Installation and
Operation Manual
SCO58
SCO40
SCO31
Previously designated:
2460, 2440, 2428

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SCO58 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shel lab SCO58

  • Page 1 AIR-JACKETED CO INCUBATORS 115 Volts Installation and Operation Manual SCO58 SCO40 SCO31 Previously designated: 2460, 2440, 2428...
  • Page 2 INCUBATORS 115 Voltage Installation and Operation Manual Part Number (Manual): 4861710 Revision: December 24, 2014 Pictured on front cover: SCO58 SCO40 SCO31 These units are compliant with the following standards for use within an ambient air pressure range of 22.14 – 31.3 inHg (75 – 106 kPa), with no flammable, volatile, or combustible materials being heated.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS INTRODUCTION ............................4 General Safety Considerations ......................... 4 Engineering Improvements ........................5 Contacting Assistance ..........................5 RECEIVING YOUR INCUBATOR ........................ 6 Inspecting the Shipment ..........................6 Recording Data Plate Information ......................10 INSTALLATION ............................11 Ambient Conditions ..........................11 Location ..............................
  • Page 4: Introduction

    This can result in injury or death. Your Shel Lab CO incubator and its recommended accessories are designed and tested to meet strict safety requirements. Only use this equipment for its intended application. Any alterations or modifications void the warranty.
  • Page 5: Engineering Improvements

    Therefore, some changes, modifications, and improvements may not be covered in this manual. If your unit’s operating characteristics or appearance differs from those described in this manual, please contact your Shel Lab dealer or distributor for assistance.
  • Page 6: Receiving Your Incubator

    Verify that the correct number of the following accessories are present: Included Accessories Model Tubing Kit Shelves Shelf Clips 115V Power Cord SCO58 SCO40 SCO31 Carefully check all packaging before discarding. Save the shipping carton until you are sure everything works properly. P a g e...
  • Page 7 RECEIVING YOUR INCUBATOR (CONTINUED) Orientation Photos SCO58 Control Panel Inlet (left) Sample Port (right) Data Ports Power Inlet Circuit Breaker Access Port OTL Temperature Probe (Top) Controller Board Temperature Probe (Bottom) Incubation Chamber Back Wall Shelf Mounting Standard Door Gasket...
  • Page 8 RECEIVING YOUR INCUBATOR (CONTINUED) SCO40 Control Panel Inlet Sample Port Data Jacks USB-Style Port OTL Temperature Probe (Top) Main Temperature Probe (Bottom) Incubation Chamber Power Inlet, Circuit Breaker, and Data Plate Door Gasket Incubator Door P a g e...
  • Page 9 RECEIVING YOUR INCUBATOR (CONTINUED) SCO31 Control Panel Inlet Sample Port Data Jacks USB-Style Port OTL Temperature Probe (Top) Controller Board Temperature Probe (Bottom) Incubation Chamber Power Inlet, Circuit Breaker, and Data Plate Door Gasket Incubator Door P a g e...
  • Page 10: Recording Data Plate Information

    The data plate contains the incubator model number and serial number. Record this information for future reference. On the SCO58 the data plate is located on the back of the unit, on the top right side. On the SCO31 and SCO40 the data plate is toward the bottom and back of the unit’s right side.
  • Page 11: Installation

    INSTALLATION MBIENT ONDITIONS The SCO Incubators are intended for use indoors, at room temperatures between 15C and 30C (59F and 86F), at no greater than 80% Relative Humidity (at 25C / 77F). Allow a minimum of 4 inches (10cm) between the incubator and walls or partitions, and 2 inches (5cm) of clearance above the top of the incubator for unobstructed airflow.
  • Page 12: Lifting

    CCESS TOPPER The SCO58 Incubator is provided with an access port located on the right side of the unit. The incubator is shipped with one (1) rubber access port stopper. The stopper should come installed in the port inside the incubation chamber. The stopper should always be installed in the chamber to obtain the best temperature uniformity and prevent condensation from forming inside the port.
  • Page 13: Connect The Co Supply To The Incubator

    INSTALLATION (CONTINUED) ONNECT THE UPPLY TO THE NCUBATOR Note: Always use medical grade CO . Use of non-medical grade CO risks introducing contaminants into the chamber, may damage the incubator, and will void the incubator’s warranty. Always install the in-line filter provided with the unit tubing kit between the gas source and the incubator.
  • Page 14: Shelving Installation

    1. Squeeze each shelf clip when installing. Insert the top tab first into shelf mounting standard, and then the bottom tab using a rocking motion. a. SCO58, install six (6) clips for each shelf. Two (2) each on the left, right, and back walls of the incubation chamber.
  • Page 15: Cleaning And Deionized Water

    Use of DI water in a Shel Lab incubator voids the unit’s warranty. Sheldon Manufacturing recommends the use of distilled water in the resistance range of 50K Ohm/cm to 1M Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0 uS/cm, for cleaning and any humidifying applications.
  • Page 16: Graphic Symbols

    GRAPHIC SYMBOLS The incubator is provided with multiple graphic symbols on its exterior and internal surfaces. These symbols identify hazards and the functions of the adjustable components, as well as important notes in the user manual. Symbol Definition Indicates that you should consult your service manual for further instructions. Indique que l'opérateur doit consulter le manuel d'utilisation pour y trouver les instructions complémentaires.
  • Page 17 GRAPHIC SYMBOLS (CONTINUED) Symbol Definition Indicates Potential Shock Hazard Signale danger électrique WEEE Directive compliant logo Indicates the unit should be recycled (Not disposed of in land-fill) Indique l’appareil doit être recyclé (Ne pas jeter dans une décharge) Indicates CO Indique gaz CO2 Indicates carbon dioxide gas content as a % Indique le gaz carbonique contenu en %...
  • Page 18: Control Panel Overview

    The main power switch on the control panel controls all power to the unit and its systems. The switch illuminates when in the I on position on the SCO58. A green pilot light labeled Power On illuminates on the SCO40 and SCO31...
  • Page 19 The circuit breaker in the SCO58 and SCO40 is a 20 amp breaker. In the SCO31 it is a 15 amp breaker.
  • Page 20: Operation

    OPERATION HEORY OF PERATION Heating and CO Control The incubator uses a microprocessor controller board wired to a solid state temperature probe, a heating element located in a recirculation duct, as well as a blower fan in the duct, to monitor and regulate the temperature within the incubation chamber.
  • Page 21 OPERATION (CONTINUED) Accessory Compatibility Make sure that any accessory equipment you will be using inside the incubator can safely and effectively operate at your selected temperature and CO set points. The Over Temperature Limit System (OTL) When set, the OTL system prevents runaway heating in the event of a failure of the microprocessor controller board or its thermometer probe by depowering the heating element whenever the temperature in the incubation chamber exceeds the OTL setting.
  • Page 22: Preparing The Incubator

    2. Verify the workspace power supply and incubator data plate requirements match. 3. SCO58 only: Verify that the side access port plug is in place. 4. You may play the sensor probe of a temperature reference device inside the incubation chamber at this time.
  • Page 23: Set The Temperature Set Point

    OPERATION (CONTINUED) ET THE EMPERATURE OINT Perform these steps to prepare the unit for calibration. The unit comes from the factory set to 37°C 1. Turn the Over Temperature Limit control dial clockwise to the maximum position indicated by the largest dot. This prevents the Over Temperature Limit system from interfering with the Set the Temperature Set Point and Temperature Calibration procedures.
  • Page 24: Calibrate The Temperature Display

    Check that the sensor is not in direct contact chamber through the door, with the shelving. or on the SCO58 through the access port. Place the port stopper back into the Note: A thermocouple sensor probe’s sleeve may be taped to the shelving, as long port after introducing the as the exposed copper end is 2 inches (5cm) above the shelf.
  • Page 25 OPERATION (CONTINUED) Temperature Calibration Continued 5. Place the unit in its temperature calibration mode and calibrate. a. Press and hold both the UP and DOWN arrow buttons simultaneously. b. The Temperature Display will show the letters “CO”, then begin flashing the current temperature value. c.
  • Page 26: Set The Over Temperature Limit

    OPERATION (CONTINUED) Temperature Calibration Continued 10. If the temperature readings of the incubator and the reference device still fall outside your laboratory protocol after three calibration attempts, contact Sheldon Technical Support for assistance. a. Three calibration attempts may be required to successfully calibrate units that are ±2°C or more out of calibration.
  • Page 27: Muting The Audible Temperature Alarm

    OPERATION (CONTINUED) UTING THE UDIBLE EMPERATURE LARM An audible and visual high or low deviation alarm will activate if the incubator chamber temperature deviates by 1°C above or below the temperature set point. The low deviation alarm has a delay of fifteen minutes.
  • Page 28: Calibrating The Co Display

    OPERATION (CONTINUED) ALIBRATING THE ISPLAY Note: The CO display and sensor come calibrated from the factory to a 5% concentration at 179 Feet (54 meters) above sea level. The calibration procedure verifies the accuracy of the CO reading and corrects for any measurement errors. Calibrate as often as required by your laboratory or production protocol, or regulatory requirements.
  • Page 29 OPERATION (CONTINUED) Calibrating the CO Display (Continued) If there is a difference between the readings, and that difference falls outside your laboratory protocol’s acceptable range, place the incubator into calibration mode by pressing and holding the CO Control Up and Down arrow buttons. ...
  • Page 30 OPERATION (CONTINUED) Calibrating the CO Display (Continued) Reference 6. Compare the reference CO analyzer reading to the reading of the incubator CO display. a. If the reference device and the incubator display readings are the same or the difference falls within the range of your laboratory protocol, the incubator is now calibrated for CO ...
  • Page 31: Muting The Audible Co Alarm

    OPERATION (CONTINUED) UTING THE UDIBLE LARM Visual high and low deviation indicator alarms will activate if the incubator’s CO level deviates 1% above or below the CO set point. An audible alarm will sound immediately for a high deviation. The low deviation audible alarm will sound after the visual low indicator alarm has been continually illuminated for fifteen (15) minutes.
  • Page 32: Loading The Incubator

    OPERATION (CONTINUED) OADING THE NCUBATOR Sheldon Manufacturing recommends allowing the unit to run for at least 24-hours heated to temperature and supplied with CO prior to loading samples after installing in a new location. Place items on the shelves inside the incubator chamber as evenly spaced as possible. Good spacing allows for maximum air circulation and a high degree of temperature uniformity.
  • Page 33: Data Output Capabilities

    A hardware driver and data logging software package for the USB-style port can be downloaded from the Shel Lab website. To download the software, visit the product page of any Shel Lab CO incubator and click on the large USB bar icon located near the bottom of the page.
  • Page 34: User Maintenance

    USER MAINTENANCE : Prior to any maintenance or service on this unit, disconnect the power cord from the power supply. Warning : Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher le cordon secteur de la Avertissement source d'alimentation. If a hazardous material or substance has spilled in the incubator, immediately initiate your site’s Hazardous Material Spill Containment protocol.
  • Page 35: Maintaining Atmospheric Integrity

    LECTRICAL OMPONENTS Electrical components do not require maintenance. If the incubator fails to operate as specified, please contact your Shel Lab Dealer or Sheldon Technical Support for assistance (please see page 5).
  • Page 36: Condensation And The Dew Point

    Remove humidity pans or other open containers of water.  SCO58 only: Is the access port stopper in place and properly seated? Reinsert the stopper, or replace if damaged or worn. 36 |...
  • Page 37: Unit Specifications

    We reserve the right to alter technical specifications at all times. EIGHT Model Shipping Net Weight 850 lbs. / 385.55 kg 710 lbs. / 322 kg SCO58 850 lbs. / 385.55 kg 710 lbs. / 322 kg SCO40 660 lbs. / 299.37 kg 610 lbs. / 276.69 kg SCO31...
  • Page 38: Temperature

    0 – 20% Concentration All Models EMPERATURE Model Range Uniformity Stability ± 0.5 @ 37C ± 0.1C SCO58 Ambient +8C to 60C ± 0.5 @ 37C ± 0.1C SCO40 Ambient +8C to 60C ± 0.5 @ 37C ± 0.1C SCO31 Ambient +8C to 60C...
  • Page 39: Parts And Consumables

    2700506 In-Line HEPA Filter 2800525 Tubing Kit 9710500 Door Gasket, Magnetic 3450585 SCO58 48 X 68 Inches Door Gasket SCO40 3450534 1 foot sections, requires 20 sections Door Gasket SCO31 3450534 1 foot sections, requires 17.25 sections Power Cord 20 Amp 2.5 Meters...