customer or the user should assure that it is used in such an environment
Emissions Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
RF Emissions CISPR 14-1
RF Emissions CISPR 15
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The CO Check is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.
Immunity Test
Conducted RF
IEC 61000-4-6
Compliance
level
Group 1
Class B
[Not Applicable]
[Not Applicable]
[See 5.2.2.1 c)
and Figure 1]
[See 5.2.2.1 c)
and Figure 1]
Complies
Complies
IEC 60601 Test level
3 Vrms 150 KHz to 80
MHz
Electromagnetic environment guidance
The CO Check uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The CO Check can be used in domestic,
light and heavy industrial environments.
The CO Check is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplied buildings
used for domestic purposes.
The CO Check is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that
supplied buildings used for domestic
purposes.
The CO Check is not suitable for
interconnection with other equipment.
The CO Check is not suitable for
interconnection with other equipment.
Compliance
Electromagnetic environment
level
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the CO Check, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance
[V
] V
1
d = [ 3.5 ] √ P
18
.
Need help?
Do you have a question about the CO Check Pro and is the answer not in the manual?