Environmental Express 35419-52 Instruction Manual

Environmental Express 35419-52 Instruction Manual

Ph electrode

Advertisement

Quick Links

This manual describes the operation of the pH electrode,
35419-52. Be sure to read this manual carefully, before
using the electrode.

Conformable standards

This equipment conforms to the following standards:
RoHS: EN IEC 63000
9. Monitoring and control instruments
including
control instruments
RoHS: BS EN IEC 63000
9. Monitoring and control instruments
including
control instruments
Authorised representative in EU and UK
 Cole-Parmer GmbH
Futtererstraße 16
Wertheim 97877
Germany
Telephone: +49 9377 9203-0
Fax: +49 9377 1388
E-Mail: sales@coleparmer.de
Website: www.coleparmer.co.de
 COLE-PARMER INSTRUMENT COMPANY UK
9 Orion Court
Ambuscade Road
St. Neots PE19 8YX
United Kingdom
Telephone: +44 (0)1480 272279
Free Call: +44 (0)1480 272279
E-Mail: sales@coleparmer.co.uk
Website: coleparmer.co.uk

Caution on use

 Safety Precautions
CAUTION
Chemical solution
The internal solution in the electrode uses potassium chloride
(3.33 mol/L-KCI). If the internal material comes in contact with the
hands or skin, wash immediately with water. If the internal solution
comes in contact with the eyes, flush with a large amount of
running water, then seek medical advice.
Glass fragments
Glass fragments can cause injury.
The outer tube and tip of the electrode are made from glass.
Be careful not to break them.
 Points of concern
 Do not shock the electrode.
 Do not measure chapped or cracked electrode. Replace
such electrode with a new one.
 Do not allow the connector to come in contact with water
or unclean hands. If not, accurate measurement cannot
be performed.
 Always use the 3.33 mol/L-KCl solution internal solution
for reference electrode.
Environmental Express, Inc.
2345A Charleston Regional Pkwy Charleston, SC 29492
E-mail: info@envexp.com
Phone number: 800-343-5319
pH Electrode 35419-52
Instruction Manual
industrial
monitoring
and
industrial
monitoring
and
CODE:M0042153200915187GZ0000649868
 If the responsive glass membrane or liquid junction is
very dirty and cannot be washed clean using pure water
(or deionized water), clean it depending on dirt
condition. (Refer to " Maintenance " (page 3).)
 Do not use this electrode at acidic or alkaline
concentrations above 0.1 mol/L. Using the electrode
under these conditions may harm performance and
shorten the life of the electrode.
 Do not measure the solutions which contain hydrofluoric
acid, which may corrode the glass.
 Do not suddenly immerse the electrode in samples
where the temperature variation is 50°C or more.
 Silver chloride can be detected inside the responsive
glass membrane. However, it will not cause any
performance problem.
 Do not use the electrode in any place where usable
temperature is out of the specified range.
 White crystal can adhere to the internal solution filler
port or protective cap after storage of the electrode.
However, it will not cause any performance problem.

Packaged contents

Name
Electrode 35419-52
Instruction manual (this book) 1 booklet
Specifications and parts description
 Part name
Internal solution
filler port  (Slide cap)
Cap
Connector cover
Temp. connector
Lead
For any questions regarding this product, please contact your local
agency, or inquire from the following website.
https://www.EnvExp.com
July,2021
Q'ty
1 pc
Responsive glass
membrane
Liquid
Stem
junction
Protective cap
Electrode connector

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 35419-52 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Environmental Express 35419-52

  • Page 1 This manual describes the operation of the pH electrode,  If the responsive glass membrane or liquid junction is 35419-52. Be sure to read this manual carefully, before very dirty and cannot be washed clean using pure water using the electrode.
  • Page 2  Specifications Preparation  Electrode model 35419-52 Note Measurement range pH 0 to 14 White crystal (KCl) can adhere to the internal solution filler port or protective cap. However, it will not cause any performance Usable temperature range 0C to 100C problem.
  • Page 3 Measurement (calibration) Maintenance   1. Open the internal solution filler port. Note organic solvents (such acetone tetrahydrofuran) or acid (above 1 mol/L) or alkaline (above 0.1 mol/L) cleaning solutions. Using such chemicals may damage the body of the electrode or cause a decline in performance. ...
  • Page 4 Storage  1. Wash the electrode well with pure water (or deionized water) to remove sample completely and wipe it with filter paper or tissue paper. 2. Close the internal solution filler port. 3. Wash the inside of the protective cap with pure water (or deionized water), then add pure water (or deionized water) until the sponge gets wet.

Table of Contents