Summary of Contents for Centurion Scientific PrO-Road

  • Page 1 Scientific Limited Instruction manual...
  • Page 2 Read before using Check the Centrifuge and the Packaging for any shipping damage. • Inform the shipping company and Centurion Scientific immediately Check the order for completeness, if not contact Centurion Scientific • Location The Centrifuge should only be operated indoors.
  • Page 3 Fixing of rotor Fixing for micro and fixed angle rotors BRK5401. BRK54XXseries. BRK53XX series. BRK52XX series Locate 3 pins of fixing, into 3 holes at the top of the rotor Push down firmly into holes. Lift and fit rotor onto the motor shaft. Fit the Screw and spring washer, screw down tight with the supplied tool.
  • Page 4: Very Important

    Rotor balancing. Very Important Balancing of tubes for Micro and Fixed angle rotors Tare (zero) your scales first Weigh a tube , note its weight Weigh another and match its weight to within 0.1 gram...
  • Page 5 Tube placement Fixed angle rotors Small fixed angle Rotors ✘ ✔ ✘ ALWAYS load your tubes opposite to each other ALWAYS weigh samples to within 0.1 Gram. For best separations. (see page 4) CHECK The rotor screw fixing tightness regularly ....................................
  • Page 6 Instructions ambient. RTL7 & RTL5 Prime and Dual models only. Ambient Rotor Speed x 10 Time (Mins) RCF O ACCELL DECELL O RPM Memory Accel Decell Memory Memory Start Pulse Open Stop (Sec) Store Function Rcf/Rpm Recall START POINT. A. Selection of your fitted Rotor First set the rotor 0-99.
  • Page 7 Instructions ambient B. Setting of Speed or Rcf(G) and acceleration or Decelleration (braking) Speed x 10 RCF O 0 4 0 0. Speed shown (Example). The speed shown is x 10 so is 4000 Rpm ACCELL DECELL O RPM Accel Decell Open Function...
  • Page 8 Instructions ambient D. Programming, 100 are available from 0 – 99 Set all parameters as previous details. Rotor, . Speed or Rcf(G), Accel, Decell. Time Memory Memory Memory Store Function Recall After setting all your required parameters. Press Function and select memory to your desired number by pressing Memory buttons Once done whilst pressing...
  • Page 9 Temperature controlled RTL7R RTL5R Prime and Dual models only.Refrigerated Temp Speed x 10 Time (Mins) RCF O ACCELL DECELL O RPM Memory Rotor Rotor Accel Decell Memory Memory Start Pulse Open Stop (Sec) Store Function Rcf/Rpm Recall ........................................START POINT. A.
  • Page 10 Temperature controlled C. Setting of Speed or Rcf(G) and acceleration or Decelleration (braking) Speed x 10 RCF O Speed shown (Example). The speed shown is x 10 so is 4000 Rpm O RPM Accel Decell Open Function Rcf/Rpm Select if you require Speed or Rcf (G) RCF (G) is the actual force that does the work for your analysis.
  • Page 11 Temperature controlled E. Programming, 10 are available from 0 – 9 Set all parameters as previous details. Rotor, . Speed or Rcf(G), Accel, Decell. Time Memory Memory Memory Store Function Recall After setting all your required parameters. Press Function and select memory to your desired number by pressing Memory buttons Once done whilst pressing...
  • Page 12: Emergency Lid Release

    Emergency Lid release Mechanical Emergency Door release During a power failure, you will not be able to open the Centrifuge lid via the Display button Mechanical override is provided to allow sample recovery However this is for Emergency use only This should Not be used for normal use.
  • Page 13 Do not site the unit in sunlight Always ensure that lids and accessories are correctly tightened before running the unit. Use only official Centurion Scientific Ltd accessories with your centrifuge. It is recommended that regular maintenance and cleaning of the unit and accessories is carried out.
  • Page 14 Care & maintenance. Rotor Always check rotors for cracks and damage, if found replace your rotor. DO NOT USE as dangerous Autoclave for the maximum number of times shown in the sales literature. In brackets( 20) Rotors have a Seven year lifespan and must be replaced . DO NOT USE beyond their Lifespan It is recommended that regular maintenance and cleaning of the rotor and accessories is carried out.
  • Page 15 CERTIFICATE Road. RTL7, RTL7R & RTL5, RTL5R Company Representative Scientific Limited The Old Stables, Church Farm, Stoughton, Chichester, West Sussex. PO18 9JL. United Kingdom.
  • Page 16 PrO-Care Service, Warranty & Maintenance pakages The lifespan of a centrifuge depends upon how well it is looked after and maintained. Many of our centrifuges have lived a life of over 15 years. This is not only due to our stringent manufacturing process and high quality components but also to a yearly servicing package.