Result screen
Message
The combined data is not saved. Continue
operation?
The data has been properly copied
The remained memory of the drive is less
than 5%
Database, Network error
Message
Network connection error
Network drive has been shut down.
Could not connect to the network
If you proceed there will be no link to the
VK (any) database.
Cannot open the database
If the examination continues, the results
will be saved in the local folder separately
from previous data.
Cause and Remedies
●
An attempt has been made to change over the screen to another
without saving the result of data combination. To save the
combined data, press [Cancel] and [Save] in this order. To not
save, press [OK] and proceed with the operation.
●
Data read has been normally completed from Kowa automatic
perimeter "AP–6000," "AP–5000," or "AP–3000" to "AP–7000."
●
This message appears when the remaining free space of the data
save drive is less than 5% at the time of software startup or
examination result save.
Cause and Remedies
●
The database or network is abnormal in some way. Select [OK]
to save the data locally. If the database or network is restored
when the instrument is started next time, the locally saved data
can be coupled to the conventional data.
●
Select [Cancel] to end KOWA AP–7000.
10 Troubleshooting
149
Need help?
Do you have a question about the AP-7000 and is the answer not in the manual?