SimpleDist:
Operation and Instruction Manual
Ammonia Nitrogen Distillation with the SimpleDist® Continued:
5.11 Turn air/vacuum valves to the off position.
5.12 Bring the volume in the collection trap up to the 25 mL line with DI water (a graduated cylinder, volumetric flask, or
other volumetric glassware may be used for volume measurements).
5.13 The distillate in the collection trap is now ready for analysis. Pour contents into an appropriate container. When
pouring, position the vent tube of the collection trap up to prevent trapping of air.
5.14 Bring the volume in the collection trap up to the 25 mL line with DI water (a graduated cylinder, volumetric flask, or
other volumetric glassware may be used for volume measurements).
+1.800.343.5319 or +1.843.881.6560 envexp.com
System Operation
Environmental Express
• 17