Page 1
eSelectM 2 PH (RX) -CL OPERATING INSTRUCTIONS AND MAINTENANCE UNI EN ISO 9001-2015...
Page 2
CERTIFICATE OF CONFORMITY ETATRON D.S. S.p.A. He ad O ff i ce Via dei Ranuncoli, 53 – 00134 ROMA • ITALY Tel. +39 06 93 49 891 • Fax +39 06 93 43 924 C.C.I.A.A. 535990 - Trib. di Velletri 5170/85 • Cod. Fisc. 06632160583 P.
W arni ngs Read the warnings below carefully. They provide important information regarding safe installation, use and maintenance. Store this manual with the utmost care for future reference. The device is built to a professional standard. Its durability and electrical and mechanical reliability will be more efficient if it is used properly and maintenance is carried out on a regular basis.
� In case of failure and/or malfunctioning of the device, switch it off and do not tamper with it. For any repairs, please contact our service centres and request the use of original spare parts. Failure to comply with the above can jeopardise the safety of the pump. �...
M ai n addi ti onal functi ons Independent settings to start up metering pumps or 3 setpoint ON-OFF peristaltic dispensers in constant mode or equipment with ON-OFF mode Setpoint Adjusts the setpoint value (ON-OFF mode) Selects a measurement range around the setpoint value, Hysteresis locking the output relays (ON-OFF) Relay outputs...
Page 9
▪ 0-2 Cl ppm = Resolution 0.01 Cl ppm Hysteresis/PWM point = 0.050 Cl ppm Chlorine measurement range Cl ppm (free or total) ▪ 0-20 Cl ppm = Resolution 0.10 Cl ppm Hysteresis/PWM point = 0.50 Cl ppm Membrane type sensors: ▪...
INSTALLATION Install the instrument in a dry place, away from heat sources at a maximum room temperature of 40°C. Comply with standards in force in the different countries regarding electrical installation (Fig. 2). If the power cord does not have a plug, the device must be connected to the mains by means of an omnipolar disconnecting switch with at least 3 mm between the contacts.
It is recommended to use voltage protections, check the earthing system and, when other equipment is connected in parallel, use a transducer. Furthermore, ETATRON recommends installing a UPS (genset) to assure continuity thus ensuring no data are lost.
NO MEASUREMENT CONN., this means there is a problem between the two boards, in that case check the flat electric cable and immediately contact the ETATRON service. Other status icons shown at the end of the row ***...
OPERATING FUNCTIONS O N -OFF m ode The unit has an ON-OFF mode which switches on (or off if the reverse mode is ON) the output relays to control Constant / ON-OFF metering pumps, peristaltic pumps or other ON-OFF equipment. DI RECT / REVERSE di recti on The setpoint relays are factory set as follows: Setpoint 1 PH: ACID mode, the output is active when the measured value is higher than the selected setpoint, the connected pump...
The chlorine measurement CL works according to the corresponding values relating to the selected chlorine range indicating the value in Cl ppm. ADVANTAGES: the proportional function is more accurate than the ON-OFF mode. DISADVANTAGES: the user needs to be a professional in order to select the most accurate settings to assure the best results. ANALOGUE OUTPUTS I N CURRENT 4 -20 m A1 / 4 -20 m A2 The instrument features 2 outputs with signal in current in mA.
To go forward quickly, press and hold one of the ◄► buttons The software version is shown when the instrument is on the lower part of the display. ETATRON The software is subject to revisions without notice. Model eSelec M2 The instrument is prepared for measuring and is then ready to operate.
Sel ect M EASUREM ENT TYPE PH or RX The instrument of the eSelect M2 series makes it possible to select the type of measurement to be used, i.e. PH or RX (Redox) in the same unit. The measurement parameter may be changed from PH to RX and the display and resolutions of the measurement change accordingly.
PROGRAMM I NG M ENU The instrument lets you choose between two programming modes: BASIC programming: simplified mode for non-professional operators. � EXPERT programming: complete programming that includes functions for more accurate control and results. � After selecting the type of programming, the Menus and sub-menus change accordingly. To help the operator choose the correct menu, the main differences between the two menus are set out below: Menu Mode Setpoi nt 1 -2 -3 <Set poi nt val ue - M et eri ng - ON/ OFF or Propo rt i onal m ode –...
MENU SETPOINT 1 (pH) MENU SETPOINT 1 (RX) Metering Metering Setpoint 1 is set for operations in Acid Direc t Setpoint 1 is set for operations in ACID mode, the output is active � MENU BASE � � MENU BASE � DIRECT mode: if the measured when the measured value is higher...
MENU SETPOINT 2 (pH) MENU SETPOINT 2 (RX) Setpoint Value Setpoint Value The setpoints activate the output relays for the metering pumps or other devices until the setpoint is reached. 7.00 pH 200 mV � MENU BASE � � MENU BASE � MENU SETPOINT 2 (pH) MENU SETPOINT 2 (RX) Metering...
Direct mode: the output is active when the measured value is lower than MENU SETPOINT 3 (Cl) MENU SETPOINT 3 (Cl) the one selected in the setpoint, the metering pump injects Chlorine. Metering Metering Reverse mode: the output is active also the measured value is higher than Reverse Direc t the one selected in the setpoint, the metering pump injects a Chlorine...
We suggest using new or however uncontaminated buffer solutions. As for the length of the pH electrode we advise max 9 m: contact ETATRON for longer distances. The mV value at the bottom of the screen shows the electrode’s efficiency (according to the manufacturer’s specifications) CALIBRAT.
MEASUREMENT Zero Wait 5/10 minutes, press ◄►until 0.00 is displayed, press O K / M E N U . 0 Cl ppm 420.1 mV Another simple method for ZERO calibration is to perform the instrument’s electronic Zero as follows: DISCONNECT THE SENSOR, short- circuit the M-0 pins of the internal terminal board, press and hold ▲...
SETTINGS > BASIC MENU Menu Selection Settings BASIC programming: simplified mode for Se�ngs Se�ngs non-professional operators Menu Mode Menu Mode BASIC EXPERT EXPERT Programming: complete programming that � MENU BASE � � MENU BASE � includes functions for more accurate control and results. After selecting, e.g.
* By selecting M e a s u r e m e n t T y p e *Selecting the setting C h l o r i n e S c a l e , S E T P O I N T 3 S E T P O I N T 1 , S E T P O I N T 2 and and C A L I B R A T I O N will be followed by values in the...
Page 26
MENU SETPOINT 1 (pH) MENU SETPOINT 1 (RX) The unit has an ON-OFF mode which switches on or off (if the reverse mode Mode Mode is ON) The output relays of Constant / ON-OFF metering pumps or other ON/OFF ON/OF F ON-OFF equipment.
Page 27
FUNCTION NOT AVAILABLE WITH PROPORTIONAL MODE MENU SETPOINT 1 (pH) MENU SETPOINT 1 (RX) Hysteresis is useful during operations to adjust the setpoints in ON-OFF Hysteresis Hysteresis mode and is used to enable and disable the output relay when the selected 0.05 pH 10 mV hysteresis has been achieved.
SETPOINT 2 PH AND SETPOINT 2 RX > EXPERT MENU SETPOINT2 PH SETPOINT2 RX Menu Selection Menu Selection The two programming stages are below: when Measurement Type pH is Setpoint 2 (pH) Setpoint 2 (RX) selected, refer to the left column; when selecting RX refer to the right column.
Page 29
MENU SETPOINT 2 (pH) MIN Active Time: programmable. Defines the minimum time for which PWM is active; prevails over the selected Min.Active Time settings. The pulse modulation function is adjusted with 3 functions according to the following formula: Active Time in accordance with the selected formula = (measured value - setpoint) / (period width * cycle time). If the 5 sec result of the formula is <...
SETPOINT 3 CL CHLORINE > EXPERT MENU The chlorine instrument is preset by default for a 0-2 ppm range. By selecting the required chlorine range, the Menu Selection values displayed change accordingly. The setpoints will activate the output relays for the metering pumps or other devices once the setpoint level is selected.
Page 31
MENU SETPOINT 3 (Cl) MENU SETPOINT 3 (Cl) MENU SETPOINT 3 (Cl) MENU SETPOINT 3 (Cl) Hysteresis Hysteresis Hysteresis Hysteresis 0.005 Cl ppm 0.05 Cl ppm 0 .0 5 Cl ppm (Tot) CLC 0.05 C l ppm � EXPERT MENU � �...
MENU SETPOINT 3 (Cl) Start-up Delay The start-up delay stops the output relays when the unit is switched on, thus allowing the sensor to polarise assuring correct measurements (programmable). 00.00 h:m � EXPERT MENU � MENU SETPOINT 3 (Cl) MENU SETPOINT 3 (Cl) ONLY WHEN USING METERING PUMPS Priming...
4-20m A1 OUTPUT PH OR RX > M ETERI NG FUNCTION ON SETPOI NT The 4-20mA1 analogue output follows the options previously selected in Measurement Type i.e. pH or RX mA1 pH mA1 RX Menu Selection Menu Selection 4-20mA1 (pH) 4-20mA1 (RX) SETTING 4-20mA1 (pH) SETTING 4-20mA1 (RX)
4-20m A2 OUTPUT CHLORI NE > Rem ote Devi ces Functi on The analogue 4-20mA2 output of the instrument follows the setting selected in C H L O R I N E S C A L E Menu Selection Please note that the displayed value is automatically adjusted according to the Chlorine Scale range previously 4-20mA2 (Cl) entered.
AUX 1 Program 01 AUX 2 Program 01 Active time (m:s) 01 : 00 Active time (m:s) 01 : 00 Start time (h:s) 10 : 30 Start time (h:s) 10 : 30 The settings of the program entail many steps, which may be Enabl.
Page 37
SETTINGS SETTINGS By enabling the pH priority function, the instrument is able to first reach the pH priority pH priority pH setpoint, then switch to product metering operations (oxidising or reducing agent), thus assuring measurement stability � EXPERT MENU � �...
We suggest using new buffer solutions or still in good conditions (not old) or a reliable test kit. As for the lengths of the pH electrode cable we advise max 9 m: contact ETATRON for longer distances. CHLORI NE el ectrodes - Cl eani ng and M ai ntenance For cleaning and maintenance of the chlorine probes, refer to the probe’s manual.
Furthermore, ETATRON recommends installing a UPS (genset) to assure continuity thus ensuring no data are lost. A system that is set up without following the proper electrical design rules, without an earthing system, with frequent ON/OFF operations, might directly undermine the printed circuit.