Translation of the viewing plane closer to MV:
Before
After
Press Image Store once satisfied with the image.
Temporal resolution
If higher temporal resolution is needed a multi-beat acquisition might be necessary.
By splitting the entire volume into several sub volumes (by keeping the selected
volume size from single beat), higher volume rates can be achieved, without losing
spatial resolution
Adjust your single beat acquisition as describe above.
Then select the number of heart cycles you want to collect.
Possible settings are 2-3-4 or 6 cycles.
Note
The volume will be split in several sub volumes according to the selection of
the number of heart cycles. This will increase the frame rate by 2 times,
3times, 4 or 6 times according to the number of cycles.
Press Multi Beat to start the stitched acquisition.
Press Image store whenever satisfied with the dataset.