idi Aspect100RT Operator's Manual

idi Aspect100RT Operator's Manual

Mobile fluoroscopic table

Advertisement

Quick Links

Aspect100RT
Mobile Fluoroscopic Table
Operator Manual
Made in the USA

Advertisement

Table of Contents
loading

Summary of Contents for idi Aspect100RT

  • Page 1 Aspect100RT Mobile Fluoroscopic Table Operator Manual Made in the USA...
  • Page 2: Table Of Contents

    Aspect 100RT Operator Manual Table of Contents PRODUCT DATA ............................1 INTRODUCTION ............................2 CUSTOMER SUPPORT ........................... 3 SYMBOLS AND DATA PLATE ......................... 4 SPECIFICATIONS ............................3 ELECTROMAGNETIC COMPATIBILITY STATEMENT ..................4 SETUP ................................5 SAFETY HAZARDS............................6 SAFETY HAZARD ALERTS ..........................7 SAFETY INSTRUCTIONS ..........................
  • Page 3: Product Data

    Aspect 100RT Operator Manual PRODUCT DATA • Low attenuation carbon fiber tabletop • Low attenuation patient pad • Standard Tabletop: 61cm x 203cm (24in x 80in) • Trendelenburg: ±15° • Height range: 77cm to 107cm (30.4in to 42.2in) • Patient capacity: 204kg (450lbs) •...
  • Page 4: Introduction

    Aspect 100RT Operator Manual INTRODUCTION Overview This manual pertains to the specified product only. It is intended for qualified medical personnel who have been trained in the use of medical equipment. It is not designed to replace or substitute for certified training in the application of this equipment. Functional capabilities and operation of the equipment described herein which can be employed in a variety of diagnostic, therapeutic, and surgical applications.
  • Page 5: Customer Support

    For technical assistance, call IDI at (978) 829-0009 or send fax to (978) 829-0027. Be prepared to give the complete model and serial number found on the dataplate on the table base at the time of contact.
  • Page 6: Symbols And Data Plate

    Aspect 100RT Operator Manual SYMBOLS AND DATA PLATE Intertek Testing Service The system was tested and found to be in compliance with the requirements of all relevant directives and standards in effect within the European Union at the time of manufacture. Attention! Consult accompanying documents.
  • Page 7 Aspect 100RT Operator Manual Recycle Some of the material can be recycled rather than discarded. Not for Patient Transport. The table should not be relocated with a patient on it nor should it be used to move a patient. Do Not Sit This label is found on parts that are not designed to support the sitting weight of a patient.
  • Page 8: Specifications

    Aspect 100RT Operator Manual SPECIFICATIONS Mode of Operation: For continuous operation with short time loading Duty Cycle: 10% (6 min/hr) Type of Equipment: Class I Type B (as defined by IEC 60601-1, UL 60601-1, EN 60601-1, CSA 601.1, IEC 60601-2-32:1994 and IEC 60601-2-46) Type B protection against electrical shock Complies with FDA standards Electrical:...
  • Page 9: Electromagnetic Compatibility Statement

    Aspect 100RT Operator Manual ELECTROMAGNETIC COMPATIBILITY STATEMENT This equipment may generate and use radio frequency energy. The equipment must be installed and used according to the manufacturer’s instructions in order to avoid radio frequency interference. If this equipment generates or receives interference: •...
  • Page 10: Setup

    Aspect 100RT Operator Manual SETUP Unpack No special handling is required for unpacking this equipment at the site. Conventional shipping materials are used. To remove the table from the shipping pallet, roll the table down a temporary ramp or lift off pallet with a forklift.
  • Page 11: Safety Hazards

    Comments and questions regarding safety should be addressed to: Customer Support Image Diagnostics, Inc. 310 Authority Drive Fitchburg, MA 01420 Or call IDI at (978) 829-0009 or send fax to (978) 829-0027. Or call toll-free at (877) 304-5434. L100-3061 REV A www.imagediagnostics.com...
  • Page 12: Safety Hazard Alerts

    Aspect 100RT Operator Manual SAFETY HAZARD ALERTS Alert Circumstances for Use DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury or...
  • Page 13: Safety Instructions

    Aspect 100RT Operator Manual SAFETY INSTRUCTIONS This symbol, when found marked on the equipment means: ”Attention, consult accompanying documentation.” This manual should be accessible to all personnel installing, operating, or servicing this equipment. Only a qualified technician may install or maintain this equipment. Only qualified persons may operate this equipment.
  • Page 14 Aspect 100RT Operator Manual CAUTION Do not use table for patient transport. CAUTION During operation, listen for any unusual sounds and watch for uneven operation. CAUTION Ensure that the tabletop does not contact other equipment as it moves. CAUTION Do not sit on the table top at the head of the table. CAUTION Do not place or store any containers or large items underneath the tabletop.
  • Page 15: Accessories

    Aspect 100RT Operator Manual ACCESSORIES • CRESCENT FACE CUSHION (supplied with some models) Use this pad for patient comfort when the patient will be positioned face down. Place a disposable cover over the pad. LATEX Crescent Face Cushion Head Pad Part Number C000-0597 Disposable Covers Part Number C000-0598 (box of 50)
  • Page 16 Aspect 100RT Operator Manual • PATIENT RESTRAINT STRAPS SLIDE THE RESTRAINT STRAP OVER THE END OF THE TABLETOP WITH THE SMALL RETAINING STRIP OVER THE TOP SURFACE OF THE TABLETOP. PHOTO OF STRAP ARRANGEMENT: NOTE: Strap configuration shown above is recommended, however, patient restraint is a case-by-case condition.
  • Page 17 Aspect 100RT Operator Manual • CLAMP-ON ACCESSORY RAIL AND ARMBOARD (OPTION) WARNING! 25lbs (11kg) Maximum Load To install clamp to tabletop: - Turn knob counterclockwise (as viewed from above) to open clamp enough to slip onto tabletop - Turn knob clockwise to tighten clamp to tabletop A –...
  • Page 18 Aspect 100RT Operator Manual • RADIATION SHIELDS, SADDLE STYLE (OPTIONAL) • 24X80 Straight Tabletop Shield Part Number: A610-096 INSTALLATION INSTRUCTIONS FOR 24” SADDLE STYLE RADIATION SCATTER SHIELD (0.5mm PbEq) Note: Individual shields have hook fasteners that are engaged into the loop fasteners of the saddle.
  • Page 19 Aspect 100RT Operator Manual • 22X80 Facial Cutout Tabletop Shield Part Number: A610-092 INSTALLATION INSTRUCTIONS FOR 22” SADDLE STYLE RADIATION SCATTER SHIELD (0.5mm PbEq) Note: Individual shields have hook fasteners that are engaged into the loop fasteners of the saddle. Overlap shields as shown. Step A: Place the saddle with the loop fastener side “up”...
  • Page 20 Test procedure and schedule to be the responsibility of the appropriate department of the facility where used. IDI imaging tables are designed to provide excellent fluoroscopic imaging access with mobile C-arms. Whenever the table is used for procedures involving the...
  • Page 21: Operator Controls

    Aspect 100RT Operator Manual OPERATOR CONTROLS HAND CONTROL CAUTION Always check that the patient is secure before activating any table motion To move tabletop, press appropriate button and hold button depressed until desired position is reached. MOVES TOP UP MOVES TOP DOWN REVERSE TRENDELENBURG TRENDELENBURG...
  • Page 22 Aspect 100RT Operator Manual If, over time, an actuator is not returning to its programmed position or an actuator or controller has been replaced, perform the following steps: Operate each actuator to the retracted position, holding the button in for about 2 to 5 seconds after the actuator stops retracting.
  • Page 23 Marking is the same as the hand control. Foot Control and Disposable Cover Foot Control Part Number A100-1960 Contact IDI Customer Service for individual components of foot control. Foot Control Disposable Cover Not made with natural rubber latex...
  • Page 24: Emergency Stop Switch

    Aspect 100RT Operator Manual EMERGENCY STOP SWITCH • The red emergency stop button is located on top of the control box • Pressing this button down will stop all motor movement • In the event that the hand or foot control malfunctions, the emergency stop control prohibits motion by removing power from all system components •...
  • Page 25: Electric System

    Aspect 100RT Operator Manual ELECTRIC SYSTEM TRENDELENBURG ELEVATING ROLL ACTUATOR ACTUATOR ACTUATOR HAND CONTROL FOOT CONTROL EMERGENCY STOP SWITCH JUNCTION BOX AC POWER CORD 12V BATTERIES INSIDE CONTROLLER CONTROLLER CONNECTIONS CONNECT OPTIONAL FOOT PORT DEVICE CONTROL HERE TREND ACTUATOR ELEVATING ACTUATOR CONNECT HAND CONTROL HERE ROLL ACTUATOR...
  • Page 26: Battery Function (Motorized Tables)

    Aspect 100RT Operator Manual BATTERY FUNCTION (MOTORIZED TABLES) When AC power is applied, the battery will begin charging until the battery reaches full capacity. When it does, the charging will taper off. If the power goes out, the table will automatically switch to battery back-up mode.
  • Page 27 Aspect 100RT Operator Manual APPROVED AND TESTED GENERAL PURPOSE CLEANERS • Simple Green ™ cleaner CLEANING STEPS 1. Move the tabletop into the horizontal position. 2. Lower the tabletop to the lowest position. 3. Power cord, hand control and foot control cords must be plugged in to protect the sockets. 4.
  • Page 28 Aspect 100RT Operator Manual CLEANING THE OPTIONAL RADIATION SHIELD CAUTION Observe cleaning product cautions found on product labels. After each procedure, the radiation shield should be carefully cleaned. CLEANING STEPS Remove the panel sections and lay flat before using the recommended cleaner in an adequately ventilated area.
  • Page 29: Maintenance

    Aspect 100RT Operator Manual MAINTENANCE ”Qualified Technician” • All maintenance procedures should be done by an experienced technician with demonstrated knowledge and skills (electrical and mechanical) relative to this type of equipment. • This individual must have access to this manual and the proper tools. ROUTINE MAINTENANCE •...
  • Page 30: User Replacement Items

    Aspect 100RT Operator Manual USER REPLACEMENT ITEMS These are easily replaced by the operator. See also previous sections for operator hand and foot controls, armboard pads and protective covers. • Standard Tabletop Pad Pads are held in place by hook and loop fasteners. Simply pull pad up gently to remove. New replacement pads are supplied with new self-adhesive fasteners installed and new mating pieces for the tabletop.
  • Page 31: Trouble-Shooting

    Aspect 100RT Operator Manual • Tabletop Pad (Optional) Tabletop Pad 2” thick Optional Tabletop Pad Part Number X100-1757 TROUBLE-SHOOTING SYMPTOM POSSIBLE CAUSE ACTION No motor sound Power cord not plugged in Connect power cord And/or movement No power at the receptacle Check appropriate circuit breaker A missing or loose connection...
  • Page 32: Disposal Of Components

    Operator Manual DISPOSAL OF COMPONENTS IDI medical tables are made up of mostly steel, copper and aluminum parts which are easily recycled. It is recommended that some components be disassembled before disposal for recycling. The table below lists components typically found in IDI products but varies with model and options.
  • Page 33: Warranty

    Aspect 100RT Operator Manual WARRANTY Warranty details for IDI Products can be obtained directly from Image Diagnostics, Inc. Image Diagnostics, Inc. 310 Authority Drive Fitchburg, MA 01420 Or call IDI at (978) 829-0009 or Toll Free (877) 304-5434. The fax number for IDI is (978) 829-0027.

Table of Contents