Tube Holder Configurations; 12-Place Horizontal Standard / Performance Plus Rotors - Horizon Fitness 853VES Operator's Manual

Laboratory centrifuge
Table of Contents

Advertisement

Tube Holder Configurations

The next two pages will describe proper balancing and tube loading. Use the
appropriate section for the rotor installed in your centrifuge. Improper loading can
result in damage to the centrifuge or test tube samples.
Twelve (12) Place Horizontal Standard and Performance Plus Rotors
This rotor is designed to hold up to twelve (12) tube holders. It comes with two tube
holder types. The purple holders are designed to hold tubes up to 10 mL (up to 17 mm x
100 mm). The blue holders are designed to hold tubes up to 5 mL (up to 17 mm x 75 mm).
Your centrifuge must contain a balanced load in order to work properly. To ensure
that the load is balanced, keep these rules in mind when inserting test tube samples
into the 12 place horizontal rotor:
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded with equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate with the
unpaired one.
Directions:
A.
1. Compare the tube to be spun with the boxes shown here.
2. Find the box that most closely matches the tube's length.
NOTE: The tube length with its stopper or cap must be no longer
than the chosen box.
3. Match the letter from the chosen box with one of the configurations
below. For Example: A tube is found to be as long as box B.
B.
Accordingly, we can use a 100 mm tube holder with a 1525 cushion
or a 75 mm tube holder with no cushion, (configurations B1 or B2).
A.
PURPLE
C.
08
100mm
Tube Holder
B1.
B2.
PURPLE
BLUE
Performance Plus (optional)
C.
75mm Tube
BLUE
Holder
with 1525
cushion
Standard

Advertisement

Table of Contents
loading

Table of Contents