Summary of Contents for PCB Piezotronics ICP 105C02
Page 1
Model 105C02 ICP® Pressure Sensor Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Assistance is needed to safely operate equipment PCB Piezotronics is an ISO-9001 certified company whose Damage is visible or suspected calibration services are accredited by A2LA to ISO/IEC Equipment fails or malfunctions 17025, with full traceability to SI through N.I.S.T.
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
OPERATION MANUAL FOR MINIATURE ICP PRESSURE SENSOR 105A AND 105C SERIES INTRODUCTION MODEL 105A This model series is designed to satisfy the need for This is the smallest of the three styles and it is measurement of dynamic pressures where space is supplied in only the 100 psi range.
OPERATION MANUAL FOR MINIATURE ICP PRESSURE SENSOR 105A AND 105C SERIES Because of its one-piece machined diaphragm MODELS 105C03, C13, 23, C33, construction, the 105A series is sensitive over its entire frontal area. Use extreme care when preparing recessed installation port to ensure that no particles, This style is the largest and is supplied in five models burrs or other residue are entrapped in the .010”...
Page 8
OPERATION MANUAL FOR MINIATURE ICP PRESSURE SENSOR 105A AND 105C SERIES The natural frequency and approximate fastest Model 480A pressure step rise time for various length passages is 002A Cable Signal Conditioner shown in the following chart. (Medium, air at 25C). PASSAGE PASSAGE APPROX.
Page 9
The ground return is provided by a metal ribbon (.003 x .020) welded to the Model 060A04 clamp nut. ICP is a registered trademark of PCB Piezotronics Drawing Number: 21106 Revision: NR...