www.sentronic.com
mailbox@sentronic.com
1 9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PyroMini Series and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents