Cerabar S FF
2.2 Mounting instructions with diaphragm seal
The Cerabar S with diaphragm seal is
screwed
in,
flanged
depending on the type of diaphragm seal.
•
The protective cap of the diaphragm
seal should only be removed just before
mounting in order to protect the
diaphragm.
•
The diaphragm of the diaphragm seal of
the Cerabar S must not be pressed in or
cleaned with pointed or hard objects.
•
The diaphragm seal and the pressure
sensor together form a closed and
calibrated system which is filled with
filling fluid through a hole in the upper
part. The following rules should be
observed:
This hole is sealed and not to be
–
opened.
The instrument should only be turned
–
by the diaphragm seal and not by the
housing.
For level measurement the Cerabar S must
always be installed below the lowest
measuring point.
•
The instrument should not be installed
within the flow of material, in the outlet
of the tank or at a place in the tank
where pressure pulses from agitator
blades may occur.
•
Calibration and function testing are
easier to carry out if the Cerabar S is
mounted downstream from a shut-off
valve.
The use of temperature spacers is
recommended
when
extreme product temperatures cause the
maximum
permissible
temperature to exceed +85 °C.
•
Note when mounting that the
temperature spacer increases the
maximum height by 100 mm.
•
Due to the hydrostatic column in the
temperature spacer, the increased
height also causes a zero point shift of
approx. 10 mbar.
Endress+Hauser
or
clamped,
the
continuous
ambient
0 - 10 bar
0 - 10 bar
2 Installation
Cerabar S
with diaphragm seal
– PMC 635
– PMP 635
Figure 2.6
When screwing in the Cerabar S
with diaphragm seals, turn by the
diaphragm seal only, not by the
housing.
Level measurement
Mounting with
temperature spacers
Figure 2.7
Temperature spacer
11