6.2.4
Delete data menu description
Go to the second menu. Press the Del key in the second menu to delete all the saved sampling
data. Please operate with caution! If you do not want to delete the data, you can press the HOME
button to return to the main interface.
6.2.5
Flow calibration (password required) interface description
This menu is intended for trained personnel only.
In the flow calibration menu, press the corresponding password box to select it and then enter the
corresponding password by using the up and down buttons to the right. When the password is
correct, press the CONFIRM key to enter the flow calibration menu, as shown in the second figure.
When entering the flow calibration menu, the sampling pump is automatically started. Press the
CALIBRATION button to select the flow parameters. After selection, the corresponding values
can be modified by means of the up and down buttons. When the flow setting reaches the
standard value, press the SAVE button to save the current flow calibration value and then return
to the main menu. If you press the home page, you will not save the modified parameters and
return to the main interface.
Setting range of calibration parameters: 0-499.