Summary of Contents for Centurion Scientific C2006
Page 2
• No changes should be made to the mechanical or electrical components without prior consultation with and the written permission of Centurion Scientific. • The centrifuge has not been manufactured from inert materials or to be explosion proof.
Page 3
Should any problems of questions arise during the installation of the instrument please refer to Centurion Scientific for advice. During installation the record for the individual centrifuge (to be found with our manual) must be completed.
Page 4
WARNING : If any liquid has entered the main body of the centrifuge it MUST NOT be used. Contact the supplier or Centurion Scientific Ltd immediately for advice. ROTOR • Remove the locking screw with the key provided. • Remove the flat washer •...
Page 5
Above shows 12000 (Rpm see dot bottom right) IE: 1200 x 10 and 30 minutes of time To set speed in Rpm To set time Press up or down arrows below speed display. Press up or down arrows below time display. To set in Rcf (G) Press Function button (keep held down) then press Lid open button.
Page 6
OPEn - You may now open the Lid. UUAIt - Wait 5 seconds press Lid open tAcho? - Check that your tubes are not too long and fouling the lid or anything else is stopping rotation of the rotor. An imbalance has occurred Please check the tubes are diagonally opposite and of the same weight.